Skip to content

AVANOS 45837Recalls & reported problems

Avanos Medical, Inc. · CORGRIP* Nasogastric/Nasointestinal Feeding Tube Retention System · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00350770458373), received between Mar 6, 2023 and Oct 25, 2023 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Loss of or Failure to Bond1

Patient problems

  • Foreign Body In Patient2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A PATIENT DEVELOPED A WOUND ON THE NOSE THAT "APPEARED TO BE RELATED TO THE STRING PULLING MORE TO THE SIDE WITH THE WEIGHT OF THE TUBE." ADDITIONAL INFORMATION RECEIVED 03-OCT-2023 STATING "THE CORGRIP WAS PLACED ON (B)(6) 2023. THE WOUND WAS ON THE BRIDGE OF THE NOSE. IT APPEARED THE STRING ITSELF WAS RUBBING ON THIS AREA, NOT THE CLIP ITSELF. PER THE NURSE, THERE WAS SPACE BETWEEN THE CLIP AND THE STRING, BUT THE ACTUAL TUBE WAS PULLING THE BRIDLE TO THE SIDE AND RUBBING ONTO THE NOSE. NONE OF THE DIETITIANS ACTUALLY VISUALIZED THIS FOR CONFIRMATION HOWEVER BECAUSE THE BRIDLE WAS ALREADY REMOVED ONCE THE WOUND WAS IDENTIFIED, SO IT IS BASED ON REPORT ONLY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18007447 ↗
  • MalfunctionSep 19, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED BY A NURSE, ¿DURING THE INSERTION OF A FIXATION SYSTEM FOR A NASOGASTRIC FEEDING TUBE (CORGRIP), ONE OF THE RODS WAS ¿INHALED¿ BY THE PATIENT TO THE BACK OF HIS THROAT, WHILE THE END IS PROVIDED WITH A THICKER TIP. A NECK X-RAY SHOWING THIS ROD IS AVAILABLE IN THE FILE PATIENT'S MEDICAL.¿ NO INJURY OR MEDICAL INTERVENTIONS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 17777356 ↗
  • MalfunctionMar 6, 2023Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED, THAT UPON REMOVAL OF THE WHITE CATHETER THE UMBILICAL TAPE WAS NOTED TO NO LONGER HAVE THE MAGNET AND THEY BELIEVED THE MAGNET WAS STILL INDWELLING IN THE PATIENT; THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED 13FEB2023 REPORTED, THE NASOGASTRIC (NG) TUBE WAS LEFT IN THE PATIENT AND SECURED WITH TAPE. ON (B)(6) 2023 THE EAR, NOSE AND THROAT (ENT) DOCTOR PERFORMED A NASAL ENDOSCOPE WHICH REVEALED THE PATIENT HAD A DEVIATED NASAL SEPTUM AND NASAL POLYPS; HOWEVER, THEY WERE UNABLE TO LOCATE THE CORGRIP MAGNET. IT WAS ADDITIONALLY REPORTED, THE FAMILY MADE THE DECISION TO NOT PERFORM ANY FURTHER TESTING TO LOCATE THE MAGNET AS THE PATIENT WAS ON END-OF-LIFE CARE FOR A DIAGNOSIS OF METASTATIC BONE CANCER. THE PATIENT EXPIRED, ON (B)(6) 2023, ¿DUE TO [THE] DIAGNOSIS OF METASTATIC BONE CANCER¿.

    Patient problems: Foreign Body In Patient

    MAUDE report 16492377 ↗

Other Tubes, gastrointestinal (and accessories) devices

All Tubes, gastrointestinal (and accessories) devices →