Clareon SY60WF (UDI 00380652251426)Recalls & reported problems

Alcon Laboratories, Inc. · STERILE UV and Blue Light Filtering Acrylic Foldable Aspheric Posterior Chamber Intraocular Lens · Class 3 · Implant

All models of Clareon (1142)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00380652251426), received between May 24, 2022 and Sep 25, 2026 — classified as 2 injuries, 16 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Scratched Material5
  • Defective Device2
  • Material Split, Cut or Torn2
  • Crack1
  • Device Damaged by Another Device1
  • Device Dislodged or Dislocated1
  • Failure to Deliver1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Insufficient Information9
  • Eye Injury1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Scratched Material

    A health care professional reported that, during intraocular lens implantation surgery, it was found scratch on the optic under microscope and there was no patient harm.Additional information has been requested but is not available at the time of this report.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50893365 ↗
  • MalfunctionAug 24, 2026Material Split, Cut or Torn

    A non-healthcare professional reported that during an intraocular lens (IOL) implant procedure, lens was cracked. Additional information has been requested and received by stating there was no patient contact and harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50408591 ↗
  • InjuryJul 8, 2026Scratched Material

    A NON HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, LOADING ERROR WAS NOTICED. ADDITIONAL INFORMATION HAS BEEN RECEIVED THAT THE LENS WAS SCRATCHED. THE LENS WAS REMOVED WITH LENS SCISSORS. THE INCISION WAS WIDENED AND 10 NYLON WAS USED TO CLOSE THE INCISION. THERE WAS NO FURTHER PATIENT IMPACT.

    Patient problems: Eye Injury

    MAUDE report 25744553 ↗
  • MalfunctionMay 22, 2026Failure to Deliver

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE THE LENS FOUND TO BE DAMAGED. ADDITIONAL INFORMATION RECEIVED AND STATED THAT THE LENS WAS NOT DEPLOYING PROPERLY. THE SURGERY WAS COMPLETED ON THE SAME DAY. THE LENS WAS NOT INSERTED IN TO THE PATIENT EYE. THE LENS REMAIN IN THE DELIVERY SYSTEM. PER THE SURGEON¿S OPINION IT WAS EITHER LENS DEFECT OR TECHNICIAN LOADING ISSUE. THERE WAS A NO IMPACT TO THE PATIENT. THE SURGEONS PROGNOSIS WAS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25255970 ↗
  • MalfunctionMay 18, 2026Break; Product Quality Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED DURING INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE LENS WAS DEFECTIVE. THERE WAS A PATIENT CONTACT, AND THE LENS WAS NOT USED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25203160 ↗
  • MalfunctionDec 30, 2025Device Damaged by Another Device; Device Damaged by Another Device

    A PHYSICIAN REPORTED THAT DURING CATARACT SURGERY WITH AN INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE IT WAS OBSERVED THAT THE INJECTOR PLUNGER COULD NOT ADVANCE THE FOLDED IOL THROUGH THE NARROW SECTION OF THE CARTRIDGE. THIS LED TO A CARTRIDGE CRACK AND SUBSEQUENT DAMAGE TO THE IOL. ON THE SAME DAY THE PATIENT WAS IMPLANTED WITH ANOTHER IDENTICAL IOL. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT, THE SURGEON ASSUMES THAT SOMETHING WAS WRONG WITH THE LENS, PREVIOUSLY THERE WERE NO ISSUES.

    Patient problems: Insufficient Information

    MAUDE report 23922799 ↗
  • MalfunctionSep 18, 2025Break; Break

    A NON-HEALTHCARE PROFESSIONAL REPORTED A TORN OR BROKEN HAPTIC. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23082909 ↗
  • MalfunctionJul 11, 2025Break; Break

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, TRAILING HAPTIC BROKE AND THE LENS HAD TO BE CUT OUT OF PATIENT'S EYE IN ORDER TO REPLACE. THE PROCEDURE WAS COMPLETED ON SAME DAY.

    Patient problems: Insufficient Information

    MAUDE report 22487839 ↗
  • MalfunctionJan 27, 2025Break; Break

    A FACILITY REPRESENTATIVE REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, HAPTIC WAS BROKEN BEFORE USE. ADDITIONAL INFORMATION RECEIVED STATING THAT HAPTIC BROKEN WHILE LOADING THE LENS. THE PROCEDURE COMPLETED THE SAME DAY WITH NEW LENS. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21240714 ↗
  • MalfunctionSep 26, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    A NON-HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, LENS TORE UPON IMPLANTATION. THE LENS WAS REMOVED AND ANOTHER LENS WAS IMPLANTED. ADDITIONAL INFORMATION WAS RECEIVED, THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20308364 ↗
  • MalfunctionSep 5, 2024Break; Scratched Material; Break; Scratched Material

    A FACILITY REPRESENTATIVE REPORTED THAT DURING THE INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, LENS WAS INDENTED AND TORN. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 20146569 ↗
  • MalfunctionMar 19, 2024Crack; Crack

    A CONSUMER REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, ONCE LENS IN THE EYE, UNDER THE MICROSCOPE SURGEON COULD SEE SEVERAL LITTLE CRACKS. HE HAD TO CUT OUT AND REMOVE THE IMPLANT AND INSERT A NEW ONE WITH NO PATIENT HARM. THE IOL WAS EXPLANTED DURING INITIAL PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18929712 ↗
  • MalfunctionJan 5, 2024Scratched Material; Scratched Material

    A NURSE REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, LENS DEFECTIVE NOTED UPON IMPLANTATION. LENS WAS EXPLANTED AT INITIAL PROCEDURE AND REPLACED WITH UNSPECIFIED LENS. ADDITIONAL INFORMATION HAS BEEN RECEIVED AND STATED THAT, THE LENS HAD SCRATCHES. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18452139 ↗
  • MalfunctionNov 23, 2023Break; Break

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE LENS FOUND TO BE BROKEN. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 18198116 ↗
  • InjuryJul 25, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A REGISTERED NURSE REPORTED ABOUT A LENS BEING EXPLANTED IN A SECONDARY PROCEDURE DUE TO LENS WHEN CHECKED POST OPERATIVELY FOUND TO BE DECENTERED. THE LENS WAS REPLACED WITH NON COMPANY LENS.

    Patient problems: Insufficient Information

    MAUDE report 17390313 ↗
  • MalfunctionApr 25, 2023Defective Device; Defective Device

    A NON HEALTH CARE PROFESSIONAL (TEAM LEADER) REPORTED THAT DURING THE INTRAOCULAR LENS (IOL), IMPLANT PROCEDURE LENS FOUND BE A DAMAGED. THERE ARE TWELVE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS REPORT. THIS IS 3 OF 12. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 16807518 ↗
  • MalfunctionApr 25, 2023Defective Device; Defective Device

    A NON HEALTH CARE PROFESSIONAL (TEAM LEADER) REPORTED THAT DURING THE INTRAOCULAR LENS (IOL), LENS FOUND BE A DAMAGED. THERE ARE TWELVE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS REPORT. THIS IS 2 OF 12.

    Patient problems: Insufficient Information

    MAUDE report 16807506 ↗
  • MalfunctionMay 24, 2022Scratched Material; Scratched Material

    AN OPHTHALMOLOGIST REPORTED THAT, FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, OPTIC OF THE IOL FOUND TO BE SCRATCHED. THE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. ADDITIONAL INFORMATION WAS RECEIVED THAT, AFTER INSERTION OF THE IOL, A VERTICAL WHITE OPACITY WAS OBSERVED ON THE POSTERIOR SURFACE OF THE IOL. THE IOL WAS REMOVED IN THREE SECTIONS AND A SPARE IOL WAS INSERTED WITH THE SAME INJECTOR AND A DIFFERENT CARTRIDGE. THE PATIENT'S POSTOPERATIVE COURSE WAS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14485657 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page