Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380652251136), received between Oct 18, 2024 and Oct 18, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Optical Decentration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Optical Decentration1

Patient problems

  • Visual Impairment1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 18, 2024Optical Decentration; Optical Decentration

    AN OTHER HEALTH CARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE IN RIGHT EYE, 3 WEEKS POSTOPERATIVELY THE PATIENT REFRACTED AT -3.00/-0.5X1 AND REPORTED REFRACTIVE SURPRISE AS TARGET WAS FOR EMMETROPIA. THE PATIENT WAS COMPLICATED WITH PRE-OPERATIVE HISTORY OF MILD ALBINISM AND NYSTAGMUS. SURGEON HAS REVIEWED THE PATIENT POST-OPERATIVELY AND CONFIDENT THAT THERE WAS NO RETAINED VISCOELASTIC. HOWEVER, IOL WAS SITTING MORE ANTERIOR THAN EXPECTED. SURGEON FELT IT WAS A PATIENT FACTOR RATHER THAN ISSUE WITH THE IOL. CONSERVATIVE MANAGEMENT WAS PLANNED. THE SURGEON WAS MONITORING THE PATIENT HAD A MYOPIC REFRACTIVE ERROR BUT OTHERWISE THE EYE WAS QUIET, AND PATIENT WAS PROCEEDING WITH SURGERY ON THE FELLOW EYE IN THE COMING DAYS. THERE WAS NO SECONDARY INTERVENTION PLANNED AND THE SURGEON WILL PROVIDE AN UPDATE IF FURTHER INFORMATION COMES FORWARD. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Visual Impairment

    MAUDE report 20481147 ↗