AVANOS 95656Recalls & reported problems
Avanos Medical, Inc. · MIC-KEY* Gastric-Jejunal Feeding Tube Kit · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00350770956565), received between Oct 22, 2020 and Feb 8, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- May 31, 2024Open, Classified
The IFU was missing some risk information requested by another country to be contained in the IFU pertaining to potential complications.
Z-2422-2024 · Root cause: Error in labeling · FDA record ↗
Most reported problems
Device problems
- Break1
- Material Separation1
Patient problems
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 8, 2021Break; Break
IT WAS REPORTED THE PATIENT HAD AND ENTERAL FEEDING TUBE PLACED IN (B)(6) OF 2020 WHICH "DEVELOPED A BREAK BETWEEN THE G AND J TUBES INSIDE, AND THE FEEDING WENT INTO THE STOMACH AND LEAKED OUT OF TUBE." THERE WAS NO REPORTED INJURY BUT THE PATIENT WAS TAKEN TO THE HOSPITAL FOR TUBE REPLACEMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11296819 ↗MalfunctionOct 22, 2020Material Separation; Material Separation
IT WAS REPORTED THAT THE FEEDING TUBE, "COMPLETED DETACHED 15CM FROM THE BUTTON. 30CM WAS LEFT IN THE STOMACH. TUBE IN STOMACH WAS REMOVED BY ENDOSCOPY. NO HARM CAME TO PATIENT."
Patient problems: Foreign Body In Patient
MAUDE report 10719193 ↗
Other Tubes, gastrointestinal (and accessories) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.