Skip to content

ACRYSOF ULTRASERT AU00T0 (REF AU00T0.235)Recalls & reported problems

Alcon Laboratories, Inc. · AcrySof(R) IQ ASPHERIC IOL SP ACRYLIC FOLDABLELENS wUltraSert(TM) DELIVERY SYSTEM UV wBLUE LIGHT FILTER. 13.0mm LENGTH 6.0mm ANTERIORASYMMETRIC BICONVEX OPTIC PLANAR HAPTICS. · Class 3 · Implant

All models of AcrySof UltraSert (102)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380652358347), received between Nov 13, 2017 and Nov 13, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Detachment Of Device Component1
  • Sticking1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2017Break; Detachment Of Device Component; Sticking

    HAPTIC BROKE OFF. THESE ARE PRE-LOADED INSERTION DEVICES AND THE HAPTIC GOT STUCK IN DEVICE. A NEW ALCON IOL WITH ULTRASERT PRE-LOADED DELIVERY SYSTEM WAS OBTAINED AND INSERTED WITHOUT DIFFICULTY. NEW LENS IMPLANTED: LENS UV WITH BLUE LIGHT FILTER ULTRASERT 22.0D POWER 13MM LENGTH 6.0MM OPTIC AU00T0 ALCON - CAT# AU00T0.220, EXP 2-29-2020.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7024567 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page