Skip to content

ACRYSOF® SN6AT5 (REF SN6AT5.195)Recalls & reported problems

Alcon Laboratories, Inc. · AcrySof(R) IQ TORIC ASTIGMATISM IOL, SINGLE-PIECE ACRYLIC FOLDABLE LENS, UV WITH BLUE LIGHTFILTER, 13.0mm LENGTH, 6.0mm BICONVEX TORICASPHERIC OPTIC, 3.00 D CYLINDER. · Class 3 · Implant

All models of ACRYSOF (353)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380652250986), received between May 31, 2022 and Jan 30, 2025 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Scratched Material” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Scratched Material2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Damaged by Another Device1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Visual Disturbances1
  • Zonular Dehiscence1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION HAS BEEN RECEIVED AND STATED THE PATIENT HAD ZONULYSIS HAPPENED WHEN DIALING THE LENS TO THE FINAL AXIS.

    Patient problems: Zonular Dehiscence

    MAUDE report 21273049 ↗
  • InjuryDec 30, 2022Scratched Material; Scratched Material

    A HEALTH CARE PROFESSIONAL REPORTED THAT FOLLOWING THE IMPLANT PROCEDURE, THE PATIENT COMPLAINT ABOUT A SCRATCH ON THE LENS. THE SCRATCH WAS INFLUENCING HIS VISION. THE LENS WAS PLANNED TO REMOVE. ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT, POST CATARACT SURGERY (WITH IOL IMPLANT) 1 WEEK FOLLOW UP, IT WAS FOUND THAT THE LENS THAT WAS INSERTED PRESENTED WITH A MARK IN THE OPTIC OF THE LENS, AFFECTING THE PATIENTS VISION. THE IOL WAS EXCHANGED IN A SECONDARY PROCEDURE. THE ISSUE HAS BEEN RESOLVED.

    Patient problems: Visual Disturbances

    MAUDE report 16077323 ↗
  • MalfunctionOct 3, 2022Scratched Material; Scratched Material

    A PHYSICIAN REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE SCRATCHES WERE NOTED CENTRALLY ON THE INTRAOCULAR LENS, THEREFORE THE LENS WAS EXPLANTED AND DISCARDED. SUBSEQUENTLY, THE STANDBY IOL WAS IMPLANTED. THERE WAS A PATIENT CONTACT BUT NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15527962 ↗
  • MalfunctionMay 31, 2022Device Damaged by Another Device; Device Damaged by Another Device

    A PURCHASING AGENT REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, LENS SCRATCHED WHILE ADVANCING THE INJECTOR. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 14548856 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates