Skip to content

Clareon CNW0T3 (UDI 00380652409674)Recalls & reported problems

Alcon Laboratories, Inc. · CLAREON BLF TORIC · Class 3 · Implant

All models of Clareon (1142)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00380652409674), received between Jun 18, 2024 and Aug 26, 2026 — classified as 4 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Device Damaged by Another Device2
  • Device Dislodged or Dislocated1
  • Failure to Eject1
  • Material Split, Cut or Torn1
  • Material Twisted/Bent1
  • Product Quality Problem1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Blurred Vision2
  • Visual Impairment2
  • Insufficient Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2026Product Quality Problem

    A non-healthcare professional reported that, during an intraocular lens (IOL) implantation procedure, a defect in the lens was detected. Additional information was received indicating that the lens was replaced with a company lens. There was no patient contact, and no further impact to the patient.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50449984 ↗
  • InjuryJul 30, 2026Adverse Event Without Identified Device or Use Problem

    A non health care professional reported that during an intraocular lens (IOL) implant procedure, the patient experienced visual disturbance, blurry vision. The IOL was explanted and exchanged for an unknown advanced technology intraocular lens (ATIOL) 2 months, 2 days following the initial implant procedure.Additional information received and the file was updated accordingly.

    Patient problems: Visual Impairment

    MAUDE report 50082144 ↗
  • MalfunctionMay 29, 2026Material Twisted/Bent

    A FACILITY REPRESENTATIVE REPORTED THAT THE BACK HAPTIC WOULD NOT FOLD. THE ISSUE WAS NOTED DURING LOADING AND THERE WAS NO PATIENT CONTACT. IN THE SURGEONS OPINION, BAD HAPTIC CAUSED OR CONTRIBUTED TO THE EVENT. PROCEDURE WAS COMPLETED ON THE SAME DAY. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25333656 ↗
  • MalfunctionApr 22, 2026Device Damaged by Another Device

    A HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE LENS WAS INSERTED INTO THE EYE WITH THE INJECTOR SET AS USUAL, AND THE IOL WAS CONFIRMED TO BE DAMAGED.

    Patient problems: Insufficient Information

    MAUDE report 24955526 ↗
  • InjuryDec 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED WITH UNABLE TO TOLERATE BLURRINESS. THE LENS WAS EXCHANGED FOR ANOTHER LENS MODEL 05 MONTHS 17 DAYS FOLLOWING THE INITIAL PROCEDURE. THE LENS WAS REPLACED WITH NONCOMPANY LENS. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Blurred Vision

    MAUDE report 23919240 ↗
  • InjuryAug 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, PATIENT WAS NOT ABLE TO READ. THE LENS WAS EXPLANTED AND EXCHANGED 4 MONTHS FOLLOWING THE INITIAL PROCEDURE. THE CLINICAL REASON FOR THE EXPLANT WAS MENTIONED AS MECHANICAL COMPLICATION. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Blurred Vision

    MAUDE report 22716837 ↗
  • InjuryApr 2, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ADDITIONAL INFORMATION WAS RECEIVED AND IT HAS BEEN DETERMINED THAT THERE WAS NO REPORTED DEFECT OR FAULT WITH A INTRAOCULAR LENS.

    Patient problems: Visual Impairment

    MAUDE report 21755746 ↗
  • MalfunctionOct 29, 2024Device Damaged by Another Device; Failure to Eject; Device Damaged by Another Device; Failure to Eject

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, WHILE LOADING THE LENS GOT STUCK IN CARTRIDGE AND CRACKED. THERE WAS NO PATIENT CONTACT. THE SURGERY WAS COMPLETED ON THE SAME DAY. AS PER SURGEON THE PRODUCT DID NOT CONTRIBUTE TO EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20558518 ↗
  • MalfunctionSep 21, 2024Scratched Material; Scratched Material

    A HEALTH CARE PROFESSIONAL REPORTED WITH THE DESCRIPTION OF, SCRATCH-LIKE OBSERVATION WAS FOUND ON THE OPTICAL PART. THERE WAS NO VISION PROBLEMS OR HEALTH PROBLEMS. THERE ARE SIX MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS REPORT. THIS IS 2 OF 6.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20275008 ↗
  • MalfunctionJun 18, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    A PHYSICIAN REPORTED THAT DURING IMPLANTATION OF INTRAOCULAR LENS (IOL), THE SURGEON NOTICED THE TEARS/CRACKS IN THE IOL AFTER IMPLANTATION AND THIS HAS TWO CRACKS IN THE PERIPHERY. THE IOL REMAINED IN THE EYE. NO COMPLICATIONS ARE EXPECTED. THERE WAS NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19559007 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates