Clareon CCWET0 (UDI 00380652405669)Recalls & reported problems

Alcon Laboratories, Inc. · Clareon Vivity UVA · Class 3 · Implant

All models of Clareon (1142)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00380652405669), received between Oct 18, 2022 and Apr 30, 2025 — classified as 4 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Scratched Material” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Scratched Material3
  • Failure to Eject1

Patient problems

  • Blurred Vision4
  • Insufficient Information3
  • Visual Impairment2
  • Dry Eye(s)1
  • No Clinical Signs, Symptoms or Conditions1
  • Visual Disturbances1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 30, 2025Scratched Material; Scratched Material

    NON-HEALTHCARE PROFESSIONAL DURING SURGERY REPORTED THAT CARTRIDGES POSSIBLY ARRIVED CRACKED, THEN THEY PULLED ALL THE CARTRIDGES. ADDITIONALLY THEY HAD TWO LENSES THAT SEEMED TO BE SCRATCHED AND MENTIONED THAT AFTER TALKING WITH THE TEAM IF SOUNDED THAT THE LENS POSSIBLY SAT TOO LONG AND WAS TOO TIGHT TO PASS THROUGH THE CARTRIDGE. THEY GRABBED A NEW LENS BUT THE SAME CARTRIDGE AND NOTICED THE CRACK. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 21935885 ↗
  • MalfunctionApr 30, 2025Scratched Material; Scratched Material

    NON-HEALTHCARE PROFESSIONAL DURING SURGERY REPORTED THAT CARTRIDGES POSSIBLY ARRIVED CRACKED, THEN THEY PULLED ALL THE CARTRIDGES. ADDITIONALLY, THEY HAD TWO LENSES THAT SEEMED TO BE SCRATCHED AND MENTIONED THAT AFTER TALKING WITH THE TEAM IF SOUNDED THAT THE LENS POSSIBLY SAT TOO LONG AND WAS TOO TIGHT TO PASS THROUGH THE CARTRIDGE. THEY GRABBED A NEW LENS BUT THE SAME CARTRIDGE AND NOTICED THE CRACK. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 21933445 ↗
  • InjuryAug 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED BLURRY VISION. THE LENS WAS EXPLANTED AND REPLACED WITH ANOTHER LENS IN A SECONDARY PROCEDURE. THE CLINICAL REASON FOR THE EXPLANT WAS VISUAL DISTURBANCE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Blurred Vision; Visual Impairment

    MAUDE report 19938905 ↗
  • MalfunctionApr 8, 2024Scratched Material; Scratched Material

    A NON HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE LENS WAS SCRATCHED. THE LENS WAS EXPLANTED AND REPLACED WITH SAME MODEL AND DIAPTOR IN THE INITIAL PROCEDURE . ADDITIONAL INFORMATION HAS BEEN RECEIVED AND STATED THAT THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19059551 ↗
  • MalfunctionDec 11, 2023Failure to Eject; Failure to Eject

    A FACILITY REPRESENTATIVE REPORTED STATING STUCK IN LOADER. THERE WAS PATIENT CONTACT. TYPE OF PROCEDURE WAS BEING PERFORMED WAS CATARACT. THE PROCEDURE COMPLETED THAT DAY AND THE PRODUCT WAS NOT AVAILABLE TO RETURN. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 18300719 ↗
  • InjuryNov 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON HEALTHCARE PROFESSIONAL REPORTED THAT, AFTER INTRAOCULAR LENS IMPLANTATION SURGERY, THE PATIENT EXPERIENCED TERRIBLE NEAR VISION. ON PRE- OP VISIT, THE SURGEON INFORMED THAT PATIENT THAT AFTER SURGERY, THE PATIENT COULD NOT BE ABLE TO SEE ANYTHING CLOSE SUCH AS COMPUTER SCREEN, CELL PHONES, BOOK OR LABELS AT A STORE ON USING THE SPECIALTY LENS WITHOUT READING GLASSES. ON (B)(6) 2023, THE PATIENT HAD CATARACT SURGERY ON ONE EYE AND ON POST OPERATIVE VISIT, THE PATIENT INFORMED THE SURGEON THAT, HIS FAR VISION WAS GOOD BUT THE PATIENT COULD NOT READ ANYTHING CLOSE. THE DOCTOR INFORMED PATIENT THAT, HIS VISION WOULD BE BETTER AFTER A WHILE. A WEEK AFTER SURGERY ON HIS FIRST EYE, THE PATIENT HAD THE SECOND LENS IMPLANTED. SEVERAL WEEKS LATER, DURING THE POST-OP MEETING WITH ANOTHER DOCTOR, THE PATIENT AGAIN INFORMED HIM THAT HIS NEAR VISION WAS TERRIBLE AND I COULDN¿T READ ANYTHING WITHOUT READING GLASSES. THE SURGEON RECOMMENDED THAT, THEY COULD NOT DO ANYTHING ABOUT THE LENS WHICH WAS IMPLANTED BECAUSE ANOTHER SURGERY MAY CAUSE FURTHER DAMAGE. ADDITIONAL INFORMATION RECEIVED FROM THE CONSUMER AND STATED THAT, THE PATIENT WAS NOT HAPPY WITH THE LENSES. THERE ARE TWO MEDICAL DEVI

    Patient problems: Blurred Vision

    MAUDE report 18204261 ↗
  • InjuryAug 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CONSUMER REPORTED FOLLOWING THE INTRAOCULAR LENS IMPLANTATION REPORTS OF HAVING ISSUES WITH CONTRAST SENSITIVITY, BRIGHTNESS. FREQUENT IRRADIC CLOUDY VISION MOSTLY RELATED TO DRYNESS FOR WHICH THE PATIENT WAS USING DROPS WITH NO CHANGE IN VISION. ALSO THE FIELD OF VISION IN THE EYE IS DECREASED. AN EXPLANT OF THE LENS HAS BEEN PLANNED. ADDITIONAL INFORMATION HAS BEEN RECEIVED STATING THE PATIENT HAD EXPERIENCED LACK OF BRIGHTNESS AND CLARITY AT NIGHT. ADDITIONAL INFORMATION HAS BEEN REQUESTED STATING THAT THE LENS WAS EXPLANTED IN A SECONDARY PROCEDURE. THE VISION WAS STILL NOT CLEAR TO BE ABLE TO SEE LETTERS CLEARLY BUT IT WAS IMPROVING.

    Patient problems: Blurred Vision; Dry Eye(s); Visual Impairment

    MAUDE report 17478404 ↗
  • InjuryOct 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A HEALTH CARE PROFESSIONAL REPORTED THAT FOLLOWING IMPLANTATION OF AN INTRAOCULAR LENS (IOL) THE PATIENT HAD EXPERIENCED BLURRY AND DOUBLE VISION. THE LENS WAS EXPLANTED IN A SECONDARY PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Blurred Vision; Visual Disturbances

    MAUDE report 15621944 ↗

Other intraocular lens devices

All intraocular lens devices →