Skip to content

ACRYSOF® SN60AT (REF SN60AT.190)Recalls & reported problems

Alcon Laboratories, Inc. · AcrySof(R) Natural SINGLE-PIECE ACRYLICFOLDABLE IOL, UV WITH BLUE LIGHT FILTER,POSTERIOR CHAMBER LENS (IOL/PC), 13.0mm LENGTH,6.0mm BICONVEX OPTIC, PLANAR HAPTICS. · Class 3 · Implant

All models of ACRYSOF (353)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380650130280), received between Oct 18, 2021 and Apr 28, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Opacification” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Material Opacification1

Patient problems

  • Blurred Vision1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 28, 2025Material Opacification; Material Opacification

    A PHYSICIAN REPORTED WITH A DESCRIPTION OF THE INTRAOCULAR LENS HAD SUPERFICIAL COATING OR FILM. THE LENS REMAINS IMPLANTED. IT WAS BARELY VISIBLE DURING THE VISUAL INSPECTION BEFORE. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 21918025 ↗
  • InjuryOct 18, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    BASED ON INFORMATION RECEIVED FOLLOWING SUBMISSION OF THE INITIAL REPORT, THIS EVENT DOES NOT MEET CRITERIA FOR REPORTING AS A SERIOUS INJURY.

    Patient problems: Blurred Vision

    MAUDE report 12646245 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page