Gastrointestinal Anchor Set, SAF-T-PEXY* T-Fasteners 98701Recalls & reported problems
Avanos Medical, Inc. · Gastrointestinal Anchor Set, SAF-T-PEXY* T-Fasteners · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Avanos SAF-T-PEXY T-Fastener gastrointestinal anchor set is used to secure a gastrostomy (G-) tube. The 63 reports mostly describe fasteners breaking, fracturing, or rupturing — sometimes during placement in surgery, sometimes before use — plus fasteners coming apart early, shifting, or becoming buried in tissue, with reports of pain and pressure injuries. One report mentions a death and 10 mention injuries, with 52 malfunctions. No recalls name this device. Reports dropped from 16 in the prior 12 months to 3 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 61
- Deaths reported
- 1
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 61 adverse event reports that list this exact device (DI 00350770987019), received between Oct 29, 2019 and Feb 19, 2026 — classified as 1 death, 9 injuries, 51 malfunctions.
The most frequently reported specific device problem is “Break” (43 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break43
- Unintended Movement5
- Fracture4
- Premature Separation3
- Adverse Event Without Identified Device or Use Problem2
- Material Integrity Problem2
- Appropriate Device Problem Term/Code Not Available1
- Crack1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Consequences Or Impact To Patient14
- Insufficient Information9
- No Known Impact Or Consequence To Patient8
- Pain6
- Device Embedded In Tissue or Plaque5
- Fever3
- High White Blood Cell Count3
Recent adverse event reports
The 25 most recent of 61 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2026Fracture
IT WAS REPORTED VIA FDA (B)(4), THERE WERE ALSO ISSUES WITH THE T-FASTENERS, POSSIBLY A DEVICE FAILURE. FASTENERS FRACTURED DURING INITIAL USE DURING PLACEMENT OF G-TUBE "[GASTRIC TUBE]".
Patient problems: Insufficient Information
MAUDE report 24388239 ↗MalfunctionOct 12, 2025Break; Break
IT WAS REPORTED THE "SUTURE BREAKS AT ANCHOR WHEN LOCKED." THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23273728 ↗MalfunctionAug 12, 2025Break; Break
ADDITIONAL INFORMATION RECEIVED 19-AUG-2025 REGARDING THE REPORTED INCIDENT NOTED, "THE SUTURE WAS BROKEN DURING THE PREPARATION STAGE, BEFORE BEING USED ON THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22784624 ↗MalfunctionApr 11, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED "BALLOON LEAK, CAME OUT OF STOMA, LEAK FROM STOMA SITE AFTER 2 DAYS OF USE... RETURN TO SURGERY¿ FOR MIC-KEY REPLACEMENT AND THAT PATIENT WAS CURRENTLY INPATIENT AFTER ¿SECOND SURGERY.... MUM REPORTS TUBE WAS LEAKING - REDUCED FEEDS TO ½, THEN NOTED TUBE HAD FALLEN OUT...PATIENT WAS AN INPATIENT AFTER THE SECOND SURGERY ¿ THE PATIENT WAS ORIGINALLY DISCHARGED FOLLOWING THE FIRST INITIAL SURGERY OF THE GASTROSTOMY INSERTION, HE WAS THEN DISCHARGED AND THEN THE MIC-KEY BUTTON FELL OUT WHILE AT HOME AND, THEREFORE, PRESENTED BACK TO HOSPITAL AND REQUIRED A SECOND SURGERY WITHIN 1 WEEK FOR THE SECOND PLACEMENT OF THE GASTROSTOMY AND REQUIRED ADMISSION ON [THE] SURGICAL WARD FOR POST-OP MANAGEMENT AND CARE FOR A FURTHER 6-DAYS. ADDITIONAL INFORMATION RECEIVED 19-MAR-2025 STATED "IT WAS AN INITIAL TUBE INSERTION WHICH FELL OUT AND REQUIRED SURGERY FOR A SECOND PLACEMENT WITHIN A WEEK." THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21829789 ↗MalfunctionApr 9, 2025Break; Break
IT WAS REPORTED TWO "SUTURES SNAPPED DURING OPERATION. NEW GASTROPEXY PACKAGE WAS OPENED AND REPLACED." THERE WAS NO REPORTED PATIENT INJURY. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21807255 ↗MalfunctionFeb 19, 2025Break; Break
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING TWO DIFFERENT EVENTS. THIS IS THE FIRST OF TWO REPORTS. REFER TO 9611594-2025-00033 FOR THE SECOND EVENT. IT WAS REPORTED THE "SUTURES RUPTURED." THE DEVICE REMAINS IMPLANTED. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21412876 ↗DeathFeb 19, 2025Break; Break
IT WAS REPORTED THE "SUTURES RUPTURED." THE PATIENT DIED FROM PERITONITIS.
Patient problems: Peritonitis
MAUDE report 21412875 ↗MalfunctionFeb 18, 2025Break; Break
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING TWO DIFFERENT EVENTS. THIS IS THE SECOND OF TWO REPORTS. REFER TO 9611594-2025-00030 FOR THE FIRST EVENT. IT WAS REPORTED THE "SUTURES RUPTURED." THERE WAS NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 21401862 ↗MalfunctionNov 27, 2024Break; Break
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT EVENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 9611594-2024-00272 FOR THE FIRST EVENT. REFER TO 9611594-2024-00273 FOR THE SECOND EVENT. IT WAS REPORTED THE SUTURES ARE TOO THIN AND BREAKS WHEN THE SURGEON PRESSED DOWN THE DILATOR TO FORM THE STOMA. THE GASTRO SURGEON STATED AT LEAST 1-2 SUTURES BROKE, WHICH RESULTED IN THE SURGEON HAVING TO USE THE STANDARD PROCEDURE INSTEAD. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20799392 ↗MalfunctionNov 27, 2024Break; Break
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT EVENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 9611594-2024-00273 FOR THE SECOND EVENT. REFER TO 9611594-2024-00274 FOR THE THIRD EVENT. IT WAS REPORTED THE SUTURES ARE TOO THIN AND BREAKS WHEN THE SURGEON PRESSED DOWN THE DILATOR TO FORM THE STOMA. THE GASTRO SURGEON STATED AT LEAST 1-2 SUTURES BROKE, WHICH RESULTED IN THE SURGEON HAVING TO USE THE STANDARD PROCEDURE INSTEAD. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20799389 ↗MalfunctionNov 27, 2024Break; Break
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT EVENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 9611594-2024-00272 FOR THE FIRST EVENT. REFER TO 9611594-2024-00274 FOR THE THIRD EVENT. IT WAS REPORTED THE SUTURES ARE TOO THIN AND BREAKS WHEN THE SURGEON PRESSED DOWN THE DILATOR TO FORM THE STOMA. THE GASTRO SURGEON STATED AT LEAST 1-2 SUTURES BROKE, WHICH RESULTED IN THE SURGEON HAVING TO USE THE STANDARD PROCEDURE INSTEAD. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20799341 ↗MalfunctionNov 21, 2024Break; Break
IT WAS REPORTED TWO OF THE "SAF-T-PEXY T-FASTENERS BROKE DURING A PLACEMENT PROCEDURE." THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED 30-OCT-2024 STATED, "THE SUTURES DID NOT LOOP CORRECTLY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20750424 ↗MalfunctionNov 20, 2024Break; Break
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME (ONE) PATIENT. THIS IS THE THIRD OF FOUR REPORTS. REFER TO 9611594-2024-00260 FOR THE FIRST REPORT REFER TO 9611594-2024-00261 FOR THE SECOND REPORT REFER TO 9611594-2024-00263 FOR THE FOURTH REPORT. IT WAS REPORTED, FOUR T-FASTENERS WERE PLACED. ALL DETACHED AFTER A FEW DAYS. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 20737550 ↗MalfunctionNov 20, 2024Break; Break
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME (ONE) PATIENT. THIS IS THE FOURTH OF FOUR REPORTS. REFER TO 9611594-2024-00260 FOR THE FIRST REPORT. REFER TO 9611594-2024-00261 FOR THE SECOND REPORT. REFER TO 9611594-2024-00262 FOR THE THIRD REPORT. IT WAS REPORTED, FOUR T-FASTENERS WERE PLACED. ALL DETACHED AFTER A FEW DAYS. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 20737468 ↗MalfunctionNov 20, 2024Break; Break
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME (ONE) PATIENT. THIS IS THE SECOND OF FOUR REPORTS. REFER TO 9611594-2024-00260 FOR THE FIRST REPORT. REFER TO 9611594-2024-00262 FOR THE THIRD REPORT. REFER TO 9611594-2024-00263 FOR THE FOURTH REPORT. IT WAS REPORTED, FOUR T-FASTENERS WERE PLACED. ALL DETACHED AFTER A FEW DAYS. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 20737439 ↗MalfunctionNov 20, 2024Break; Break
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME (ONE) PATIENT. THIS IS THE FIRST OF FOUR REPORTS. REFER TO 9611594-2024-00261 FOR THE SECOND REPORT. REFER TO 9611594-2024-00262 FOR THE THIRD REPORT. REFER TO 9611594-2024-00263 FOR THE FOURTH REPORT. IT WAS REPORTED, FOUR T-FASTENERS WERE PLACED. ALL DETACHED AFTER A FEW DAYS. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 20737058 ↗MalfunctionNov 12, 2024Break; Break
IT WAS REPORTED DURING USE "THE SUTURE STRINGS BROKE ON THE ANCHORS." THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20665589 ↗MalfunctionOct 17, 2024Crack; Crack
IT WAS REPORTED THREE GASTROPEXY ANCHORS HAD CRACKS NOTED DURING PROCEDURE. ADDITIONAL INFORMATION RECEIVED 26-SEP-2024 NOTED "THE KIT THAT MALFUNCTIONED WAS USED AND WE PLACED ONE OF THE ANCHOR SET."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20474826 ↗InjurySep 24, 2024Unintended Movement; Premature Separation; Unintended Movement; Premature Separation
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE THIRD OF THREE REPORTS. REFER TO 9611594-2024-00199 FOR THE FIRST REPORT. REFER TO 9611594-2024-00200 FOR THE SECOND REPORT. IT WAS REPORTED, THE METAL BAR OF THE T-FASTENERS FELL OFF WITHIN 24 HOURS. PER ADDITIONAL INFORMATION RECEIVED ON 08SEP2024, PATIENT REQUIRED TWO ADDITIONAL VISITS TO THE INTERVENTIONAL RADIOLOGY (IR) SUITE: CONTRAST STUDY VIA RIG ON WAS PERFORMED ON (B)(6) 2024. GASTROPEXY WAS REATTEMPTED ON (B)(6) 2024. PATIENT REQUIRED ADDITIONAL HOSPITAL STAY AND A COURSE OF INTRAVENOUS ANTIBIOTICS. IT WAS ADDITIONALLY REPORTED, ¿26AUG2024 PROCEEDED TO HAVE A RADIOLOGICALLY INSERTED GASTROSTOMY (RIG) PROCEDURE. POST OP PAIN IN LEFT SHOULDER SECONDARY TO PNEUMOPERITONEUM, RESOLVED OVER NEXT FEW DAYS. POST-OP DAY 1 NOTED RIG LOOSE DISCS AND COMING AWAY FROM RIG - TAKEN BACK TO DSA FOR REVIEW AND FIXING, KEPT ON NASOGASTRIC TUBE (NGT) FEEDS ONLY. POST-OP DAY 2 ABDOMINAL PAIN, TAKEN BACK TO DSA FOR PARACENTESIS OF PNEUMOPERITONEUM AND REATTEMPT AT GASTROPEXY - FAILED TO PERFORM GASTROPEXY WITH ANCHORS, OPT
Patient problems: Device Embedded In Tissue or Plaque; Fever; High White Blood Cell Count; Pain; Pneumoperitoneum; Skin Inflammation/ Irritation; Tachycardia
MAUDE report 20294590 ↗InjurySep 24, 2024Unintended Movement; Premature Separation; Unintended Movement; Premature Separation
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE SECOND OF THREE REPORTS. REFER TO 9611594-2024-00199 FOR THE FIRST REPORT. REFER TO 9611594-2024-00201 FOR THE THIRD REPORT. IT WAS REPORTED, THE METAL BAR OF THE T-FASTENERS FELL OFF WITHIN 24 HOURS. PER ADDITIONAL INFORMATION RECEIVED ON 08SEP2024, PATIENT REQUIRED TWO ADDITIONAL VISITS TO THE (B)(6): CONTRAST STUDY VIA RIG ON WAS PERFORMED ON (B)(6) 2024. GASTROPEXY WAS REATTEMPTED ON (B)(6) 2024. PATIENT REQUIRED ADDITIONAL HOSPITAL STAY AND A COURSE OF INTRAVENOUS ANTIBIOTICS. IT WAS ADDITIONALLY REPORTED, ¿(B)(6) 2024 PROCEEDED TO HAVE A RADIOLOGICALLY INSERTED GASTROSTOMY (RIG) PROCEDURE. POST OP PAIN IN LEFT SHOULDER SECONDARY TO PNEUMOPERITONEUM, RESOLVED OVER NEXT FEW DAYS. POST-OP DAY 1 NOTED RIG LOOSE DISCS AND COMING AWAY FROM RIG - TAKEN BACK TO DSA FOR REVIEW AND FIXING, KEPT ON NASOGASTRIC TUBE (NGT) FEEDS ONLY. POST-OP DAY 2 ABDOMINAL PAIN, TAKEN BACK TO DSA FOR PARACENTESIS OF PNEUMOPERITONEUM AND REATTEMPT AT GASTROPEXY - FAILED TO PERFORM GASTROPEXY WITH ANCHORS, OPTED TO RELY ON TENSION FROM
Patient problems: Device Embedded In Tissue or Plaque; Fever; High White Blood Cell Count; Pain; Pneumoperitoneum; Skin Inflammation/ Irritation; Tachycardia
MAUDE report 20290098 ↗InjurySep 24, 2024Unintended Movement; Premature Separation; Unintended Movement; Premature Separation
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF THREE REPORTS. REFER TO 9611594-2024-00200 FOR THE SECOND REPORT. REFER TO 9611594-2024-00201 FOR THE THIRD REPORT. IT WAS REPORTED, THE METAL BAR OF THE T-FASTENERS FELL OFF WITHIN 24 HOURS. PER ADDITIONAL INFORMATION RECEIVED ON 08SEP2024, PATIENT REQUIRED TWO ADDITIONAL VISITS TO THE INTERVENTIONAL RADIOLOGY (IR) SUITE: CONTRAST STUDY VIA RIG ON WAS PERFORMED ON (B)(6) 2024. GASTROPEXY WAS REATTEMPTED ON (B)(6) 2024. PATIENT REQUIRED ADDITIONAL HOSPITAL STAY AND A COURSE OF INTRAVENOUS ANTIBIOTICS. IT WAS ADDITIONALLY REPORTED, ¿(B)(6) 2024 PROCEEDED TO HAVE A RADIOLOGICALLY INSERTED GASTROSTOMY (RIG) PROCEDURE. POST OP PAIN IN LEFT SHOULDER SECONDARY TO PNEUMOPERITONEUM, RESOLVED OVER NEXT FEW DAYS. POST-OP DAY 1 NOTED RIG LOOSE DISCS AND COMING AWAY FROM RIG - TAKEN BACK TO DSA FOR REVIEW AND FIXING, KEPT ON NASOGASTRIC TUBE (NGT) FEEDS ONLY. POST-OP DAY 2 ABDOMINAL PAIN, TAKEN BACK TO DSA FOR PARACENTESIS OF PNEUMOPERITONEUM AND REATTEMPT AT GASTROPEXY - FAILED TO PERFORM GASTROPEXY WITH ANCHORS, O
Patient problems: Device Embedded In Tissue or Plaque; Fever; High White Blood Cell Count; Pain; Pneumoperitoneum; Skin Inflammation/ Irritation; Tachycardia
MAUDE report 20290070 ↗MalfunctionSep 5, 2024Break; Break
IT WAS REPORTED "THE CONCERNS EXPRESSED BY THE SURGEONS ARE THE T-FASTENERS BREAK VERY EASILY AND THEN THE TUBE IS DISLODGED INTO THE ABDOMINAL CAVITY AND NO LONGER ADHERED TO THE STOMACH. THIS HAS CAUSED INTRAPERITONEAL FLUID TO LEAK INTO THE ABDOMINAL CAVITY." NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20149695 ↗MalfunctionJun 7, 2024Break; Break
IT WAS REPORTED, "THE ANCHOR BREAKS WHILE IN/ON PATIENT. ANCHOR COMES OUT OF PATIENT." NO INJURY REPORTED. PER ADDITIONAL INFORMATION RECEIVED ON 1-MAY-2024, "ANCHOR BREAKS DURING PROCEDURE. IN THIS CASE ANCHOR CAME OUT."
Patient problems: Insufficient Information
MAUDE report 19490384 ↗MalfunctionJun 6, 2024Break; Break
IT WAS REPORTED, "THE ANCHOR BREAKS WHILE IN/ON PATIENT. ANCHOR COMES OUT OF PATIENT". NO INJURY REPORTED. PER ADDITIONAL INFORMATION RECEIVED ON 1-MAY-2024, "ANCHOR BREAKS DURING PROCEDURE. IN THIS CASE ANCHOR CAME OUT".
Patient problems: Insufficient Information
MAUDE report 19479701 ↗MalfunctionMay 23, 2024Break; Break
IT WAS REPORTED, ¿THE ANCHOR BREAKS WHILE IN/ON PATIENT. ANCHOR COMES OUT OF PATIENT.¿ NO INJURY REPORTED. PER ADDITIONAL INFORMATION RECEIVED ON 1-MAY-2024, ¿ANCHOR BREAKS DURING PROCEDURE. IN THIS CASE ANCHOR CAME OUT.¿
Patient problems: Insufficient Information
MAUDE report 19383084 ↗
Other TUBE, GASTRO-ENTEROSTOMY devices
- AVANOS 98431Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY93 reports
- AVANOS 98432Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY72 reports
- Halyard 98701Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death40 reports
- Halyard 98431Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY35 reports
- AVANOS 98433Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY34 reports
- AVANOS 98430Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY1 death29 reports
- Halyard 98433Avanos Medical, Inc. · TUBE, GASTRO-ENTEROSTOMY27 reports
- Bard Button Replacement Gastrostomy Device 000282Bard Access Systems, Inc. · TUBE, GASTRO-ENTEROSTOMY25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.