Skip to content

NGENUITY 00380650003997Recalls & reported problems

Alcon Laboratories, Inc. · Microsurgical Camera · Class 1

All models of Ngenuity (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. May 23, 2024Open, Classified

    After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display incorrectly, which could result in incorrect placement of toric IOL axis, which could cause astigmatic error under/over correction resulting in decreased uncorrected visual acuity.

    Z-2412-2024 · Root cause: Software design · Recall details · FDA record ↗