AVANOS 20-7368Recalls & reported problems
Avanos Medical, Inc. · CORFLO* Nasogastric/Nasointestinal Feeding Tube with Stylet · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00350770457550), received between Dec 13, 2022 and May 15, 2026 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Anxiety1
- Foreign Body In Patient1
- Perforation1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2026Break
IT WAS REPORTED THAT A NASOGASTRIC (NG) TUBE DEVELOPED A LARGE HOLE AND SEPARATED INTO TWO PARTS DURING USE AFTER THE PATIENT VOMITED, REQUIRING REMOVAL AND REPLACEMENT. THERE WAS NO REPORTED INJURY.
Patient problems: Anxiety; No Clinical Signs, Symptoms or Conditions
MAUDE report 25185477 ↗InjuryJan 13, 2023Break; Break
IT WAS REPORTED VIA FDA FDA USER FACILITY REPORT # MW5113771 THE FOLLOWING INFORMATION: "SMALL BORE FEEDING TUBE FRACTURE." ADDITIONAL INFORMATION RECEIVED 21-DEC-2022 STATED THE NASOGASTRIC TUBE (NG) TUBE WAS NOTED TO BE MALPOSITIONED. WHEN REMOVED, THE WEIGHTED TIP WAS RETAINED IN THE PATIENT. THE PATIENT REQUIRED ENDOTRACHEAL TUBE EXCHANGE AND A BRONCHOSCOPY FOR FOREIGN BODY REMOVAL. THE NG TUBE WAS IN PLACE 'LESS THAN 30 MINUTES, LIKELY EVEN LESS' PRIOR TO THE FRACTURE OCCURRING. THE NG TUBE WAS NOT FLUSHED WITH 10 CC OF WATER AFTER PLACEMENT WAS CONFIRMED, AND NOT FLUSHED PRIOR TO THE REMOVAL OF THE STYLET TO ACTIVE THE INTERNAL LUBRICANT OF THE NG TUBE. THE NG TUBE WAS IMMEDIATELY NOTED TO BE MALPOSITIONED AND THUS THE STYLET WAS NEVER REMOVED, NOR WAS ANYTHING USED TO FLUSH THE NG TUBE. THE FRACTURE APPEARS WHERE THE WEIGHTED TIP MEETS THE REST OF THE TUBE/STYLET. THERE WAS NO INDICATION THE NG TUBE WAS CLOGGED BECAUSE IT WAS NEVER USED. THE WEIGHTED NG TUBE TIP WAS LEFT INSIDE THE PATIENT¿S BODY. THE PATIENT WAS INTUBATED FOLLOWING CEREBRAL EDEMA FROM A SUBOCCIPITAL CRANIOTOMY FOR TUMOR DEBULKING. THE NG TUBE WAS PLACED PRIOR TO EXTUBATION OF THE ENDOTRACHEAL TUBE TO HELP F
Patient problems: Foreign Body In Patient
MAUDE report 16164715 ↗InjuryJan 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, THE ¿SMALL BOWEL FOLLOW-THROUGH (SBFT)¿ WAS BLINDLY PLACED, AN X-RAY CONFIRMED A LUNG PLACEMENT. THE TUBE WAS REMOVED; HOWEVER, A TIP ARTIFACT WAS LEFT IN THE BODY AND WAS REMOVED WITHOUT INCIDENT. THE PATIENT WAS REPORTEDLY DOING FINE.
Patient problems: Perforation
MAUDE report 16118586 ↗InjuryDec 13, 2022Fracture
SMALL BORE FEEDING TUBE FRACTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15979684 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports