AcrySof® SA6AT9 (REF SA6AT9U175)Recalls & reported problems
Alcon Laboratories, Inc. · AcrySof® Aspheric UV Absorbing Toric Posterior Chamber Intraocular Lens (IOL) · Class 3 · Implant
All models of ACRYSOF (353)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380652362597), received between Nov 16, 2023 and Nov 16, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated1
Patient problems
- Blurred Vision1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 16, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED BLURRED VISION, CLINICAL REASON FOR EXPLANT BEING DISPLACEMENT OF IOL. THE IOL WAS EXPLANTED AND EXCHANGED FOR AN UNKNOWN ATIOL FOLLOWING THE INITIAL IMPLANT PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND RECEIVED STATING THAT THERE WAS NO PATIENT HARM.
Patient problems: Blurred Vision
MAUDE report 18149095 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page