Skip to content

AcrySof™ IQ Vivity™ DFT515 (REF DFT515.235)Recalls & reported problems

Alcon Laboratories, Inc. · Sterile UV and Blue Light Filtering Acrylic Foldable Aspheric Wavefront-Shaping Toric Posterior Chamber Intraocular Lenses · Class 3 · Implant

All models of AcrySof IQ Vivity (225)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380652422536), received between Jun 1, 2021 and Jun 1, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Malposition of Device1

Patient problems

  • Capsular Bag Tear1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 1, 2021Malposition of Device; Malposition of Device

    A NURSE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE IOL BECAME DISLOCATED. THE IOL WAS REMOVED IN A SECONDARY PROCEDURE. ADDITIONAL INFORMATION WAS PROVIDED BY THE SURGEON THAT THE POSTERIOR CAPSULE APPEARED INTACT PRIOR TO INSERTING THE LENS, WHEN ROTATING THE LENS, A POSTERIOR TEAR WAS NOTED AND THE LENS FELL IN TO THE VITREOUS. THE EYE WAS CLOSED AND THE PATIENT WAS SENT TO A RETINA SPECIALIST FOR EVALUATION THE SAME DAY. THE IOL WAS EXCHANGED THE NEXT DAY. IN THE SURGEON'S OPINION, THE IOL DID NOT CAUSE THE EVENT. THE EVENT WAS CAUSED BY THE POSTERIOR CAPSULAR TEAR. THERE WAS NO PATIENT HARM, THE PATIENT'S ISSUES HAVE RESOLVED AND THE PROGNOSIS IS GOOD. THERE IS A MILD RESOLVING VITREOUS HEMORRHAGE.

    Patient problems: Capsular Bag Tear

    MAUDE report 11915146 ↗

Other Extended depth of focus intraocular lens devices

All Extended depth of focus intraocular lens devices →

Sources and updates