Skip to content

AVANOS 44627Recalls & reported problems

Avanos Medical, Inc. · MIC* Standard PEG Kit - Pull Method · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00350770446271), received between Jun 20, 2022 and May 13, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

14 other recalls cover devices cleared under the same FDA submission (K924065); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Material Too Soft/Flexible1

Patient problems

  • Foreign Body In Patient2
  • Obstruction/Occlusion2
  • Abscess1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2026Break

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENTS, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING TWO DIFFERENT PATIENTS. THIS IS THE SECOND OF TWO REPORTS. REFER TO 3006646024-2026-00031 FOR THE FIRST REPORT. REFER TO 3006646024-2026-00032 FOR THE SECOND REPORT. IT WAS REPORTED THAT A PATIENT WITH OESOPHAGEAL APLASIA UNDERWENT AN OESOPHAGECTOMY. POST-OPERATIVE OESOPHAGEAL STRICTURE REQUIRING THREE SESSIONS OF OESOPHAGEAL DILATION UP TO 18 MM. IT WAS NOT POSSIBLE TO REMOVE THE FEEDING TUBE COLLAR ENDOSCOPICALLY. THE FEEDING TUBE COLLAR WAS LEFT INTRAGASTRIC TO PASS NATURALLY. AN OPERATION TOOK PLACE ON (B)(6) 2026. THE PATIENT IS FINE NOW.

    Patient problems: Foreign Body In Patient; Obstruction/Occlusion

    MAUDE report 25161707 ↗
  • InjuryMay 13, 2026Break

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENTS, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING TWO DIFFERENT PATIENTS. THIS IS THE FIRST OF TWO REPORTS. REFER TO 3006646024-2026-00031 FOR THE FIRST REPORT. REFER TO 3006646024-2026-00032 FOR THE SECOND REPORT. IT WAS REPORTED THAT A GASTROSTOMY TUBE REMOVED ON (B)(6) 2026. IT WAS NOT POSSIBLE TO PASS THE UPPER OESOPHAGEAL SPHINCTER ENDOSCOPICALLY. THE TUBE COLLAR WAS LEFT IN THE STOMACH TO PASS NATURALLY. SUBSEQUENTLY, AN OBSTRUCTION OCCURRED. THE PATIENT UNDERWENT AN OPERATION ON (B)(6) 2026. THE PATIENT IS FINE NOW.

    Patient problems: Foreign Body In Patient; Obstruction/Occlusion

    MAUDE report 25161701 ↗
  • InjuryJun 20, 2022Material Too Soft/Flexible; Material Too Soft/Flexible

    IT WAS REPORTED 'THE FLANGE THAT HOLDS THE TUBE IS SO FLEXIBLE THAT THE TUBE COMES OUT OF THE STOMACH.' THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED 24-MAY-2022 STATED 'THE PATIENT'S CONDITION IS DEGRADED. PASSAGE OF NUTRITION OUTSIDE THE STOMACH, WALL ABSCESS. SURGICAL INTERVENTION TO DRAIN THE ABSCESS.'

    Patient problems: Abscess

    MAUDE report 14750572 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K924065. One submission often covers many models, so check the product and lot details in each record.

  • Z-2348-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors 20 Fr PUSH OTW Model/Catalog Number: 8170-20
  • Z-2342-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit 20 Fr PUSH OTW Model/Catalog Number: 7170-20 Product Description: Feeding Tube Kit Component: No
  • Z-2345-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit 20 Fr PULL Model/Catalog Number: 7180-20
  • Z-2350-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors 14 Fr PULL Model/Catalog Number: 8180-14
  • Z-2341-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit 14 Fr PUSH OTW Model/Catalog Number: 7170-14 Product Description: Feeding Tube Kit
  • Z-2344-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit 14 Fr PULL Model/Catalog Number: 7180-14
  • Z-2352-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors 24 Fr PULL Model/Catalog Number: 8180-24
  • Z-2343-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit 24 Fr PUSH OTW Model/Catalog Number: 7170-24 Product Description: Feeding Tube Kit Component: No
  • Z-2351-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors 20 Fr PULL Model/Catalog Number: 8180-20
  • Z-2347-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors 14 Fr PUSH OTW Model/Catalog Number: 8170-14 Product Description: Feeding Tube Kit Comp
  • Z-2346-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit 24 Fr PULL Model/Catalog Number: 7180-24
  • Z-2349-2026 · Apr 30, 2026 · Open, Classified
    Product Name: MIC Safety Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors 24 Fr PUSH OTW Model/Catalog Number: 8170-24
  • Z-0566-2022 · Dec 22, 2021 · Terminated
    MIC* Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors - 20 Fr-PULL
  • Z-0565-2022 · Dec 22, 2021 · Terminated
    MIC* Percutaneous Endoscopic Gastrostomy PEG Kit - 20 Fr-PULL

Other Tubes, gastrointestinal (and accessories) devices

All Tubes, gastrointestinal (and accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page