Skip to content

AcrySof® UltraSert® AU00T0 (UDI 00380652394932)Recalls & reported problems

Alcon Laboratories, Inc. · Sterile UV and Blue Light Filtering Acrylic Foldable Single-Piece Posterior Chamber Lenses with the UltraSert® Pre-loaded Delivery System · Class 3 · Implant

All models of AcrySof UltraSert (102)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380652394932), received between Mar 19, 2021 and Oct 15, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged by Another Device1
  • Device Difficult to Setup or Prepare1
  • Failure to Align1
  • Failure to Eject1
  • Material Too Rigid or Stiff1
  • Unintended Ejection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Capsular Bag Tear1
  • Insufficient Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2025Device Difficult to Setup or Prepare; Device Damaged by Another Device; Device Difficult to Setup or Prepare; Device Damaged by Another Device

    A PHYSICIAN REPORTED THAT DURING INTRAOCULAR LENS IMPLANTATION SURGERY, THE HAPTIC OF THE LENS WAS TIGHTLY WEDGED BETWEEN THE PLUNGER AND CARTRIDGE. AND THE HAPTIC BROKE OFF. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23279323 ↗
  • MalfunctionNov 8, 2023Unintended Ejection; Unintended Ejection

    A NURSE REPORTED THAT DURING THE INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE LENS CAME WRONG OUT FROM THE INJECTOR. THE PROCEDURE WAS COMPLETED WITH THE REPLACEMENT LENS. THERE WAS PATIENT CONTACT BUT SURGEON REMOVED IT WITH NO COMPLICATIONS. THE RE IS NO PATIENT ARM REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18091902 ↗
  • InjuryMay 13, 2023Failure to Align; Failure to Eject; Failure to Align; Failure to Eject

    A NURSE REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE IOL WAS OVERSHOT WHEN IMPLANTING IN TO THE EYE, WHICH RESULTED IN PENETRATION OF THE PATIENT'S THE POSTERIOR CAPSULE. ADDITIONAL INFORMATION HAS BEEN RECEIVED AND STATED THAT, THE FAULTY IOL DAMAGED THE CAPSULAR BAG SO THE ORIGINAL PROCEDURE WAS ABANDONED AND ALTERNATIVE INSTIGATED. THE SURGEON HAD DONE ANTERIOR VITRECTOMY AND EXPLANT OF FAULTY IOL AND ALTERNATIVE 3 PIECE IOL PLACED IN SULCUS. THE PATIENT WAS NOT HOSPITALIZED FOR THE EVENT.

    Patient problems: Capsular Bag Tear

    MAUDE report 16925335 ↗
  • MalfunctionMar 19, 2021Material Too Rigid or Stiff; Material Too Rigid or Stiff

    A CUSTOMER REPORTED THE PLUNGER OF A PRELOADED INTRAOCULAR LENS (IOL) DELIVERY SYSTEM WAS STIFF. THERE WAS NO REPORTED PATIENT IMPACT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 11530921 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page