Skip to content

Clareon CNA0T3 (REF CNA0T3.210)Recalls & reported problems

Alcon Laboratories, Inc. · CLAREON TORIC AUTONOME · Class 3 · Implant

All models of Clareon (1142)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00380652398800), received between Aug 29, 2024 and Sep 22, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Material Discolored” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Discolored3
  • Failure to Fold2
  • Defective Component1
  • Failure to Advance1
  • Failure to Deliver1
  • Scratched Material1

Patient problems

  • Insufficient Information5
  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026Scratched Material

    A healthcare professional reported that during an intraocular lens implant procedure dirt was noticed (adherence) when implanting. An attempt was made to remove the adherent material during surgery, but as it could not be removed. Upon reviewing the footage of the positioning step, a deposit or scratch-like mark is visible before the plunger advances the lens. The lens was exchanged during the initial procedure and replaced with unspecified lens.

    Patient problems: Insufficient Information

    MAUDE report 50852203 ↗
  • MalfunctionJan 7, 2026Failure to Deliver

    ADDITIONAL INFORMATION WAS RECEIVED AND STATED, PLUNGER DID NOT ENGAGE IOL PROPERLY AND BENT UNDERNEATH. THE ISSUE WAS NOTICED DURING PRIMING OF THE DEVICE AND THERE WAS NO PATIENT CONTACT.

    Patient problems: Insufficient Information

    MAUDE report 23988105 ↗
  • MalfunctionApr 29, 2025Material Discolored; Material Discolored

    A PHYSICIAN REPORTED THAT AFTER IMPLANTATION OF LENS DURING THE INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE IOL LOOKED BLUE IMMEDIATELY. THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. IT WAS UNKNOWN WHETHER THESE FINDINGS DISAPPEARED OR CONTINUED. THE LENS REMAINED IMPLANTED. THERE HAVE BEEN NO REPORTED HEALTH PROBLEMS OR HARM TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE (B)(4) MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS IS (B)(4) OF (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21928041 ↗
  • MalfunctionApr 29, 2025Material Discolored; Material Discolored

    A PHYSICIAN REPORTED THAT AFTER IMPLANTATION OF LENS DURING THE INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE IOL LOOKED BLUE IMMEDIATELY. THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. IT WAS CURRENTLY UNKNOWN WHETHER THESE FINDINGS WERE DISAPPEARED OR CONTINUING. THERE HAVE BEEN NO REPORTED HEALTH PROBLEMS OR HARM TO THE PATIENT. THE LENS REMAIN IMPLANTED. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE 224 MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS IS 199 OF 224.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21925867 ↗
  • MalfunctionApr 29, 2025Material Discolored; Material Discolored

    A PHYSICIAN REPORTED THAT AFTER IMPLANTATION OF LENS DURING THE INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE IOL LOOKED BLUE IMMEDIATELY. THE SURGERY WAS COMPLETED WITHOUT PRODUCT REPLACEMENT. IT WAS UNKNOWN WHETHER THESE FINDINGS DISAPPEARED OR CONTINUED. THE LENS REMAINED IMPLANTED. THERE HAVE BEEN NO REPORTED HEALTH PROBLEMS OR HARM TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE 224 MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS EVENT. THIS IS 181 OF 224.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21925447 ↗
  • MalfunctionAug 29, 2024Failure to Fold; Failure to Fold

    A HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, LENS DID NOT FOLD PROPERLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 20102014 ↗
  • MalfunctionAug 29, 2024Failure to Fold; Defective Component; Failure to Fold; Defective Component

    A HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, HAPTIC TWISTED, LENS DID NOT FOLD PROPERLY. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 20096383 ↗
  • MalfunctionAug 29, 2024Failure to Advance; Failure to Advance

    A HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, LENSES DIDN¿T FOLD APPROPRIATELY ALSO LENS GOT STUCK TO ROOF OF THE CARTRIDGE.

    Patient problems: Insufficient Information

    MAUDE report 20096339 ↗

Other Intraocular Lens devices

All Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page