Cortrak 2 Enternal access device 46044Recalls & reported problems
Avanos Medical, Inc. · CORTRAK* 2 Nasogastric/Nasointestinal Feeding Tube with Electromagnetic Transmitting Stylet with ENFit® Connector · Class 2
All models of Cortrak 2 Enternal access device (14)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00350770460444), received between May 27, 2022 and Jun 9, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Material Fragmentation1
- Material Integrity Problem1
Patient problems
- Foreign Body In Patient4
- No Clinical Signs, Symptoms or Conditions1
- Unspecified Gastrointestinal Problem1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2026Break
THE CUSTOMER REPORTED THAT A FEEDING TUBE BROKE WHILE IN USE AND THAT A PORTION OF THE DEVICE BROKE INSIDE THE PATIENT. THE CUSTOMER SUBSEQUENTLY REQUESTED MANUFACTURER FLUSHING GUIDELINES FOR THE FEEDING TUBE. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 25448977 ↗MalfunctionJun 9, 2026Break
IT WAS REPORTED THAT A CORTRAK FEEDING TUBE WAS FOUND TO BE FRAYED AND PARTIALLY BROKEN IN THE MID-SECTION AFTER REMOVAL FROM THE PATIENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25443423 ↗MalfunctionNov 14, 2023Break; Break
IT WAS REPORTED VIA FDA USER FACILITY REPORT MW REPORT MW5146866 THE FOLLOWING INFORMATION: ON (B)(6) 2023, THE PATIENT'S NASAL DUODENAL (ND) TUBE CLOGGED AND DESPITE MULTIPLE ATTEMPTS TO FLUSH AND USE CLOT-BUSTER MEDICATIONS, OBSTRUCTION WAS NOT CLEARED. WHEN ND WAS PULLED, THE DISTAL PORTION BROKE OFF AND REMAINS IN PATIENT'S ESOPHAGUS. GI WAS CONSULTED AND PATIENT WAS TAKEN TO OPERATING ROOM ON (B)(6) 2023 TO HAVE SURGICALLY REMOVED. PORTION OF THE PRODUCT WAS SAVED. ADDITIONAL INFORMATION RECEIVED 01-NOV-2023 STATING THE PATIENT RETURNED TO (OPERATING ROOM) OR FOR REMOVAL OF THE DEVICE. THE DEVICE WAS INITIALLY INSERTED ON (B)(6) 2023. THE PATIENT IS CURRENTLY DECEASED, NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 18133936 ↗InjuryAug 4, 2022Material Fragmentation; Material Fragmentation
FDA MEDWATCH / FDA USER FACILITY REPORT # MW 0504540000-2022-8016, REPORTED, THE 8FR CORTRAK 2 NASOJEJUNAL (NJ) TUBE WAS PLACED TO ADMINISTER NJ GASTRIC FEEDINGS AND ORAL MEDICATIONS. 10 DAYS LATER THE TUBE WAS CLAMPED, CLOG ZAPPER USED, MEDICATIONS AND FLUSH. AN X-RAY WAS PERFORMED APPROXIMATELY 2 WEEKS LATER AND THE ENTERIC TUBE APPEARED FRACTURED WITH THE TIP REMAINING AT THE FOURTH PORTION OF THE DUODENUM. THE TUBE WAS REMOVED WITH 3CM OF THE TUBE MISSING. THE FRACTURED ENTERIC TUBE WAS SEEN ON IMAGING AND THE DISTAL PORTION HAD MIGRATED BEYOND POINT OF ENDOSCOPIC RETRIEVAL; NO ENTERAL TUBE WAS PLACED. THE PATIENT REPORTEDLY RECEIVED PACKET RED BLOOD CELL (PRBC) X 1, PLATELET COUNT (PLT) X 2 AND 2 PLATELET TRANSFUSIONS DUE TO GASTROINTESTINAL (GI) BLEEDING. IT IS UNKNOWN IF THE DISTAL PORTION WAS REMOVED VIA MEDICAL INTERVENTION.
Patient problems: Foreign Body In Patient; Unspecified Gastrointestinal Problem
MAUDE report 15166834 ↗MalfunctionMay 27, 2022Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE CORTRAK TUBE WAS PLACED ON (B)(6) 2022 AND WAS REMOVED, BY ESOPHAGOGASTRODUODENOSCOPY (EGD) DUE TO IT BEING DAMAGED, (IT WAS BROKEN IN HALF). THERE WAS NO REPORTED INJURY.
Patient problems: Foreign Body In Patient
MAUDE report 14529558 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page