Dual-Ended Cleaning Brush 92647Recalls & reported problems
Avanos Medical, Inc. · Dual-Ended Cleaning Brush · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00350770926476), received between Dec 27, 2024 and Mar 10, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 10, 2026Break
IT WAS REPORTED VIA FDA MW5182019, "ON (B)(6) 2025 THE PATIENT UNDERWENT A COLONOSCOPY. DURING SNARING OF A POLYP IN THE COLON, A CLEANING BRUSH THAT HAD BROKEN OFF AND REMAINED INSIDE THE SCOPE FROM A PRIOR COLONOSCOPY DISLODGED AND FELL INTO THE PATIENT'S COLON. ON (B)(6) 2025 GI ENDO PROCEDURE IN PROGRESS AND A SMALL BRUSH CAME OUT OF THE SCOPE AND LANDED IN THE PTS STOMACH, IT WAS THEN REMOVED BY THE GI DOC (GASTROINTESTINAL DOCTOR). ALL BRUSHES WERE REMOVED FROM STOCK, MANUFACTURER NOTIFIED."
Patient problems: Foreign Body In Patient
MAUDE report 24563254 ↗MalfunctionDec 27, 2024Break; Break
IT WAS REPORTED THE "SMALL 5.0MM DIAMETER BRUSH HEAD HAS DETACHED AND LODGED INSIDE OUR GASTROSCOPE...WE DO NOT KNOW HOW LONG THE BRUSH HEAD WAS STUCK INSIDE THE SCOPE (THINKING A DAY) SINCE WHEN WE TAKE BIOPSIES THE FORCEPS DIAMETER WILL PUSH OUT ANY FOREIGN OBJECTS. WE USED FORCEPS 2 -DAYS AGO SO I'M ASSUMING IT WAS FROM REPROCESSING AFTER THAT DAY USE. WE LOCATED THE BRUSH HEAD AFTER THE SCOPE FAILED TO SUCTION, THE SCOPE WAS INSIDE OF A PATIENT, SO WE ARE MONITORING FOR INFECTION." ADDITIONAL INFORMATION RECEIVED 06-DEC-2024 STATED THE USER FACILITY DID NOT KNOW HOW LONG THE BRUSH HEAD WAS STUCK INSIDE THE SCOPE (THINKING A DAY) SINCE WHEN [THEY] TAKE BIOPSIES THE FORCEPS DIAMETER WILL PUSH OUT ANY FOREIGN OBJECTS."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21029000 ↗
Other ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE devices
- Hedgehog SBD-289Boston Scientific Corporation · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE129 reports
- Channel cleaning brush BW-20TOlympus Medical Systems Corp. · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE90 reports
- MetriSPONGE 10-4025Metrex Research, LLC · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
- VERIFY LCC101Steris Corporation · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
- MEDLINE DYK500FSAWMedline Industries, Inc. · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
- MEDLINE DYK500FSMedline Industries, Inc. · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
- MEDLINE DYK1000TSEMedline Industries, Inc. · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
- MEDLINE DYK215FSMedline Industries, Inc. · ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page