Portex 100/399/218Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Portex (409)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00351688068142), received between Dec 11, 2018 and Oct 12, 2022 — classified as 4 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component8
- Premature Separation4
- Device Contamination with Body Fluid1
- Difficult to Open or Close1
- Disconnection1
- Fracture1
Patient problems
- Awareness During Anesthesia4
- Needle Stick/Puncture4
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 12, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE PLASTIC CATHETER SECUREMENT DEVICE DISLODGED FROM THE SOFT DRESSING . THE RUPTURE OF THE DEVICE LED TO AN ACCIDENTAL REMOVAL OF THE CATHETER. A SECOND EPIDURAL CATHETER HAD TO BE INSERTED. NO FURTHER INCIDENT OBSERVED AFTERWARD. INCIDENT REPORTEDLY OCCURS AT A FREQUENCY RANGE OF 1 TO 4 TIMES PER MONTH. ADDITIONAL INFORMATION RECEIVED ON 29-SEP-2022: THE DEVICE WAS CHANGED. NO FURTHER ISSUE OBSERVED.
Patient problems: Needle Stick/Puncture
MAUDE report 15587370 ↗MalfunctionOct 11, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE PLASTIC CATHETER SECUREMENT DEVICE DISLODGED FROM THE SOFT DRESSING. THE RUPTURE OF THE DEVICE LED TO AN ACCIDENTAL REMOVAL OF THE CATHETER. A SECOND EPIDURAL CATHETER HAD TO BE INSERTED. NO FURTHER INCIDENT OBSERVED AFTERWARD. INCIDENT FREQUENCY REPORTEDLY RANGES FROM 1 TO 4 TIMES PER MONTH. ADDITIONAL INFORMATION RECEIVED ON 29-SEP-2022: DEVICE WAS CHANGED. NO FURTHER ISSUE OBSERVED.
Patient problems: Needle Stick/Puncture
MAUDE report 15581994 ↗MalfunctionOct 11, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE PLASTIC CATHETER SECUREMENT DEVICE DISLODGED FROM THE SOFT DRESSING. THE RUPTURE OF THE DEVICE LED TO AN ACCIDENTAL REMOVAL OF THE CATHETER. A SECOND EPIDURAL CATHETER HAD TO BE INSERTED. NO FURTHER INCIDENT OBSERVED AFTERWARD. INCIDENT FREQUENCY RANGE IS REPORTEDLY 1 TO 4 TIMES PER MONTH. ADDITIONAL INFORMATION RECEIVED ON 29-SEP-2022: DEVICE WAS CHANGED. NO FURTHER ISSUE OBSERVED.
Patient problems: Needle Stick/Puncture
MAUDE report 15580851 ↗MalfunctionOct 11, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE PLASTIC CATHETER SECUREMENT DEVICE DISLODGED FROM THE SOFT DRESSING. THE RUPTURE OF THE DEVICE LED TO AN ACCIDENTAL REMOVAL OF THE CATHETER. A SECOND EPIDURAL CATHETER HAD TO BE INSERTED. NO FURTHER INCIDENT OBSERVED AFTERWARD. ADDITIONAL INFORMATION RECEIVED ON 29-SEP-2022: DEVICE WAS CHANGED. NO FURTHER ISSUE OBSERVED.
Patient problems: Needle Stick/Puncture
MAUDE report 15580840 ↗MalfunctionAug 29, 2022Disconnection; Disconnection
IT WAS REPORTED THAT THE CATHETER FELL OUT. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15316107 ↗MalfunctionAug 26, 2022Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING THE USE OF THE PRODUCT, THE CUSTOMER NOTICED THE LID WAS UNABLE TO BE CLOSED, WHICH DID NOT ALLOW HIM TO USE IT PROPERLY. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15299984 ↗MalfunctionJul 12, 2022Fracture; Device Contamination with Body Fluid; Fracture; Device Contamination with Body Fluid
IT WAS REPORTED THAT DURING THE USE OF THE PRODUCT, THE CUSTOMER NOTICED IT WAS BROKEN. A BLOOD MARK WAS FOUND ON IT. NO PATIENT INJURY OR ADVERSE AFFECTS. NO ADDITIONAL INFORMATION IS AVAILABLE FOR THIS COMPLAINT AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14986294 ↗InjuryDec 11, 2018Detachment of Device or Device Component; Premature Separation; Detachment of Device or Device Component; Premature Separation
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PORTEX LOCKIT PLUS REGIONAL ANESTHESIA CATHETER SECUREMENT DEVICE BECAME DETACHED WHILE IN USE WITH A PATIENT. THE REPORTER STATED THE DETACHMENT RESULTED IN THE CATHETER PULLING OUT OF THE EPIDURAL SPACE, WHICH CAUSED THE EPIDURAL PAIN CONTROL TO BE INSUFFICIENT. SUBSEQUENTLY, THE NURSES HAD TO INSERT A NEW EPIDURAL CATHETER AND A MORE INVASIVE METHOD OF ANESTHESIA INTERVENTION WAS REQUIRED, NOTED AS EITHER, "GENERAL ANESTHESIA OR EMERGENCY CESAREAN SECTION". NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Awareness During Anesthesia
MAUDE report 8152168 ↗InjuryDec 11, 2018Detachment of Device or Device Component; Premature Separation; Detachment of Device or Device Component; Premature Separation
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PORTEX LOCKIT PLUS REGIONAL ANESTHESIA CATHETER SECUREMENT DEVICE BECAME DETACHED WHILE IN USE WITH A PATIENT. THE REPORTER STATED THE DETACHMENT RESULTED IN THE CATHETER PULLING OUT OF THE EPIDURAL SPACE, WHICH CAUSED THE EPIDURAL PAIN CONTROL TO BE INSUFFICIENT. SUBSEQUENTLY, THE NURSES HAD TO INSERT A NEW EPIDURAL CATHETER AND A MORE INVASIVE METHOD OF ANESTHESIA INTERVENTION WAS REQUIRED, NOTED AS EITHER, "GENERAL ANESTHESIA OR EMERGENCY CESAREAN SECTION". NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Awareness During Anesthesia
MAUDE report 8152166 ↗InjuryDec 11, 2018Detachment of Device or Device Component; Premature Separation; Detachment of Device or Device Component; Premature Separation
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PORTEX LOCKIT PLUS REGIONAL ANESTHESIA CATHETER SECUREMENT DEVICE BECAME DETACHED WHILE IN USE WITH A PATIENT. THE REPORTER STATED THE DETACHMENT RESULTED IN THE CATHETER PULLING OUT OF THE EPIDURAL SPACE, WHICH CAUSED THE EPIDURAL PAIN CONTROL TO BE INSUFFICIENT. SUBSEQUENTLY, THE NURSES HAD TO INSERT A NEW EPIDURAL CATHETER AND A MORE INVASIVE METHOD OF ANESTHESIA INTERVENTION WAS REQUIRED, NOTED AS EITHER, "GENERAL ANESTHESIA OR EMERGENCY CESAREAN SECTION". NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Awareness During Anesthesia
MAUDE report 8151930 ↗InjuryDec 11, 2018Detachment of Device or Device Component; Premature Separation; Detachment of Device or Device Component; Premature Separation
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL PORTEX LOCKIT PLUS REGIONAL ANESTHESIA CATHETER SECUREMENT DEVICE BECAME DETACHED WHILE IN USE WITH A PATIENT. THE REPORTER STATED THE DETACHMENT RESULTED IN THE CATHETER PULLING OUT OF THE EPIDURAL SPACE, WHICH CAUSED THE EPIDURAL PAIN CONTROL TO BE INSUFFICIENT. SUBSEQUENTLY, THE NURSES HAD TO INSERT A NEW EPIDURAL CATHETER AND A MORE INVASIVE METHOD OF ANESTHESIA INTERVENTION WAS REQUIRED, NOTED AS EITHER, "GENERAL ANESTHESIA OR EMERGENCY CESAREAN SECTION". NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Awareness During Anesthesia
MAUDE report 8151924 ↗
Other CATHETER, CONDUCTION, ANESTHETIC devices
- ARROW IPN925037Teleflex Incorporated · CATHETER, CONDUCTION, ANESTHETIC7 recalls0 reports
- ARROW IPN924999Teleflex Incorporated · CATHETER, CONDUCTION, ANESTHETIC3 recalls0 reports
- ARROW IPN924964Teleflex Incorporated · CATHETER, CONDUCTION, ANESTHETIC2 recalls1 report
- ARROW IPN925052Teleflex Incorporated · CATHETER, CONDUCTION, ANESTHETIC2 recalls0 reports
- ARROW IPN924975Teleflex Incorporated · CATHETER, CONDUCTION, ANESTHETIC2 recalls0 reports
- ARROW IPN922619Teleflex Incorporated · CATHETER, CONDUCTION, ANESTHETIC2 recalls0 reports
- ARROW IPN925020Teleflex Incorporated · CATHETER, CONDUCTION, ANESTHETIC2 recalls0 reports
- ARROW IPN927743Teleflex Incorporated · CATHETER, CONDUCTION, ANESTHETIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page