Skip to content

Clareon CNW0T3 (UDI 00380652409704)Recalls & reported problems

Alcon Laboratories, Inc. · CLAREON BLF TORIC · Class 3 · Implant

All models of Clareon (1142)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00380652409704), received between Apr 2, 2024 and May 22, 2026 — classified as 3 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Break2
  • Failure to Advance1
  • Failure to Deliver1
  • Material Discolored1
  • Product Quality Problem1
  • Scratched Material1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions2
  • Blurred Vision1
  • Capsular Bag Tear1
  • Visual Impairment1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 22, 2026Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING THE CATARACT SURGERY WITH INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE THE PATIENT EXPERIENCED BLURRY VISION. THE IOL HAD BEEN EXCHANGED IN A SECONDARY PROCEDURE. CLINICAL REASON FOR EXPLANT WAS PATIENT DISSATISFACTION WITH IOL.

    Patient problems: Blurred Vision

    MAUDE report 25255967 ↗
  • MalfunctionMar 25, 2026Material Discolored

    A PHYSICIAN REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, INTRAOCULAR LENS (IOL) LOOKED BLUE BY THE SLIT LAMP. IT WAS ALSO STATED THAT THERE WAS NO SUBJECTIVE SYMPTOMS AND VISUAL ACUITY AND NO REMOVAL OF LENS WAS PLANNED.

    Patient problems: Insufficient Information

    MAUDE report 24684513 ↗
  • MalfunctionJul 30, 2025Failure to Deliver; Failure to Deliver

    A PHYSICIAN REPORTED THAT DURING A CATARACT SURGERY WITH INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, IMPLANTATION WAS STOPPED DUE TO STRANGE FELT DURING IMPLANTATION. THE SURGERY WAS COMPLETED AFTER REPLACING WITH ANOTHER UNSPECIFIED LENS.

    Patient problems: Insufficient Information

    MAUDE report 22666392 ↗
  • MalfunctionJul 8, 2025Break; Scratched Material; Break; Scratched Material

    A PHYSICIAN REPORTED THAT, WHILE PREPARING THE LENS FOR IMPLANTATION, THE OPTIC OF THE LENS WAS SCRATCHED AND THE HAPTIC WAS BROKEN. HENCE ANOTHER LENS WAS USED TO COMPLETE THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22434243 ↗
  • MalfunctionMar 20, 2025Product Quality Problem; Product Quality Problem

    NON-HEALTHCARE PROFESSIONAL DURING INTRAOCULAR (IOL) LENS IMPLANT PROCEDURE THAT THE IOL WAS DEFECTIVE AND IT WAS WASTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21652400 ↗
  • InjuryNov 28, 2024Break; Break

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, WHEN LENS WAS IMPLANTED AND BEING ROTATED INTO POSITION, THE PERIPHERAL CAPSULE WAS NOTED TO TEAR OPEN ALONG THE OPTIC EDGE ANTERIORLY. THE LENS WAS REMOVED AND REPLACEMENT SULCUS LENS WAS IMPLANTED. THERE WAS NO PATIENT IMPACT AND SYMPTOMS HAS BEEN RESOLVED. THE SURGEON SAID THERE WAS NO DIFFICULTY NOTED WITH LOADING THE IOL OR INSERTION OF THE IOL AND IT MAY OCCURRED AT ANY POINT OF TIME.

    Patient problems: Capsular Bag Tear

    MAUDE report 20805765 ↗
  • InjuryOct 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED AND REPORTED THAT THE IOL (INTRA OCULAR LENS) WAS EXCHANGED FOR THE SAME LENS MODEL BUT THREE DIOPTER LESS POWER INDICATING THE CORRECT LENS POWER WAS NOT IMPLANTED INITIALLY. THERE IS NO REPORTED DEFECT OR FAULT WITH THE IOL.

    Patient problems: Visual Impairment

    MAUDE report 20406849 ↗
  • MalfunctionApr 2, 2024Failure to Advance; Failure to Advance

    A PHARMACIST REPORTED THAT DURING SURGERY, AN INTRAOCULAR LENS WAS STUCK IN THE CARTRIDGE DURING THE OPERATION. THE BACKUP IMPLANT WAS USED TO OVERCOME THE PROBLEM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 19020156 ↗