AcrySof® UltraSert® ACU0T0 (REF ACU0T0.255)Recalls & reported problems
Alcon Laboratories, Inc. · Sterile UV and Blue Light Filtering Acrylic Foldable Single-Piece Posterior Chamber Lenses with the UltraSert® Pre-loaded Delivery System · Class 3 · Implant
All models of AcrySof UltraSert (102)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00380652395366), received between Jul 2, 2021 and Feb 16, 2026 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Advance3
- Break2
- Defective Device2
- Failure to Eject2
- Crack1
- Device Difficult to Setup or Prepare1
- Difficult to Advance1
- Failure to Deliver1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information3
- Blurred Vision1
- Corneal Edema1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 16, 2026Scratched Material
A PHARMACIST REPORTED WITH A DESCRIPTION, THE LENS DISC WAS SCRATCHED. THE SURGERY WAS COMPLETED ON THE SAME DAY. THERE WAS NO PATIENT CONTACT AND NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED AND STATED, THE SCRATCH WAS DETECTED DURING THE SURGICAL PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24343769 ↗MalfunctionNov 9, 2025Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
A HEALTH CARE PROFESSIONAL REPORTED THAT DURING THE CATARACT SURGERY WITH INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE HAPTIC DID NOT COME OUT CORRECTLY. THE PROCEDURE WAS COMPLETED WITH ANOTHER LENS. THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23507061 ↗MalfunctionOct 24, 2025Break; Break
A HEALTH CARE PROFESSIONAL REPORTED THAT DURING THE CATARACT SURGERY WITH INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE UPON INSERTION OF THE IOL INTO THE PATIENT'S EYE, THE TRAILING HAPTIC WAS FOUND TO BE BROKEN AND REMAINED LODGED WITHIN THE LENS INTRODUCER. AS A RESULT, THE IMPLANTED LENS HAD TO BE CAREFULLY CUT AND REMOVED. A NEW IOL WAS INSERTED WITHOUT COMPLICATIONS, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT IMPACT REPORTED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23380434 ↗MalfunctionJul 17, 2025Failure to Advance; Failure to Advance
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PLUNGER OVERRODE THE LENS. THERE WAS NO PATIENT HARM. THE SURGERY COMPLETED WITH BACKUP LENS. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22533534 ↗MalfunctionApr 23, 2025Failure to Advance; Failure to Advance
A HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, IT WAS NOTED THAT THE INJECTOR WENT OVER THE LENS. THERE WAS NO PATIENT HARM. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21889381 ↗InjuryOct 24, 2024Crack; Crack
A HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE LENS WAS INJECTED AND UNFORTUNATELY ONCE OPENED IN THE EYE IT WAS NOTICED THAT THE LENS HAD A LARGE CRACK IN THE CENTER. THE LENS REMOVED AND REPLACED AND THERE WAS NO COMPLICATIONS. IN FOLLOW-UP VISIT THE SURGEON OBSERVED CORNEAL EDEMA DUE TO PROLONGED SURGERY AND ONLY ABLE TO VISUALLY COUNT FINGERS AND TREATMENT WAS GIVEN. ADDITIONAL INFORMATION HAS BEEN RECEIVED THAT PATIENT HAD SUTURE OF WOUND.
Patient problems: Blurred Vision; Corneal Edema
MAUDE report 20520672 ↗MalfunctionFeb 2, 2024Defective Device; Defective Device
A NON HEALTHCARE PROFESSIONAL REPORTED WITH THE DESCRIPTION FAULTY. ADDITIONAL INFORMATION HAS BEEN REQUESTED. THERE ARE TWO MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS FILE. THIS REPORT IS ASSOCIATED WITH THE 1 OF 2 FILES.
Patient problems: Insufficient Information
MAUDE report 18629088 ↗MalfunctionAug 14, 2023Break; Failure to Eject; Break; Failure to Eject
A HEALTHCARE PROFESSIONAL REPORTED DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE LENS DID NOT INJECT PROPERLY INTO THE EYE, PART OF THE LENS WAS STILL IN THE DELIVERY SYSTEM. THE LEADING HAPTIC WAS RIPPED AND IT WAS STILL IN THE DELIVERY SYSTEM. THE SURGERY WAS COMPLETED WITH ANOTHER LENS.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17540379 ↗MalfunctionAug 4, 2023Defective Device; Defective Device
A NURSING CHIEF REPORTED THAT DURING A CATARACT EXTRACTION WITH INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE IOL CAME DEFECTIVE INTO THE EYE. ADDITIONAL INFORMATION HAS BEEN REQUESTED. THERE ARE ELEVEN MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS REPORT. THIS REPORT IS FOR 11 OF 11.
Patient problems: Insufficient Information
MAUDE report 17456619 ↗MalfunctionJul 13, 2021Failure to Advance; Failure to Advance
A PHYSICIAN REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANTATION THE LENS WAS STUCK IN THE PISTON AND IN ORDER TO INSERT IT INTO THE PATIENT'S EYE, THE SURGEON HAD TO HOLD IT USING THE CANNULA. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12159448 ↗MalfunctionJul 2, 2021Failure to Deliver; Difficult to Advance; Failure to Eject; Physical Resistance/Sticking; Failure to Deliver; Difficult to Advance; Failure to Eject; Physical Resistance/Sticking
A SAFETY MANAGER REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE THE LENS DELIVERY FAILED. THE SURGEON NOTICED THE RESISTANCE IN THE INJECTOR, THE INJECTOR ARM DID NOT ENGAGE THE LENS AND PASSED OVER THE IOL. THERE WAS A PATIENT CONTACT WITH NO PATIENT HARM. ADDITIONAL INFORMATION WAS REQUESTED AND RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12108885 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page