Clareon CCWET0 (UDI 00380652405805)Recalls & reported problems
Alcon Laboratories, Inc. · Clareon Vivity UVA · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Alcon Clareon Vivity UVA is a lens implanted during cataract surgery. The 31 reports mostly describe patients seeing blurry or unclear vision, halos and glare, and lens surfaces found scratched during surgery; in many cases the lens was removed and replaced with a different one. There were24 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 24
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 00380652405805), received between Jul 8, 2022 and Jun 26, 2026 — classified as 24 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Defective Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem19
- Defective Device3
- Scratched Material3
- Appropriate Device Problem Term/Code Not Available2
- Break1
- Defective Component1
- Device Contamination with Chemical or Other Material1
- Device Damaged by Another Device1
Patient problems
- Blurred Vision13
- Insufficient Information9
- Visual Disturbances6
- Visual Impairment6
- No Clinical Signs, Symptoms or Conditions4
- Halo1
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
InjuryJun 26, 2026Adverse Event Without Identified Device or Use Problem
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED INADEQUATE VISION AND CLINICAL REASON VISION WAS NOT CLEAR ENOUGH FOR HIS SHOOTING EYE. THE IOL WAS EXCHANGED FOR MONOFOCAL NON-COMPANY REPLACEMENT LENS 07 DAYS FOLLOWING THE INITIAL IMPLANT PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Blurred Vision
MAUDE report 25633985 ↗InjuryApr 5, 2026Adverse Event Without Identified Device or Use Problem
A NON-HEALTH CARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT VISION WAS NOT CLEAR IN THE RIGHT EYE AND HALOS ARE WORSE. THE LENS WAS EXCHANGED FOR ANOTHER LENS MODEL 03 MONTHS 14 DAYS FOLLOWING THE INITIAL PROCEDURE. CLINICAL REASON FOR EXPLANT WAS MENTIONED AS REFRACTION. ADDITIONAL INFORMATION WAS RECEIVED AND STATED, 1 WEEK POST-OPERATIVE VISION WAS STILL BLURRY AND SEEING HALOS. VISION NEVER BECAME CLEAR. THERE WAS NO PATIENT HARM.
Patient problems: Blurred Vision; Visual Disturbances; Visual Impairment
MAUDE report 24786612 ↗InjuryMar 17, 2026Adverse Event Without Identified Device or Use Problem
AN OPHTHALMIC SURGEON REPORTED THAT AFTER THE INTRAOCULAR LENS (IOL) IMPLANTATION SURGERY, THE PATIENT HAD REPORTED BLURRY VISION, SHADOWS, AND AN INABILITY TO WORK CUTTING HAIR. THE CLINICAL REASON FOR EXPLANT WAS DYSPHOTOPSIA AND MECHANICAL COMPLICATIONS OF THE IOL. THE LENS WAS REMOVED AND REPLACED WITH A MONOFOCAL LENS IN A SECONDARY PROCEDURE." ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 24618060 ↗MalfunctionMar 7, 2026Scratched Material
A NON-HEALTHCARE PROFESSIONAL REPORTED DURING INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE SURGEON NOTICED SCRATCHES IN BOTH LENSES IN THE SAME CASE. THE LENSES WERE EXPLANTED AND REPLACED WITH ALTERNATIVE REPLACEMENT LENS WITH 21.0 DIOPTER DURING INITIAL PROCEDURE. THE PROCEDURE WAS COMPLETED ON SAME DAY WITHOUT ISSUES. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND RECEIVED STATING SURGEON OBSERVED A SCRATCH ON 2 LENSES THAT WERE INJECTED INTO THE EYE, BOTH WERE REMOVED. THE LENS WAS REPLACED WITH A LENS WITHOUT ISSUE AND PROCEDURE COMPLETED ON SAME DAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24541103 ↗MalfunctionMar 7, 2026Scratched Material
A NON-HEALTHCARE PROFESSIONAL REPORTED DURING INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE SURGEON NOTICED SCRATCHES IN BOTH LENSES IN THE SAME CASE. THE LENSES WERE EXPLANTED AND REPLACED WITH ALTERNATIVE REPLACEMENT LENS WITH 21.0 DIOPTER DURING INITIAL PROCEDURE. THE PROCEDURE WAS COMPLETED ON SAME DAY WITHOUT ISSUES. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND RECEIVED STATING SURGEON OBSERVED A SCRATCH ON 2 LENSES THAT WERE INJECTED INTO THE EYE, BOTH WERE REMOVED. THE LENS WAS REPLACED WITH A LENS WITHOUT ISSUE AND PROCEDURE COMPLETED ON SAME DAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24539865 ↗InjuryMar 3, 2026Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, PATIENT REASON AND CLINICAL REASON AS MECHANICAL COMPLICATION. THE IOL WAS EXPLANTED AND EXCHANGED FOR AN UNKNOWN ADVANCED TECHNOLOGY INTRAOCULAR LENS FOLLOWING THE INITIAL IMPLANT PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND RECEIVED STATING THAT THE PATIENT ANGLE ALPHA IS ALMOST 6OU, HE WAS SEEING EDGE OF BUTTON ON THE LENS. IN THE SURGEONS OPINION, LENS CONTRIBUTED TO THE EVENT. THE IOL WAS EXPLANTED AND EXCHANGED FOR A NON COMPANY LENS. THERE WAS NO FURTHER IMPACT TO THE PATIENT. ISSUES WERE RESOLVED.
Patient problems: Visual Disturbances
MAUDE report 24496964 ↗InjuryOct 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT HAD BEEN FRUSTRATED BECAUSE SHE HAD NOT BEEN ABLE TO READ WITHOUT HER GLASSES. THE CLINICAL REASON WAS MENTIONED AS MECHANICAL ISSUES WITH THE IOL. CATARACT SURGERY HAD BEEN PERFORMED FOR HER NARROW ANGLE GLAUCOMA, AND THE PATIENT HAD NOT BEEN FUNCTIONING WELL. THE IOL WAS EXPLANTED AND EXCHANGED WITH AN UNSPECIFIED MONOFOCAL LENS DURING THE SECONDARY IMPLANT PROCEDURE. ADDITIONAL INFORMATION WAS RECEIVED BY STATING, THERE WAS NO FURTHER IMPACT TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Visual Impairment
MAUDE report 23348981 ↗InjuryJul 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT COULD NOT ADJUST WITH COMPANY LENS AND HAD VISION BLURRY. CLINICAL REASON FOR EXPLANT INCLUDED PATIENT WAS NOT SATISFIED WITH VISUAL OUTCOME. THE IOL WAS EXCHANGED FOR MONOFOCAL NON-COMPANY LENS 06 MONTHS 15 DAYS FOLLOWING THE INITIAL IMPLANT PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Blurred Vision
MAUDE report 22521551 ↗MalfunctionMay 22, 2025Device Damaged by Another Device; Device Damaged by Another Device
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT, DURING INTRAOCULAR LENS IMPLANTATION SURGERY, UPON INSERTION SURGEON NOTED THAT THE LENS WAS SCRATCHED, BUT HE FELT THAT THE INSERTER ARM WAS BENT AND HAD DRAGGED OVER THE TOP OF THE LENS. HENCE THE LENS WAS REMOVED AND A BACKUP LENS WAS INSERTED USING A NEW CARTRIDGE AND LOADER. THERE WAS PATIENT CONTACT. PROCEDURE COMPLETED ON THE SAME DAY USING ANOTHER LENS.
Patient problems: Insufficient Information
MAUDE report 22069111 ↗InjuryApr 30, 2025Break; Device Dislodged or Dislocated; Break; Device Dislodged or Dislocated
A NON-HEALTH CARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT IOL HAD SHIFTED INFERIOR OR LENS WAS NOT IN PLACE IN THE RIGHT EYE. THE LENS WAS EXCHANGED FOR ANOTHER LENS MODEL 09 DAYS FOLLOWING THE INITIAL PROCEDURE. CLINICAL REASON FOR EXPLANT WAS MENTIONED AS; THE DISPLACED IOL HAD BROKEN HAPTIC. ADDITIONAL INFORMATION WAS RECEIVED AND STATED, DUE TO THE DAMAGE OF THE HAPTIC IOL WAS DISLOCATED. IN SURGEONS' OPINION IT IS UNCLEAR HOW OR WHEN THE HAPTIC WAS DAMAGED AS NO ISSUES WERE NOTED UPON INITIAL PLACEMENT OF THE IOL. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21934468 ↗InjuryFeb 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT VISION IS NOT CRISP AND CLEAR. THE IOL WAS EXCHANGED FOR UNSPECIFIED LENS MODEL 3MONTHS AND 11 DAYS AFTER THE INITIAL IMPLANT PROCEDURE. CLINICAL REASON FOR EXPLANT WAS REPORTED AS VISUAL DISTURBANCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 21417910 ↗MalfunctionDec 11, 2024Defective Component; Defective Component
A NON-HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE HAPTIC WAS BROKEN INSIDE THE EYE AND WERE ABLE TO RETRIEVE THE LENS AND HAPTICS IN PIECES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 20901621 ↗InjuryOct 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT VISUAL ACUITY WAS WORSENING. THE IOL WAS EXCHANGED FOR UNSPECIFIED LENS MODEL 28 DAYS AFTER THE INITIAL IMPLANT PROCEDURE. CLINICAL REASON FOR EXPLANT WAS MYOPIC SURPRISE AND -1.25 D OFF PLANO TARGET. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Blurred Vision; Visual Impairment
MAUDE report 20553086 ↗InjuryOct 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A CONSUMER REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, HE CANNOT READ PRINT WITHOUT READING GLASSES. HE HAS SPOKEN WITH DOCTOR REGARDING THIS, SHE PRESCRIBED READING GLASSES. HE WOULD LIKE COMPANY TO REACH OUT TO THE SURGEON FOR FURTHER INFORMATION OF HIS OUTCOME. ADDITIONAL INFORMATION RECEIVED, THE CAUSE OF EVENT WAS LIMITATION OF EXTENDED DEPTH OF FOCUS TECHNOLOGY. THERE WAS NO PATIENT HARM. THERE ARE TWO MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS ASSOCIATED WITH THE LEFT EYE.
Patient problems: Blurred Vision
MAUDE report 20552342 ↗InjuryJun 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE IOL WAS EXCHANGED FOR ANOTHER LENS MODEL FOUR MONTHS FOLLOWING THE INITIAL IMPLANT PROCEDURE. THE REASON FOR EXPLANT WAS NOT REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 19523307 ↗InjuryApr 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING THE INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED BLURRY VISION AND GHOSTING. THE WAS EXCHANGED FOR THE SAME LENS MODEL 3 MONTHS FOLLOWING THE INITIAL PROCEDURE. CLINICAL REASON FOR EXPLANT MENTIONED AS INTOLERABLE ANISOMETROPIA AND IMBALANCE BETWEEN EYES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Blurred Vision; Visual Disturbances
MAUDE report 19039764 ↗InjuryJan 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A NON-HEALTHCARE PROFESSIONAL REPORTED WITH DESCRIPTION EXPLANTED IOL WITH REASON LENS EXCHANGE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 18518459 ↗MalfunctionOct 10, 2023Scratched Material; Scratched Material
ADDITIONAL INFORMATION HAS BEEN RECEIVED INCLUDES PATIENT SYMPTOMS WERE RESOLVED AND THERE WAS NO PATIENT HARM.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17903259 ↗InjuryJul 3, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
ADDITIONAL INFORMATION WAS PROVIDED. ACCORDING TO THE SURGEON, THE OBJECT THAT WAS REMOVED WAS NOT IDENTIFIED AS A METAL ITEM. INSTEAD, IT WAS DESCRIBED AS A FOREIGN BODY THAT THE PATIENT MENTIONED AS BEING REFLECTIVE TO LIGHT DURING A POST-OPERATION EXAMINATION.
Patient problems: Insufficient Information
MAUDE report 17248862 ↗InjuryJun 26, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
A NON HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPLANT WITH COMPANY LENS AS THE PATIENT WAS UNHAPPY. THERE ARE TWO MEDICAL REPORTS ASSOCIATED WITH THIS REPORT. THIS IS 2 OF 2. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 17197541 ↗InjuryJun 12, 2023Defective Device; Defective Device
A NURSE REPORTED THAT AFTER IMPLANTATION OF INTRAOCULAR LENS (IOL), THE PATIENT HAD LENS COMPLICATIONS. THE LENS WAS EXPLANTED IN A SECONDARY PROCEDURE AND REPLACED WITH A MONOFOCAL LENS. THE CLINICAL REASON FOR EXPLANT WAS MYOPIC OUTCOME. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Visual Impairment
MAUDE report 17106492 ↗InjuryMay 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED REFRACTIVE SURPRISE AND INTOLERABLE ANISOMETROPIA . THE IOL WAS EXCHANGED FOR ANOTHER LENS MODEL 1 MONTHS FOLLOWING THE INITIAL IMPLANT PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Visual Impairment
MAUDE report 16982780 ↗InjuryMay 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED FUZZY VISION. THE IOL WAS EXCHANGED FOR UNSPECIFIED ATIOL LENS FOLLOWING THE INITIAL IMPLANT PROCEDURE. CLINICAL REASON FOR EXPLANT MENTIONED AS MECHANICAL COMPLICATION. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Blurred Vision
MAUDE report 16944288 ↗InjuryApr 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A CONSUMER REPORTED THAT AFTER IMPLANTATION OF INTRAOCULAR LENS (IOL), SINCE SURGERY, BARELY ABLE TO SEE OUT OF THE EYE. SHORTLY AFTER THE SURGERY, THE PHYSICIAN PERFORMED A YAG LASER. THE CONSUMER MENTIONED THAT THE PHYSICIAN HAS NOT TOLD THAT EYE WAS BEING TREATED WITH YAG, JUST MENTIONED THAT HE WAS CLEANING UP THE EYE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Blurred Vision
MAUDE report 16830487 ↗InjuryApr 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AN ACCOUNT MANAGER REPORTED FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED REFRACTIVE SURPRISE. THE LENS WAS EXPLANTED AND REPLACED WITH OTHER LENS IN A SECONDARY PROCEDURE. ANTERIOR CELL GROWTH WAS NOTED ON LENS PRIOR TO EXPLANT. THE EXPLANT WAS NOT DUE TO THE ANTERIOR CELL GROWTH. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Visual Impairment
MAUDE report 16677195 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page