Clareon CNA0T3 (REF CNA0T3.180)Recalls & reported problems
Alcon Laboratories, Inc. · CLAREON TORIC AUTONOME · Class 3 · Implant
All models of Clareon (1142)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00380652398749), received between Apr 17, 2024 and Mar 9, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Material Discolored” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Discolored7
- Device Difficult to Setup or Prepare1
- Failure to Fold1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2026Device Difficult to Setup or Prepare
A HEALTHCARE PROFESSIONAL STATED THAT HAPTIC WAS TWISTED DURING SURGERY. WHEN PREPARING TO LOAD THE LENS, THE SURGEON NOTICED A KINK IN THE HAPTIC. THE SURGERY WAS COMPLETED ON THE SAME DAY. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24546539 ↗MalfunctionApr 29, 2025Material Discolored; Material Discolored
A PHYSICIAN REPORTED THAT AFTER IMPLANTATION OF LENS DURING THE INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE IOL LOOKED BLUE IMMEDIATELY. THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. IT WAS UNKNOWN WHETHER THESE FINDINGS DISAPPEARED OR CONTINUED. THE LENS REMAINED IMPLANTED. THERE HAVE BEEN NO REPORTED HEALTH PROBLEMS OR HARM TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE 224 MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS IS 32 OF 224.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21929198 ↗MalfunctionApr 29, 2025Material Discolored; Material Discolored
A PHYSICIAN REPORTED THAT AFTER IMPLANTATION OF LENS DURING THE INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE IOL LOOKED BLUE IMMEDIATELY. THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. IT WAS UNKNOWN WHETHER THESE FINDINGS DISAPPEARED OR CONTINUED. THE LENS REMAINED IMPLANTED. THERE HAVE BEEN NO REPORTED HEALTH PROBLEMS OR HARM TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE 224 MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS IS 31 OF 224.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21929194 ↗MalfunctionApr 29, 2025Material Discolored; Material Discolored
A PHYSICIAN REPORTED THAT AFTER IMPLANTATION OF LENS DURING THE INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE IOL LOOKED BLUE IMMEDIATELY. THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. IT WAS UNKNOWN WHETHER THESE FINDINGS DISAPPEARED OR CONTINUED. THE LENS REMAINED IMPLANTED. THERE HAVE BEEN NO REPORTED HEALTH PROBLEMS OR HARM TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE 224 MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS IS 143 OF 224.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21928192 ↗MalfunctionApr 29, 2025Material Discolored; Material Discolored
A PHYSICIAN REPORTED THAT AFTER IMPLANTATION OF LENS DURING THE INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE IOL LOOKED BLUE IMMEDIATELY. THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. IT WAS UNKNOWN WHETHER THESE FINDINGS DISAPPEARED OR CONTINUED. THE LENS REMAINED IMPLANTED. THERE HAVE BEEN NO REPORTED HEALTH PROBLEMS OR HARM TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE 224 MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS IS 1 OF 224.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21927045 ↗MalfunctionApr 29, 2025Material Discolored; Material Discolored
A PHYSICIAN REPORTED THAT AFTER IMPLANTATION OF LENS DURING THE INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE IOL LOOKED BLUE IMMEDIATELY. THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. IT WAS UNKNOWN WHETHER THESE FINDINGS DISAPPEARED OR CONTINUED. THE LENS REMAINED IMPLANTED. THERE HAVE BEEN NO REPORTED HEALTH PROBLEMS OR HARM TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE 224 MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS IS 56 OF 224.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21926401 ↗MalfunctionApr 29, 2025Material Discolored; Material Discolored
A PHYSICIAN REPORTED THAT AFTER IMPLANTATION OF LENS DURING THE INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE IOL LOOKED BLUE IMMEDIATELY. THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. IT WAS UNKNOWN WHETHER THESE FINDINGS DISAPPEARED OR CONTINUED. THE LENS REMAINED IMPLANTED. THERE HAVE BEEN NO REPORTED HEALTH PROBLEMS OR HARM TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE 224 MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS IS 64 OF 224
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21926400 ↗MalfunctionDec 12, 2024Material Discolored; Material Discolored
A HEALTHCARE PROFESSIONAL REPORTED THAT E HAD NOTICED RESIDUE ON THE IOL INTRAOPERATIVELY WHICH DISAPPEAR POSTOPERATIVELY. THE LENS REMAINS IMPLANTED IN EYE.
Patient problems: Insufficient Information
MAUDE report 20916555 ↗MalfunctionApr 17, 2024Failure to Fold; Failure to Fold
A NON-HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, FAULTY LENS WAS NOT FOLDED PROPERLY. ADDITIONAL INFORMATION WAS REQUESTED AND RECEIVED. THE LENS WAS REPLACED WITH THE UNSPECIFIED LENS (BACK UP LENS WAS USED). THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19126161 ↗
Other Intraocular Lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page