AVANOS 46223Recalls & reported problems

Avanos Medical, Inc. · CORFLO* Standard PEG Kit - Pull Method · Class 2 · Implant

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00350770462233), received between Mar 6, 2023 and Oct 29, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation3
  • Break2

Patient problems

  • Foreign Body In Patient5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2025Break; Break

    IT WAS REPORTED THAT WHEN TRYING TO TRACTION REMOVE THE CORFLO PEG [PERCUTANEOUS ENDOSCOPIC GASTROSTOMY] AFTER 3-MONTHS, THE INNER BUMPER GOT LOOSE AND THE TUBING CAME OUT. THE INNER BUMPER WAS STILL IN THE PATIENT. A GASTROSCOPY WAS NEEDED TO REMOVE THE INNER BUMPER. ADDITIONAL INFORMATION RECEIVED 20-OCT-2025 STATED "THE CORFLO PEG WAS INSERTED (B)(6) 2025 AND REMOVED (B)(6) 2025." THE PATIENT DIDN'T NEED A MIC-KEY BUTTON, THE PATIENT COULD EAT FOOD NORMALLY BY HIMSELF.

    Patient problems: Foreign Body In Patient

    MAUDE report 23416824 ↗
  • MalfunctionJan 10, 2025Break; Break

    IT WAS REPORTED THE PEG [PERCUTANEOUS ENDOSCOPIC GASTROSTOMY] TUBE "INNER FIXATION STOP CAME LOOSE FROM THE TUBING WHEN TRYING TO PULL IT OUT/TRACTION REMOVE." THE DEVICE WAS IN USE FOR SIX MONTHS. ADDITIONAL INFORMATION RECEIVED 17-DEC-2024 NOTED THE CLINICIAN LEFT "STOP PLATE IN PLACE AND INFORMED THE PATIENT OF THE SYMPTOMS TO WATCH OUT FOR." ADDITIONAL INFORMATION RECEIVED (B)(6) 2024 12-19 NOTED THE PATIENT WAS "GOOD WHEN LEFT THE CLINIC, LIVING AT HOME."

    Patient problems: Foreign Body In Patient

    MAUDE report 21130025 ↗
  • MalfunctionMar 6, 2023Material Separation; Material Separation

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 3006646024-2023-00013 FOR THE FIRST REPORT. REFER TO 3006646024-2023-00014 FOR THE SECOND REPORT. IT WAS REPORTED, "THE INTERNAL STOP PLATE COMES OFF AND REMAINS IN THE STOMACH AND HAS TO COME OUT THE NATURAL WAY. IT HAS HAPPENED SEVERAL TIMES." NO INJURY OR MEDICAL INTERVENTIONS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 16486871 ↗
  • MalfunctionMar 6, 2023Material Separation; Material Separation

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 3006646024-2023-00013 FOR THE FIRST REPORT. REFER TO 3006646024-2023-00015 FOR THE THIRD REPORT. IT WAS REPORTED, "THE INTERNAL STOP PLATE COMES OFF AND REMAINS IN THE STOMACH AND HAS TO COME OUT THE NATURAL WAY. IT HAS HAPPENED SEVERAL TIMES." NO INJURY OR MEDICAL INTERVENTIONS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 16486868 ↗
  • MalfunctionMar 6, 2023Material Separation; Material Separation

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 3006646024-2023-00014 FOR THE SECOND REPORT. REFER TO 3006646024-2023-00015 FOR THE THIRD REPORT. IT WAS REPORTED, "THE INTERNAL STOP PLATE COMES OFF AND REMAINS IN THE STOMACH AND HAS TO COME OUT THE NATURAL WAY. IT HAS HAPPENED SEVERAL TIMES." NO INJURY OR MEDICAL INTERVENTIONS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 16486863 ↗

Other Tubes, gastrointestinal (and accessories) devices

All Tubes, gastrointestinal (and accessories) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.