Clareon™ CCA0T0 (UDI 00380652397193)Recalls & reported problems

Alcon Laboratories, Inc. · STERILE UV Absorbing Hydrophobic Acrylic Foldable Aspheric Posterior Chamber Intraocular Lens with the AutonoMe™ Automated Pre-loaded Delivery System · Class 3 · Implant

All models of Clareon (1142)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00380652397193), received between Aug 26, 2022 and Apr 23, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged by Another Device3
  • Defective Device1
  • Device Difficult to Setup or Prepare1
  • Failure to Advance1
  • Product Quality Problem1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2026Device Damaged by Another Device

    A NURSE REPORTED THAT DURING INTRAOCULAR IMPLANTATION SURGERY THE SURGEON NOTICED THE HAPTIC WAS BROKEN PRIOR TO INSERTING THE LENS. SCHEDULED PROCEDURE COMPLETED THE SAME DAY AND THERE WAS NO PATIENT CONTACT. IN THE SURGEON¿S OPINION, DELIVERY SYSTEM CAUSED OR CONTRIBUTED TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24966470 ↗
  • MalfunctionDec 4, 2025Device Damaged by Another Device; Device Damaged by Another Device

    A NONHEALTH CARE PROFESSIONAL REPORTED THAT DURING THE INTRAOCULAR LENS (IOL)IMPLANTATION SURGERY, THE HAPTIC WAS BROKEN. THE PROCEDURE WAS COMPLETED ON THE SAME DAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23717459 ↗
  • MalfunctionSep 12, 2025Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    A NURSE REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE FRONT HAPTIC HAD KINKED OFF WHILE ATTEMPTING TO ADVANCE THE LENS. THE SURGEON HAD REQUESTED A NEW LENS. THE SURGERY WAS COMPLETED USING ANOTHER COMPANY'S LENS. THERE HAD BEEN NO PATIENT CONTACT AND NO HARM TO THE PATIENT. THE WOUND HAD NOT BEEN ENLARGED, AND NO SUTURE HAD BEEN NEEDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23039308 ↗
  • MalfunctionAug 22, 2025Product Quality Problem; Product Quality Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, PLUNGER SLID PAST IOL PRIOR TO LENS INSERTION. THE NEW LENS GOT IMPLANTED AND THE PROCEDURE WAS COMPLETED. ADDITIONAL INFORMATION HAS BEEN RECEIVED STATING THAT FAULTY LENS WHILE LOADING DEVICE WITH CONFIRMED NO CONTACT WITH PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED HOWEVER NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22874449 ↗
  • MalfunctionJun 23, 2025Scratched Material; Scratched Material

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING INTRAOCULAR LENS (IOL) IMPLANTATION, THERE WAS A SCRATCH ON LENS THAT WAS NOTICED AFTER IT WAS IMPLANTED. ADDITIONAL INFORMATION RECEIVED STATING THAT PROCEDURE WAS COMPLETED ON SAME DAY AND LENS REMAINS IMPLANTED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22305949 ↗
  • MalfunctionApr 2, 2025Device Damaged by Another Device; Device Damaged by Another Device

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, LENS DAMAGED FROM THE AUTONOME AUTOLOADER. THIS LENS HAD NOT NEEDED TO BE EXPLANTED BECAUSE IT WAS NOT DAMAGED THE CENTER OPTIC, HOWEVER LENS NICKED OR DAMAGED ON THE PERIPHERY EDGE OF THE OPTIC. ADDITIONAL INFORMATION RECEIVED STATING THAT COMPANY DEVICE USED AS USUAL AND THE PERIPHERY EDGE OF THE LENS WAS DAMAGED. AS PER THE SURGEON, THE INSERTER DAMAGED THE LENS. THERE WAS NO PATIENT INJURY. THERE ARE TWO MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS REPORT. THIS FILE IS 1 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21755883 ↗
  • MalfunctionNov 15, 2024Failure to Advance; Failure to Advance

    A NURSE REPORTED WITH A DESCRIPTION OF INTRAOCULAR LENS GOT STUCK/CAUGHT IN INCISION DURING INSERTION. SUBSEQUENTLY THE LENS WAS REMOVED DURING INITIAL PROCEDURE AND COMPLETED WITH ANOTHER LENS. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20692579 ↗
  • MalfunctionAug 26, 2022Defective Device; Defective Device

    A REPORT RECEIVED FROM THE HEALTH AUTHORITY THAT, DURING CATARACT SURGERY WITH INTRAOCULAR LENS IMPLANT PROCEDURE, THE LENS FOUND TO BE DEFECTIVE. THE LENS WAS REMOVED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 15302820 ↗

Other intraocular lens devices

All intraocular lens devices →