Skip to content

Clareon SY60WF (UDI 00380652251228)Recalls & reported problems

Alcon Laboratories, Inc. · STERILE UV and Blue Light Filtering Acrylic Foldable Aspheric Posterior Chamber Intraocular Lens · Class 3 · Implant

All models of Clareon (1142)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380652251228), received between Aug 28, 2024 and Sep 27, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Positioning Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure2
  • Break1
  • Material Twisted/Bent1

Patient problems

  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 27, 2024Positioning Failure; Material Twisted/Bent; Positioning Failure; Material Twisted/Bent

    A PHYSICIAN REPORTED THAT DURING CATARACT SURGERY WITH INTRAOCULAR LENS (IOL) IMPLANTATION, WHILE IMPLANTING THE IOL INTO THE PATIENT EYE, THE IOL WAS EJECTED INVERTED. THEY TRIED TO FLIP IT BACK OVER, BUT COULD NOT, SO REMOVED THE IOL AND REPLACED WITH A COMPANY LENS IN AN INITIAL PROCEDURE. FURTHERMORE, A HAPTIC WAS TORN OFF WHILE THE IOL WAS BEING REMOVED. ADDITIONAL INCISION WAS CRATED WHEN TRIED TO FLIP THE IOL BACK OVER. THE INCISION WAS ENLARGED FOR THE REMOVAL OF THE IOL. SUBCONJUNCTIVAL INJECTION WAS APPLIED IN THE END. AFTER REVIEW OF SURGICAL VIDEO PROVIDED BY THE CUSTOMER, IT WAS NOTICED THAT THE SURGERY INVOLVED THREE LENSES. WITH THE FIRST IOL, THE HAPTIC BROKE DURING DELIVERY DUE TO A PLUNGER UNDERRIDE/EXTENDED HAPTIC. THE SECOND LENS WAS DELIVERED UPSIDE DOWN DUE TO A STRAIGHT LEADING HAPTIC AND NON ROTATION OF THE CARTRIDGE. THE THIRD LENS, A MULTI PIECE LENS REMAINS IMPLANTED. THE FIRST IOL UNFOLDED POSTERIOR SURFACE UP DUE TO THE TRAPPED TRAILING PORTION OF THE OPTIC AND HAPTIC, WHICH WERE STILL IN THE CARTRIDGE. WHEN THE PLUNGER/CARTRIDGE WERE RETRACTED, THIS BROKE THE TRAILING HAPTIC. THE LENS WAS PULLED INTACT OUT OF THE INCISION WITH FORCEPS. INCISION WAS NOT

    Patient problems: Insufficient Information

    MAUDE report 20320278 ↗
  • MalfunctionAug 28, 2024Break; Positioning Failure; Break; Positioning Failure

    A PHYSICIAN REPORTED THAT DURING CATARACT SURGERY WITH INTRAOCULAR LENS (IOL) IMPLANTATION, WHILE IMPLANTING THE IOL INTO THE PATIENT'S EYE, THE IOL WAS EJECTED INVERTED. THEY TRIED TO FLIP IT BACK OVER, BUT COULDN'T, SO REMOVED THE IOL AND REPLACED WITH A COMPANY LENS IN AN INITIAL PROCEDURE. FURTHERMORE, A HAPTIC WAS TORN OFF WHILE THE IOL WAS BEING REMOVED. ADDITIONAL INCISION WAS CRATED WHEN TRIED TO FLIP THE IOL BACK OVER. THE INCISION WAS ENLARGED FOR THE REMOVAL OF THE IOL. SUBCONJUNCTIVAL INJECTION WAS APPLIED IN THE END.

    Patient problems: Insufficient Information

    MAUDE report 20083959 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page