AVANOS 40-1368Recalls & reported problems

Avanos Medical, Inc. · CORFLO* Nasogastric/Nasointestinal Feeding Tube with ENFit® Connector · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00350770459998), received between Jul 28, 2023 and Sep 18, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Material Puncture/Hole1
  • Material Rupture1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • Vomiting1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED "PARENT REPORTED THAT PATIENT HAD REMOVED NG [NASOGASTRIC] AND STATED THERE APPEARED TO BE A DEFECTIVE AREA TO TUBE. ON INSPECTION OF 8FR NON-WEIGHTED NG TUBE, THERE WAS AN AREA ALONG THE TUBE THAT APPEARED TO BE A DEFLATED/EXPANDED AREA, OTHERWISE NG WAS INTACT FROM TIP TO WHERE MEDICATIONS WOULD BE INSERTED. WATER FLUSH PASSED THROUGH TUBE IN WHICH IT WOULD LEAK AT THE DEFECTIVE AREA WHILE ALSO REACHING THE TIP OF THE TUBE. WHEN NG WAS INSERTED, [REPORTER] WAS ABLE TO OBTAIN GASTRIC CONTENTS AND PUSH WATER THROUGH. AFTER MEDICATION WAS INSERTED TO TUBE, RESISTANCE WAS MET. FLUSHES AND MEDS WERE PUSHED THROUGH, IN WHICH RESISTANCE IMPROVED SLIGHTLY WITH EACH FLUSH. NO SIGNS OF DISTRESS FROM PATIENT WITH EACH FLUSH SUCH AS COUGHING OR VOMITING. THERE WAS NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23084729 ↗
  • MalfunctionSep 11, 2025Break; Break

    IT WAS REPORTED THE CLINICIANS "WERE NOT ABLE TO GET STOMACH CONTENTS, CHECKED WITH AIR, AIR POP HEARD. RESISTANCE WITH FLUSHING NG [NASOGASTRIC]. REMOVAL OF NG COMPLETED AT 1410, BRIDAL CUT, DID NOT CUT NG TUBE. NG REMOVED WITH NO RESISTANCE, NG FOUND TO BE BROKEN AT 19CM MARK, X-RAY COMPLETED AND NOTED REST OF NG APPROXIMATELY 19 CM IN STOMACH. [PATIENT] TO BE SENT DOWN TO REMOVE SAME VIA SCOPE." THERE WAS NO SERIOUS HARM REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 23034346 ↗
  • MalfunctionJun 24, 2024Break; Break

    IT WAS REPORTED "A CHEST [X-RAY WAS] PERFORMED ON (B)(6) 2024 AND INCIDENTAL FINDING OF THE BROKEN NG [NASOGASTRIC] TUBE. THE NG TUBE WAS REMOVED ON (B)(6) 2024 AND INTACT BUT THE TIP WAS BARLEY CONNECTED." THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED 30-MAY-2024 STATED NO MEDICAL INTERVENTION WAS REQUIRED, THE TUBE WAS IN PLACE FOR 1-DAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19593327 ↗
  • MalfunctionJul 28, 2023Material Rupture; Stretched; Material Rupture; Stretched

    IT WAS REPORTED, THE NASOGASTRIC TUBE (NGT) WAS PLACED BY INTERVENTIONAL RADIOLOGY ON (B)(6) 2023. THE PATIENT VOMITED THE TUBE ON (B)(6) 2023, AND PART OF IT BROKE OFF IN THE PATIENT¿S STOMACH. IT WAS LEFT THERE AND ANOTHER TUBE WAS PLACED IN THE INFUSION CENTER THAT DAY. THE BROKEN PART WAS EVENTUALLY REMOVED AT A LATER DATE. PER ADDITIONAL INFORMATION RECEIVED ON 17JUL2023, IT IS UNKNOWN IF THE TUBE WAS EVER CLOGGED. ESOPHAGOGASTRODUODENOSCOPY WITH FOREIGN BODY REMOVAL WAS PERFORMED ON (B)(6) 2023. THE PATIENT RECOVERED WELL.

    Patient problems: Vomiting

    MAUDE report 17415195 ↗

Other Tubes, gastrointestinal (and accessories) devices

All Tubes, gastrointestinal (and accessories) devices →