Skip to content

AcrySof® UltraSert® AU00T0 (UDI 00380652394888)Recalls & reported problems

Alcon Laboratories, Inc. · Sterile UV and Blue Light Filtering Acrylic Foldable Single-Piece Posterior Chamber Lenses with the UltraSert® Pre-loaded Delivery System · Class 3 · Implant

All models of AcrySof UltraSert (102)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00380652394888), received between Mar 23, 2021 and Jan 6, 2026 — classified as 2 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged by Another Device4
  • Defective Component3
  • Failure to Eject3
  • Device Difficult to Setup or Prepare2
  • Failure to Advance2
  • Adverse Event Without Identified Device or Use Problem1
  • Defective Device1
  • Inaccurate Delivery1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information4
  • Eye Injury2

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 6, 2026Device Difficult to Setup or Prepare; Inaccurate Delivery

    A PHARMACIST REPORTED WITH DESCRIPTION THAT DURING PROCEDURE WHEN THE LENS WAS DESTERILIZED, THE SURGEON NOTICED PROBLEM WHILE INJECTING THE LENS. ADDITIONAL INFORMATION WAS REQUESTED AND RECEIVED STATING THAT THE LENS WAS NOT COMPLETELY INSIDE THE EYE. THE PROCEDURE WAS COMPLETED USING REPLACEMENT LENS.

    Patient problems: Insufficient Information

    MAUDE report 23973606 ↗
  • MalfunctionOct 17, 2025Defective Component; Defective Component

    A PHARMACIST REPORTED THAT DURING CATARACT SURGERY WITH AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE INJECTOR TIP WAS UNSTABLE AND DETACHES/FALLS OFF DURING USE. THE SURGERY WAS COMPLETED DURING THE SAME PROCEDURE, USED ANOTHER IMPLANT. THE MEDICAL DEVICE DIDN'T TOUCH THE PATIENT'S EYE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23323093 ↗
  • MalfunctionJul 22, 2025Device Difficult to Setup or Prepare; Failure to Advance; Device Damaged by Another Device; Device Difficult to Setup or Prepare; Failure to Advance; Device Damaged by Another Device

    AN ADMINISTRATOR/SUPERVISOR REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PLUNGER WENT OVER THE LENS, DAMAGING BOTH LENS AND THE POSTERIOR HAPTICS WITH NO PATIENT CONTACT, AND CONSEQUENTLY THE LENSES COULD NOT BE INSERTED. THE SURGERY WAS COMPLETED WITH A SECOND LENS. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22566181 ↗
  • InjurySep 24, 2024Defective Component; Adverse Event Without Identified Device or Use Problem; Defective Component; Adverse Event Without Identified Device or Use Problem

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, HE HAD TO WIDEN THE INCISION FOR EVERY SUSPECT DEVICE USED. THE SURGEON STATED THAT THE USUAL INCISION SIZE WAS 2.4 MM, BUT THIS WASN'T ENOUGH FOR THE SUSPECT DEVICE, SO HE HAD TO WIDEN THE INCISION TO 2.5 TO 2.6 MM TO INSERT THE CARTRIDGE TIP INTO THE PATIENT'S EYE. THERE WAS PATIENT CONTACT, BUT NO IMPACT ON THE PATIENT AS THE CATARACT OPERATION WAS COMPLETED WITHOUT ANY PROBLEMS AFTER THE INCISIONS WERE WIDENED. THERE ARE SEVEN MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS 7 OF 7.

    Patient problems: Eye Injury

    MAUDE report 20291654 ↗
  • InjuryJul 29, 2024Defective Component; Defective Component

    A HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, SURGEON INJECTS THE LENS INTO THE POSTERIOR CHAMBER BAG IN THE LEFT EYE. WHILE INSIDE THE EYE, THE INTRAOCULAR LENS WAS DETECTED THAT A HANDLE (HAPTIC) WAS BROKEN. ADDITIONAL INFORMATION WAS RECEIVED, THE INCISION WAS EXPANDED IN THE MAIN PORT WITH A KNIFE FOR IOL EXPLANTATION DURING INITIAL PROCEDURE AND THE PATIENT WAS RECOVERED.

    Patient problems: Eye Injury

    MAUDE report 19850523 ↗
  • MalfunctionJul 9, 2024Failure to Eject; Failure to Eject

    A HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, LENS MISFIRED AND DID NOT COME OUT THE INJECTOR PROPERLY AT TIME OF SURGERY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 19701122 ↗
  • MalfunctionJun 13, 2024Device Damaged by Another Device; Failure to Eject; Device Damaged by Another Device; Failure to Eject

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE PISTON WENT OVER THE LENS AND DAMAGED IT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 19528605 ↗
  • MalfunctionJan 19, 2024Defective Device; Defective Device

    A PHARMACY STUDENT REPORTED FAILURE PLACEMENT WITH THE INTRAOCULAR LENS (IOL) IMPLANT DEVICE. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 18542177 ↗
  • MalfunctionSep 4, 2023Device Damaged by Another Device; Device Damaged by Another Device

    A PHYSICIAN REPORTED THAT THE INTRAOCULAR LENS (IOL) BROKE DUE TO INJECTOR FAILURE. ALSO, THE TWO HAPTICS WERE FOLDED CORRECTLY BUT WHEN INJECTED INSIDE ONE OF THE HAPTICS WAS FRACTURED. ADDITIONAL INFORMATION RECEIVED STATED THAT PATIENT HAD RIGHT EYE PHACO SURGERY. THE LENS WAS REMOVED AND REPLACED WITH ANOTHER LENS IN AN INITIAL PROCEDURE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17676782 ↗
  • MalfunctionMar 13, 2022Device Damaged by Another Device; Failure to Eject; Device Damaged by Another Device; Failure to Eject

    A NURSE REPORTED DURING THE INTRAOCULAR LENS IMPLANTATION ON THE THIRD TRY THE LENS ADVANCED SMOOTHLY AND THE LENS WAS IMPLANTED. THE NURSE THOUGHT SHE HAD SEEN SMALL SCRATCH ON THE SURFACE OF THE LENS. IT WAS ALSO INSTRUCTED THAT THE LENS SHOULD NOT BE USED IF THE PLUNGER GOES ABOVE THE LENS. THERE WAS CONTACT WITH THE PATIENT AND THE LENS REMAINED IMPLANTED. THERE WAS NO PATIENT HARM AND HOSPITALIZATION OR INTERVENTION. THE SURGERY WAS COMPLETED ON THE SAME DAY WITH THE SAME LENS, NO REPLACEMENT LENS WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13751376 ↗
  • MalfunctionMar 23, 2021Failure to Advance; Failure to Advance

    THERE WAS NO CONTACT BETWEEN THE DEVICE AND THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11556452 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page