Skip to content

Clareon CNW0T4 (UDI 00380652410168)Recalls & reported problems

Alcon Laboratories, Inc. · CLAREON BLF TORIC · Class 3 · Implant

All models of Clareon (1142)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00380652410168), received between Sep 28, 2021 and Aug 10, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Crack1
  • Material Split, Cut or Torn1
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Blurred Vision1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 10, 2026Adverse Event Without Identified Device or Use Problem

    A healthcare professional reported that, following intraocular lens implantation surgery, the patient experienced decreased visual acuity. Hence the lens was explanted and replaced with a non-company lens in a secondary procedure.Additional information has been requested but is not available at the time of this report.

    Patient problems: Blurred Vision

    MAUDE report 50193912 ↗
  • MalfunctionJun 5, 2026Break

    A NON HEALTH CARE PROFESSIONAL REPORTED DURING INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE HAPTICS HAD DETACHED FROM THE LENS, NECESSITATING THAT IT BE CUT AND REMOVED. ADDITIONAL INFORMATION HAS BEEN RECEIVED AND STATED THAT THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25410007 ↗
  • MalfunctionDec 12, 2025Product Quality Problem; Product Quality Problem

    A FACILITY REPRESENTATIVE REPORTED THAT DURING SURGERY, AN ATTEMPT WAS MADE TO IMPLANT THE LENS, BUT IT WAS DEFECTIVE / TORN. THERE WAS NO HARM TO THE PATIENT, AND A BACKUP UNIT WAS USED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23790729 ↗
  • MalfunctionNov 29, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    A FACILITY REPRESENTATIVE REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, LENS WAS CRACKED AFTER INSERTION. SUBSEQUENTLY THE LENS WAS REMOVED DURING INITIAL PROCEDURE AND COMPLETED WITH COMPANY ANOTHER LENS. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20809231 ↗
  • MalfunctionSep 28, 2021Crack; Crack

    A PHYSICIAN REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE IOL OPTIC GOT CRACKED FROM THE HAPTIC ROOT TO THE CENTER. THE SURGERY WAS COMPLETED AND THE LENS REMAINS IMPLANTED IN THE PATIENT'S EYE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12538740 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates