Clareon SY60WF (UDI 00380652251624)Recalls & reported problems
Alcon Laboratories, Inc. · STERILE UV and Blue Light Filtering Acrylic Foldable Aspheric Posterior Chamber Intraocular Lens · Class 3 · Implant
All models of Clareon (1142)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380652251624), received between Jul 20, 2023 and Jan 7, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Crack” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Crack1
- Failure to Deliver1
- Material Discolored1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 7, 2026Failure to Deliver
A NON-HEALTHCARE PROFESSIONAL REPORTED AN ISSUE DURING INTRAOCULAR LENS (IOL) IMPLANTATION. THE INITIAL ATTEMPT INVOLVED USING THE COMPANY¿S CARTRIDGE AND INSERTER; HOWEVER, THE IOL DID NOT ADVANCE INTO THE PROPER POSITION. A DIFFERENT INSERTER WAS THEN UTILIZED, BUT THAT DIDN'T WORK EITHER. THEY THEN REPLACED THE LENS AND CARTRIDGE WHILE CONTINUING TO USE THE SAME COMPANY INSERTER AND VISCOELASTIC. FOLLOWING THE SAME STEPS, THE NEW IOL WAS SUCCESSFULLY IMPLANTED. THE ONLY CHANGE BETWEEN THE UNSUCCESSFUL AND SUCCESSFUL ATTEMPTS WAS THE USE OF A NEW CARTRIDGE AND NEW IOL. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 23990557 ↗MalfunctionApr 29, 2025Material Discolored; Material Discolored
A PHYSICIAN REPORTED THAT AFTER IMPLANTATION OF LENS DURING THE INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE IOL LOOKED BLUE IMMEDIATELY. THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. IT WAS CURRENTLY UNKNOWN WHETHER THESE FINDINGS WERE DISAPPEARED OR CONTINUING. THERE HAVE BEEN NO REPORTED HEALTH PROBLEMS OR HARM TO THE PATIENT. THE LENS REMAIN IMPLANTED. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE 224 MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS IS 57 OF 224.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21925436 ↗MalfunctionJul 20, 2023Crack; Crack
A PHYSICIAN REPORTED DURING THE INTRAOCULAR LENS IMPLANTATION A MICROFRACTURE WAS VISIBLE UNDER A MICROSCOPE ON THE SURFACE OF THE IMPLANT AFTER INJECTION. THERE WAS NO IMPACT TO THE PATIENT. THE LENS REMAINS IMPLANTED IN THE EYE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17358639 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page