Skip to content

AcrySof® UltraSert® ACU0T0 (REF ACU0T0.135)Recalls & reported problems

Alcon Laboratories, Inc. · Sterile UV and Blue Light Filtering Acrylic Foldable Single-Piece Posterior Chamber Lenses with the UltraSert® Pre-loaded Delivery System · Class 3 · Implant

All models of AcrySof UltraSert (102)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00380652395120), received between Jun 15, 2022 and Sep 19, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Advance2
  • Defective Device1
  • Device Damaged by Another Device1
  • Device Difficult to Setup or Prepare1
  • Failure to Eject1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 19, 2025Device Damaged by Another Device; Device Damaged by Another Device

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE WHILE LOADING, THE LENS WAS TWISTED OR SQUASHED ON THE RIGHT SIDE. THE SURGERY WAS COMPLETED ON THE SAME DAY WITH ANOTHER LENS. THERE WERE NO PATIENT CONTACT AND NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23095166 ↗
  • MalfunctionAug 11, 2025Failure to Advance; Failure to Advance

    A HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, IT WAS NOTICED THAT THE LENS WAS TRAPPED UNDER THE PLUNGER AND THEREFORE NOT COMING OUT OF THE INJECTOR. THERE WAS PATIENT CONTACT AND NO PATIENT HARM. THE PROCEDURE COMPLETED ON THE SAME DAY WITH ANOTHER UNSPECIFIED LENS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22773253 ↗
  • MalfunctionMar 24, 2025Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    NON-HEALTHCARE PROFESSIONAL REPORTED DURING INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, WITH A DESCRIPTION OF TRAILING HAPTIC FAULTY AS NOT POSITIONED CORRECTLY ON INJECTOR. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21678001 ↗
  • MalfunctionAug 27, 2024Defective Device; Defective Device

    A HEALTH CARE PROFESSIONAL REPORTED FOUR DEFECTIVE PRODUCTS. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 20073987 ↗
  • MalfunctionApr 15, 2024Failure to Advance; Failure to Advance

    A SERVICE MANAGER REPORTED ABOUT A COMPLAINT OF LENS STUCK IN THE INJECTOR. THERE WAS NO PATIENT HARM. THE SURGERY WAS COMPLETED WITH A NEW LENS. THERE ARE THREE MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS REPORT. THIS IS 3 OF 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19105410 ↗
  • MalfunctionJun 15, 2022Failure to Eject; Failure to Eject

    A NURSE REPORTED THAT, DURING INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PLUNGER WENT OVER THE IOL. THERE WAS A CONTACT WITH THE PATIENT BUT NO PATIENT HARM INVOLVED. THE SURGERY WAS DELAYED FOR A MINUTE. THE SURGERY WAS COMPLETED BY USING A BACK-UP LENS AND WITHOUT ANY OTHER PROBLEM. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14696666 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page