AVANOS 40-4431Recalls & reported problems
Avanos Medical, Inc. · CORFLO* Nasogastric/Nasointestinal Feeding Tube with Stylet with ENFit® Connector · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00350770460185), received between Apr 15, 2022 and May 23, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Material Split, Cut or Torn1
- Physical Resistance/Sticking1
Patient problems
- Foreign Body In Patient3
- High White Blood Cell Count1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 23, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THE PATIENT'S, "NG [NASOGASTRIC TUBE] WAS NEARLY FULLY PULLED OUT, WITH TUBE FEED STILL RUNNING AT 10 ML/HR. TUBE FEED STOPPED IMMEDIATELY, NG FULLY REMOVED. THE REASON THE [PATIENT] TUBE FEED KEPT RUNNING WITH THE NG PARTIALLY REMOVED WAS BECAUSE THE WEIGHTED, WIDER TIP OF THE NG WAS STUCK IN THE NASOPHARYNX AND REQUIRED SOME DECENT MANIPULATION TO FULLY REMOVE." THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22081887 ↗MalfunctionJun 24, 2024Break; Break
IT WAS REPORTED "CORFLO TUBE [BROKE] APART INSIDE PATIENT. PATIENT HAD TUBE IN FOR 33-DAYS." THE "PATIENT IS VERY SICK NOW AND [THE CLINICIANS] WERE LOOKING FOR SOURCE SO DID A CT SCAN AND SAW SOMETHING WRONG." ADDITIONAL INFORMATION RECEIVED 17-JUN-2024 STATING "THE PATIENT DID EXPERIENCE LEUKOCYTOSIS WHICH PEAKED WHEN IT IS ASSUMED THE TUBE FIRST BROKE ((B)(6) 2024). PATIENT HAS HAD TO UNDERGO SERIAL IMAGING TO SEE WHERE THE RETAINED PART OF THE TUBE HAS MIGRATED. AT THIS TIME DUE TO PATIENT CONDITION, WE ARE UNDERGOING CONSERVATIVE MANAGEMENT, BUT SURGICAL REFERRAL HAS BEEN RECOMMENDED BY GI (GASTROINTESTINAL) AS THE PART IS NOT REACHABLE BY ENDOSCOPIC MEANS." THE USER FACILITY CLINICIAN HAS THE TOP HALF OF THE TUBE BUT AS STATED ABOVE THE BOTTOM HALF REMAINS IN THE PATIENT. NO MEDICAL INTERVENTION HAS OCCURRED AT THIS TIME.
Patient problems: Foreign Body In Patient; High White Blood Cell Count
MAUDE report 19593848 ↗MalfunctionJul 14, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED, A PATIENT PULLED THE NASOGASTRIC TUBE (NGT) TUBE OUT. APPROXIMATELY FOUR INCHES OF THE DISTAL PORTION OF THE NGT BROKE OFF IN THE STOMACH. AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE WAS PERFORMED TO REMOVE THE REMAINING PORTION OF THE NGT FROM THE PATIENT'S STOMACH. NO INJURY REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 17324398 ↗MalfunctionApr 15, 2022Break; Break
IT WAS REPORTED THAT A NASOGASTRIC (NG) TUBE BROKE IN TWO PIECES AFTER 15 DAYS IN PLACE. THE BROKEN TUBE WAS DISCOVERED WHEN FEEDING CONTENTS WERE NOTED IN THE PATIENT'S MOUTH AND NARES; AN X-RAY OF THE ABDOMEN WAS PERFORMED. UPON REMOVAL OF THE TUBE THE DISTAL PORTION OF THE TUBE WAS RETAINED WITHIN THE PATIENT. AN EGD (ESOPHAGOGASTRODUODENOSCOPY) WAS PERFORMED BY GASTROENTEROLOGY TO RETRIEVE THE RETAINED PORTION OF THE NG TUBE; A NEW FEEDING TUBE WAS PLACED. ADDITIONAL INFORMATION RECEIVED 23MAR2022; THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION.
Patient problems: Foreign Body In Patient
MAUDE report 14119635 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗