Skip to content

ACRYSOF® SA60WF (REF SA60WF.260)Recalls & reported problems

Alcon Laboratories, Inc. · AcrySof(R) ASPHERIC IOL, SINGLE-PIECE ACRYLICFOLDABLE POSTERIOR CHAMBER LENS,UV-ABSORBING, 13.0mm LENGTH, 6.0mm ANTERIORASYMMETRIC BICONVEX OPTIC, PLANAR HAPTICS. · Class 3 · Implant

All models of ACRYSOF (353)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380652332781), received between Oct 20, 2021 and Jan 30, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Failure to Unfold or Unwrap” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Failure to Unfold or Unwrap1
  • Material Twisted/Bent1
  • Scratched Material1

Patient problems

  • Blurred Vision1
  • Eye Injury1
  • No Clinical Signs, Symptoms or Conditions1
  • Visual Disturbances1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 30, 2026Failure to Unfold or Unwrap; Material Twisted/Bent

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, WHEN SURGEON INTENDED TO IMPLANT THE LENS, WHILE GRASPING THE LENS AND PREPARING TO INSERT IT INTO THE CARTRIDGE, SURGEON NOTICED THAT SOMETHING WAS WRONG WITH THE LENS. UPON CLOSER INSPECTION, IT WAS OBSERVED THAT ONE OF THE LENS HAPTICS WAS CAUGHT WITHIN THE OPTICAL PORTION OF THE LENS, CAUSING THE LENS TO TEAR. THE BACKUP LENS WAS USED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24222369 ↗
  • InjuryJan 31, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON HEALTH CARE PROFESSIONAL REPORTED DURING THE INTRAOCULAR LENS IMPLANTATION THE LENS WAS PARTIALLY IN THE EYE AND HAD TO BE REMOVED TO MAKE A LARGER INCISION. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND RECEIVED STATING THE PATIENT WAS NOT HOSPITALIZED, NO INTERVENTION WAS NEEDED FOR THE EVENT AND THERE WAS NO HARM TO THE PATIENT. THE SURGERY WAS COMPLETED ON THE SAME DAY WITH A BACK UP LENS OF THE SAME MODEL AND DIOPTER.

    Patient problems: Eye Injury

    MAUDE report 13391155 ↗
  • InjuryOct 20, 2021Scratched Material; Scratched Material

    A NON HEALTHCARE PROFESSIONAL REPORTED FOLLOWING AN IOL IMPLANT PROCEDURE, THE LENS WAS FOUND OUT TO BE SCRATCHED AT THE ANTERIOR SURFACE. THE PATIENT EXPERIENCE GLARE AND BLURRED VISION, THEREFORE THE LENS WAS EXPLANTED IN A SECONDARY PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Blurred Vision; Visual Disturbances

    MAUDE report 12665815 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page