Skip to content

AVANOS 45836Recalls & reported problems

Avanos Medical, Inc. · CORGRIP* Nasogastric/Nasointestinal Feeding Tube Retention System · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00350770458366), received between Feb 13, 2020 and May 23, 2024 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Use of Device Problem1

Patient problems

  • Foreign Body In Patient2
  • Insufficient Information2
  • Skin Erosion1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 23, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED, PATIENT WAS TAKEN TO MRI WITH THE MAGNET INTACT DURING THE TESTING. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 19388059 ↗
  • MalfunctionAug 17, 2022Break; Break

    IT WAS REPORTED "THE MAGNET DISCONNECTED FROM THE PROBE WHILE PLACING THE CORGRIP. THE PATIENT NEEDED FURTHER ASSISTANCE FROM THE ENT [EAR, NOSE AND THROAT] PHYSICIAN TO REMOVE THE MAGNET FROM THE NASAL CAVITY." THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED (B)(6)2022 STATED ENT WAS CONSULTED AND THE ENT SPENT 3-HOURS AT THE BEDSIDE TRYING TO RETRIEVE THE MAGNET AND "MAY HAVE CAUSED SOME TRAUMA." THERE WERE NO REPORTS OF ADVERSE EFFECTS RELATED TO THE REMOVAL PROCEDURE.

    Patient problems: Foreign Body In Patient; Unspecified Tissue Injury

    MAUDE report 15245775 ↗
  • InjuryMar 2, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING AN ENDOSCOPIC PROCEDURE, THE CORGRIP BECAME LODGED INSIDE THE PATIENT¿S NARE (NOSTRIL); THE CORGRIP WAS UNABLE TO BE EXTRACTED. AN (ENT) EAR NOSE AND THROAT PHYSICIAN WAS CONSULTED AND THE CORGRIP WAS REMOVED. THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED 07-FEB-2022 PROVIDED PATIENT INFORMATION. ADDITIONAL INFORMATION RECEIVED 11-FEB-2022 REPORTED THAT THE CLINICIAN BELIEVED IT WAS THE CLIP THAT WAS ATTACHED TO THE CORTRAK TUBE THAT WAS INSIDE THE NARE.

    Patient problems: Insufficient Information

    MAUDE report 13655753 ↗
  • InjuryFeb 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE CORGRIP TUBE RETENTION DEVICE CAUSED NASAL EROSION AND WOUND CARE THERAPY WAS REQUIRED FOR THE PATIENT. ADDITIONAL INFORMATION RECEIVED 28-JAN-2021 STATING THE DEVICE WAS PLACED ON (B)(6) 2021 AND THE WOUND WAS NOTED ON (B)(6) 2021. MEDICAL INTERVENTION ENTAILED A WOUND CARE CONSULT, WITH ADDITIONAL WOUND FOLLOW-UP AND THE TUBE RETENTION DEVICE WAS REPOSITIONED. THE PATIENT IS REPORTED TO BE "RECOVERING."

    Patient problems: Skin Erosion

    MAUDE report 11349866 ↗
  • MalfunctionFeb 13, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A "NURSE WAS PLACING THE CORGRIP IN THE PATIENT. SHE HEARD THE SNAP WHEN THE TWO ENDS CONNECTED. SHE WHEN SHE TRIED TO PULL THE STRING THROUGH, SHE COULDN'T GET IT THROUGH. SHE REMOVED THE TWO ENDS, AND THE MAGNET WAS MISSING." NO PATIENT INJURY WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 9705732 ↗

Other Tubes, gastrointestinal (and accessories) devices

All Tubes, gastrointestinal (and accessories) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.