AVANOS 42-9438Recalls & reported problems
Avanos Medical, Inc. · CORFLO* Nasogastric/Nasointestinal Feeding Tube with Stylet with ENFit® Connector · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00350770460963), received between Apr 26, 2024 and May 12, 2026 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Fluid/Blood Leak2
- Material Puncture/Hole2
- Leak/Splash1
- Material Protrusion/Extrusion1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient3
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 12, 2026Leak/Splash; Material Puncture/Hole
IT WAS REPORTED THAT A TUBE WAS INSERTED AND ITS POSITION CHECKED. THE GUIDEWIRE WAS REMOVED. FOLLOWING FLUSHING, LEAKAGE OF WATER WAS SEEN FROM THE END OF THE TUBE (CLOSE TO THE CONNECTOR). APPROXIMATELY 5-10MM FROM THE CONNECTOR THERE WAS A VISIBLE SMALL PUNCTURE (0.5MM) IN THE TUBE. THE TUBE WAS THEREFORE REMOVED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25154847 ↗MalfunctionMay 4, 2026Material Protrusion/Extrusion
IT WAS REPORTED THAT THE GUIDEWIRE OF THE NASOGASTRIC TUBE IS SHOWING THROUGH THE TIP. NORMALLY IT STOPS BEFORE THE TIP. NO INJURIES OR MEDICAL INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25071716 ↗MalfunctionApr 23, 2026Material Puncture/Hole
IT WAS REPORTED THAT A HOLE WAS FOUND IN THE NASOGASTRIC TUBING BY THE NURSING STAFF BY THE CONNECTOR. NO MEDICAL INTERVENTION, OTHER THAN REPLACEMENT THE TUBE, WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24975126 ↗MalfunctionApr 11, 2026Fluid/Blood Leak
IT WAS REPORTED THERE WAS A TEAR BETWEEN FROM Y-CONNECTOR AND TUBE AND NUTRITION LEAKED FROM THERE. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24849161 ↗MalfunctionJul 2, 2025Break; Break
IT WAS REPORTED THE "DEVICE WAS BLOCKED OVERNIGHT AND NURSING STAFF FLUSHED IN ATTEMPT TO UNBLOCK. A REPORTED 'POP' WAS FELT; HOWEVER THE NG (NASOGASTRIC) TUBE WAS REMOVED AT THE TIME WHERE THE END WAS FOUND TO BE FRAYED AND THE DISTAL 20CM OF TUBE WAS MISSING. X-RAY IMAGING CONFIRMED THE PRESENCE OF 20CM OF RETAINED DISTAL NG [NASOGASTRIC] TUBE WITHIN THE PATIENT'S STOMACH AND SMALL BOWEL. THERE WAS NO REPORTED INJURY.
Patient problems: Foreign Body In Patient
MAUDE report 22384202 ↗MalfunctionApr 11, 2025Break; Break
IT WAS REPORTED "APPROXIMATELY 20 CM OF TUBE BECAME DETACHED AND WAS RETAINED AFTER REMOVAL. IT WAS SEEN TO BE IN THE DUODENUM ON AN ABDOMINAL CT [COMPUTED TOMOGRAPHY] SCAN." ADDITIONAL INFORMATION RECEIVED 21-MAR-2025 STATED, "THE BROKEN END OF THE TUBE WAS REMOVED VIA ENDOSCOPY."
Patient problems: Foreign Body In Patient
MAUDE report 21829822 ↗MalfunctionOct 1, 2024Break; Break
IT WAS REPORTED, ¿DURING FLUSHING TUBE SPLIT AND PATIENT NEEDED TO BE SENT TO ENDOSCOPY FOR RETRIEVAL OF THE LOOSE END OF TUBE LEFT INSIDE OF PATIENT FOLLOWING SPLITTING." THE PATIENT WAS REPORTED TO BE ¿ALIVE AND WELL.¿
Patient problems: Foreign Body In Patient
MAUDE report 20346383 ↗MalfunctionApr 26, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED "THERE WAS A LEAK FROM THE TUBE. THE TUBE WAS STARTED TO USE SINCE (B)(6) 2024. WHEN THE TUBE WAS FLUSHED ON (B)(6) 2024, A NURSE NOTICED THAT PATIENT'S NOSE WAS WET. THE NURSE CHECKED AND FOUND A TEAR ON THE TUBE." THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19191709 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page