AVANOS 46941Recalls & reported problems
Avanos Medical, Inc. · CORFLO* Safety PEG Kit - Pull Method · Class 2 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00350770469416), received between Feb 26, 2024 and Dec 17, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Material Separation1
Patient problems
- Foreign Body In Patient1
- Localized Skin Lesion1
- Pain1
- Skin Inflammation/ Irritation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, ¿A 52-YEAR-OLD PATIENT WAS HOSPITALIZED ON (B)(6) 2025, FOR A GASTROSTOMY TUBE REPLACEMENT. CHRONOLOGY OF EVENTS: (B)(6) 2025: INITIAL PLACEMENT OF A CORFLO GPE GASTROSTOMY TUBE USING THE PULL TECHNIQUE. THE ENDOSCOPY NURSES PROVIDED THE PATIENT'S NURSES WITH INFORMATION REGARDING GASTROSTOMY CARE. (B)(6) 2025: A FOLLOW-UP APPOINTMENT WITH THE PHYSICIAN AT 3 WEEKS WAS SCHEDULED. DURING THIS APPOINTMENT, A SKIN EROSION AROUND THE GASTROSTOMY SITE WAS OBSERVED, DUE TO PRESSURE FROM THE EXTERNAL FLANGE. A HYDROCOLLOID DRESSING WAS APPLIED TO THE EROSION, AND A PRESCRIPTION WAS ISSUED TO CONTINUE CARE. THE TUBE'S SKIN FIXATION SYSTEM IS A RIGID CLIP SYSTEM, DIFFICULT TO DISLODGE. THE PATIENT WAS GIVEN THE TUBE SUPPLIER'S INFORMATION BOOKLET. (B)(6) 2025: FOLLOWING A CONSULTATION WITH A HEPATOLOGIST, IT WAS NOTED THAT THE GASTROSTOMY OPENING WAS INFLAMED AND PAINFUL. (B)(6) 2025: THE GASTROSTOMY TUBE WAS CHANGED TO A STANDARD BALLOON TUBE WITH A FLANGED FIXATION SYSTEM. PATIENT'S CURRENT CONDITION: SIGNIFICANT LESIONS AT THE GASTROSTOMY SITE. REOPERATION FOR TUBE REPLACEMENT IN A FRAIL PATIENT. THE PATIENT IS CURRENTLY HOSPITALIZED IN A VASCULAR REHABILITATION UNIT (IN
Patient problems: Localized Skin Lesion; Pain; Skin Inflammation/ Irritation
MAUDE report 23827761 ↗MalfunctionFeb 26, 2024Material Separation; Material Separation
IT WAS REPORTED, IN THE DIGESTIVE EXPLORATION DEPARTMENT, WHEN THE PROBE WAS BEING REMOVED, THE RETAINING BUMPER BROKE OFF. THE RETAINING BUMPER FELL INTO THE STOMACH. A FORCEPS EXTRACTION WAS PERFORMED, PROLONGING THE OPERATING TIME. THIS WAS AN ISOLATED INCIDENT. NO INJURY REPORTED. PATIENT WAS REPORTED TO BE OK.
Patient problems: Foreign Body In Patient
MAUDE report 18782074 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.