Skip to content

AcrySof™ IQ Vivity™ DAT015 (REF DAT015.220)Recalls & reported problems

Alcon Laboratories, Inc. · Sterile UV Absorbing Acrylic Foldable Aspheric Wavefront-Shaping Posterior Chamber Intraocular Lens · Class 3 · Implant

All models of AcrySof IQ Vivity (225)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Alcon AcrySof IQ Vivity intraocular lens is a foldable lens implanted during cataract surgery to give a wider range of clear vision. The 29 reports mostly describe patients who could not adapt to the lens or had blurred vision, distorted vision or difficulty seeing, with several lenses removed and replaced in a second surgery; a few mention a cracked or scratched lens or one that would not load into the injector. There are no deaths, 23 injuries and 6 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
23
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00380652424677), received between Mar 18, 2021 and Sep 12, 2024 — classified as 23 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Crack” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem18
  • Crack2
  • Defective Device2
  • Scratched Material2
  • Application Program Problem: Power Calculation Error1
  • Appropriate Device Problem Term/Code Not Available1
  • Contamination1
  • Defective Component1

Patient problems

  • Blurred Vision11
  • Visual Impairment11
  • Visual Disturbances10
  • No Clinical Signs, Symptoms or Conditions4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Insufficient Information3
  • Dry Eye(s)2
  • Eye Pain1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • InjurySep 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT HAD MALADAPTATION. THE IOL WAS EXCHANGED FOR UNSPECIFIED LENS MODEL AFTER THE INITIAL IMPLANT PROCEDURE. CLINICAL REASON FOR EXPLANT WAS REPORTED AS MALADAPTATION. THERE ARE TWO MEDICAL REPORTS ASSOCIATED WITH THIS PATIENT. THIS FILE BELONGS TO RIGHT EYE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Blurred Vision

    MAUDE report 20199793 ↗
  • InjuryApr 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED SUBJECTIVE VISUAL DISTURBANCE. THE IOL WAS EXCHANGED FOR ANOTHER UNSPECIFIED LENS MODEL APPROXIMATELY 8 MONTHS FOLLOWING THE INITIAL IMPLANT PROCEDURE. THE CLINICAL REASON FOR EXCHANGE WAS SUBJECTIVE VISUAL DISTURBANCE. ADDITIONAL INFORMATION HAS BEEN RECEIVED AND STATED THAT THE PATIENT WAS NOT ADJUSTING TO OPTIC OF IOL AND INABILITY TO ADAPT.

    Patient problems: Visual Impairment

    MAUDE report 19012530 ↗
  • InjuryJan 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT IS NOT SATISFIED WITH DISTANCE VISION AND PATIENT WAS UNABLE TO ADAPT TO THE TECHNOLOGY OF LENS. THE IOL WAS EXCHANGED FOR ANOTHER LENS MODEL FOLLOWING THE INITIAL IMPLANT PROCEDURE. ADDITIONAL INFORMATION RECEIVED AND STATED THAT THE PATIENT WAS NOT HOSPITALIZED FOR THE EVENT AND THERE WAS NO PATIENT HARM.

    Patient problems: Blurred Vision

    MAUDE report 18502083 ↗
  • InjuryOct 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON HEALTHCARE PROFESSIONAL REPORTED THAT, AFTER INTRAOCULAR LENS IMPLANTATION SURGERY, THE PATIENT DID NOT ADAPT TO THE LENS AND SAW DIAGONAL LINES. HENCE THE LENS WAS EXPLANTED AND REPLACED WITH MONOFOCAL LENS IN A SECONDARY PROCEDURE. THE PATIENT WAS NOT HOSPITALIZED AS A RESULT OF THIS EVENT AND THERE WAS NO PATIENT HARM.

    Patient problems: Visual Disturbances

    MAUDE report 17905099 ↗
  • MalfunctionOct 4, 2023Defective Component; Defective Component

    ADDITIONAL INFORMATION WAS REQUESTED AND RECEIVED STATING THAT WHEN TRYING TO LOAD THE LENS IN THE INJECTOR THE HAPTIC ON THE LENS WOULD NOT CURL AND WAS UNABLE TO BE LOADED CORRECTLY IN PREPARATION. THERE WAS NO PATIENT CONTACT WITH THE PRODUCT.

    Patient problems: Insufficient Information

    MAUDE report 17866993 ↗
  • MalfunctionApr 25, 2023Crack; Crack

    AN OPHTHALMIC SURGEON REPORTED THAT DURING A CATARACT EXTRACTION WITH AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE IOL WAS FOUND TO BE CRACKED. HENCE THE LENS WAS REMOVED DURING THE PROCEDURE AND REPLACED WITH ANOTHER LENS. ADDITIONAL INFORMATION WAS REQUESTED, BUT FURTHER NO INFORMATION WAS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 16808111 ↗
  • InjuryApr 18, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    A FACILITY REPRESENTATIVE REPORTED FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED DIFFICULTY SEEING TO DRIVE. THE LENS WAS EXPLANTED AND REPLACED WITH OTHER LENS IN A SECONDARY PROCEDURE. THE CLINICAL REASON FOR THE EXPLANT WAS MISSED TARGET, -0.75 RESIDUAL CORRECTION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Visual Impairment

    MAUDE report 16759901 ↗
  • InjuryMar 21, 2023Optical Decentration; Optical Decentration

    A MATERIALS MANAGER REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED DECENTERED POSTERIOR CAPSULE INTRA OCULAR LENS (PCIOL) AND A DISTORTED VISION. THE IOL WAS EXPLANTED AND EXCHANGED IN A SECONDARY PROCEDURE FOR MONOFOCAL LENS. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Visual Disturbances

    MAUDE report 16585202 ↗
  • InjuryAug 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED DOUBLE VISION. THE IOL WAS EXCHANGED FOR ANOTHER LENS MODEL 1 MONTH FOLLOWING THE INITIAL IMPLANT PROCEDURE. THE CLINICAL REASON FOR THE EXPLANT WAS DYSPHOTOPSIA SECONDARY TO MULTIFOCAL. ADDITIONAL INFORMATION HAS BEEN RECEIVED AND STATED THAT, AS PER THE SURGEONS OPINION IOL DID NOT CONTRIBUTED THE EVENT AND IT WAS DUE TO MONOCULAR DOUBLING AND SHADOWING. NO PATIENT HARM WAS REPORTED.

    Patient problems: Visual Impairment

    MAUDE report 15137405 ↗
  • InjuryJun 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NURSE REPORTED FOLLOWING THE INTRAOCULAR LENS IMPLANTATION THE PATIENT WAS SLIGHTLY UNHAPPY WITH THE VISION. THE LENS WAS EXCHANGED IN A SECONDARY PROCEDURE. THE CLINICAL REASON FOR EXPLANT WAS "SLIGHTLY MYOPIC". ADDITIONAL INFORMATION HAS BEEN RECEIVED THAT, THERE IS NO PATIENT HARM INVOLVED. THE PATIENT WAS NOT HOSPITALIZED.

    Patient problems: Visual Disturbances

    MAUDE report 14683269 ↗
  • InjuryApr 26, 2022Defective Device; Defective Device

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT STATES THAT, SHE FELT ISSUE WITH THE LENS (UNSPECIFIED) AS IT WAS NOT PERFORMING AS EXPECTED. THE PATIENT DEPTH OF PERCEPTION IS OFF, SHE IS UNABLE TO DRIVE COMFORTABLY. THE PATIENT HAS A SWEET SPOT AT ABOUT THREE FEET THAT SHE CAN SEE CLEARLY. THE PATIENT HAS A DARK SHADOW PERIPHERALLY (HALF MOON), THAT SHE FELT IT IS GETTING LARGER. THE PATIENT HAS SEEN A RETINAL SPECIALIST AND WAS TOLD EVERYTHING WAS FINE BUT DID HAVE MILD MACULAR DEGENERATION. THE PATIENT PHYSICIAN STATED, SHE NEEDS TO HAVE YTTRIUM ALUMINUM GARNET (YAG) BUT ONCE THAT IS DONE WOULD NOT BE ABLE TO HAVE ANY ADDITIONAL PROCEDURES. THE PATIENT¿S DOCTOR ALSO PERFORMED "BAG" PROCEDURE WITH A MEMBRANE THAT WAS PLACED ON HER CORNEA TO HELP HEALING, AFTER THE SURGERY IN HER LEFT EYE SHE HAD TO HAVE "TAP" DUE TO INCREASE PRESSURE. THE PATIENT USED ADDITIONAL DROPS THAT SAMPLES WERE PROVIDED TO HELP WITH INTRAOCULAR PRESSURE (IOP). THERE ARE TWO MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT, THIS REPORT IS FOR LEFT EYE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Visual Disturbances

    MAUDE report 14204557 ↗
  • InjuryApr 26, 2022Defective Device; Defective Device

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT STATES SHE FEELS UNSPECIFIED LENS ISSUE AS IT WAS NOT PERFORMING AS EXPECTED. HER DEPTH PERCEPTION IS OFF, SHE IS UNABLE TO DRIVE COMFORTABLY. SHE HAS A "SWEET" SPOT AT ABOUT THREE FEET THAT SHE CAN SEE CLEARLY. HAS SEEN A RETINAL SPECIALIST AND WAS TOLD EVERYTHING WAS FINE BUT DID HAVE MILD MACULAR DEGENERATION. HER PHYSICIAN STATED SHE NEEDS TO HAVE YAG BUT ONCE THAT IS DONE WOULD NOT BE ABLE TO HAVE ANY ADDITIONAL PROCEDURES. HER DOCTOR ALSO PERFORMED "BAG" PROCEDURE WITH A MEMBRANE THAT WAS PLACED ON HER CORNEA TO HELP HEALING. AFTER SURGERY IN HER LEFT EYE SHE HAD TO HAVE "TAP" DUE TO INCREASE PRESSURE. USED ADDITIONAL DROPS THAT SAMPLES WERE PROVIDED TO HELP WITH INTRAOCULAR PRESSURE (IOP). THERE ARE TWO MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS FOR RIGHT EYE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Visual Disturbances

    MAUDE report 14204183 ↗
  • InjuryApr 13, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED GLARE AND DECREASED VISION. THE IOL WAS EXCHANGED FOR ANOTHER LENS MODEL 5 MONTHS FOLLOWING THE INITIAL IMPLANT PROCEDURE. THE PATIENT WAS DISSATISFIED WITH THE IOL. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Blurred Vision; Visual Disturbances

    MAUDE report 14103050 ↗
  • MalfunctionMar 25, 2022Crack; Crack

    A FACILITY REPRESENTATIVE REPORTED THAT DURING INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE THE LENS WAS TORE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13904805 ↗
  • InjuryFeb 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A ACCOUNT MANAGER REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE THE PATIENT REPORTED DIFFICULTY SEEING HER FAMILY MEMBER'S FACES AND CANNOT SEE HER GRANDCHILDREN'S FACES AT THE SCHOOL PROGRAMS. SHE WAS NOT HAPPY WITH HER VISION AT NIGHT AND UNABLE TO DRIVE DUE TO THE GLARE. THERE WAS NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Visual Disturbances; Visual Impairment

    MAUDE report 13620525 ↗
  • InjuryFeb 24, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED BLURRED VISION. THE PATIENT INFORMED THIS TO HIS OPERATING PHYSICIAN (OP). THE PHYSICIAN KEPT SAYING THAT, IT WAS DRY EYE, BUT THE EYE DROPS DIDN'T HELP AT ALL. BEFORE THE SURGERY, THE PATIENT NEEDED CHEATER TO SEE, HOWEVER NOW THE PATIENT CAN'T SEE THE BIG SIGNS ON THE ROAD. ADDITIONAL INFORMATION RECEIVED AND STATED THAT HAVING TROUBLE IN DISTANCE VISION IN RIGHT EYE, READING IS STILL DIFFICULT, AND THERE IS A NEED FOR GLASSES, STILL SUFFERING FROM DRY EYES.

    Patient problems: Blurred Vision; Dry Eye(s); Visual Impairment

    MAUDE report 13596507 ↗
  • InjuryFeb 17, 2022Application Program Problem: Power Calculation Error; Application Program Problem: Power Calculation Error

    A NON HEALTHCARE PROFESSIONAL REPORTED FOLLOWING AN INTRAOCULAR LENS IMPLANT PROCEDURE, THE LENS WAS AGAIN EXCHANGED IN A SECONDARY PROCEDURE DUE TO INCORRECT LENS POWER. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 13549665 ↗
  • InjuryDec 17, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A HEALTH CARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED BLURRED VISION. THE CLINICAL REASON WAS REFRACTIVE SURPRISE. THE IOL WAS EXCHANGED IN A SECONDARY PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Blurred Vision; Visual Disturbances

    MAUDE report 13026600 ↗
  • MalfunctionOct 6, 2021Scratched Material; Scratched Material

    AN ACCOUNT MANAGER REPORTED THAT, DURING INTRAOCULAR LENS (IOL) IMPLANTATION, IT WAS NOTICED THAT THE LENS HAD A LINE/MARK THAT WAS NOT CAUSED BY FORCEPS DURING THE LOADING PROCESS. THEN LENS WAS CUT OUT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12584602 ↗
  • InjurySep 16, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A SURGEON HAS REPORTED THAT FOLLOWING A CATARACT EXTRACTION WITH INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT COMPLAINS OF POOR DISTANCE VISION. ADDITIONAL INFORMATION WAS PROVIDED BY THE SURGEON THAT APPROXIMATELY ONE MONTH FOLLOWING HER CATARACT EXTRACTION WITH IOL IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED SLIGHT INFLAMMATION AND LIGHT SENSITIVITY. HER POSTOPERATIVE STEROID DROP WAS INCREASED. THE SURGEON FELT THAT THE VISION WAS REDUCED DUE TO DRYNESS. THE PATIENT WAS GIVEN ARTIFICIAL TEARS AND AN ANTI-INFLAMMATORY TOPICAL MEDICATION. THE PATIENT RETURNED 6 WEEKS FOLLOWING THE PROCEDURE, COMPLAINING OF DOUBLE VISION, PAIN, AND POOR VISION. THIS IS GRADUALLY WORSENING. THE SYMPTOMS SEEM TO BE WORSE IN THE SUNSHINE. THESE PROBLEMS HAVE BEEN FAIRLY STABLE SINCE THE ONSET AND HAVE BEEN PERSISTENT. SHE ALSO REPORTS PAIN IN HER FACE AND AROUND HER EYES/CHEEK AREAS WHICH BECOME WORSE BY THE END OF THE DAY. THE VISION IS STILL VERY BLURRY FAR AWAY, BUT OKAY AT NEAR/INTERMEDIATE. SHE IS UNABLE TO READ STREET SIGNS AND FEELS UNSAFE TO DRIVE AT THE MOMENT. THE PATIENT HAS A HISTORY OF POSTOPERATIVE IRITIS WHICH IS IMPROVING. POSTERIOR CAPSULE OPACIFICATION WAS NOTED ON EXAMI

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Dry Eye(s); Eye Pain; Inflammation; Iritis; Visual Disturbances; Visual Impairment

    MAUDE report 12479743 ↗
  • MalfunctionSep 8, 2021Contamination; Contamination

    A FACILITY REPRESENTATIVE REPORTED AN INTRAOCULAR LENS (IOL) WITH A DESCRIPTION OF SOMETHING INSIDE THE IOL. THE LENS WAS PLACED IN THE EYE AND NOTICED SOMETHING ABNORMAL INSIDE THE LENS. THE SURGEON RINSED THE EYE WITHOUT REMOVAL OF OBJECT AND THEN REMOVED THE LENS FROM THE PATIENT'S EYE. PROCEDURE WAS COMPLETED THE SAME DAY WITH A NEW LENS. THE SURGEON FELT AS IF THE LENS HAD A FOREIGN OBJECT IN IT AND DID CAUSE THE EVENT TO OCCUR. NO PATIENT HARM WAS REPORTED AND NO ISSUES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12438330 ↗
  • InjurySep 7, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NURSE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT¿S VISION WAS DECREASED. THE IOL WAS EXCHANGED FOR ANOTHER LENS IN A SECONDARY PROCEDURE 105 DAYS FOLLOWING THE INITIAL PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED; HOWEVER, FURTHER INFORMATION HAS NOT BEEN RECEIVED.

    Patient problems: Visual Impairment

    MAUDE report 12435211 ↗
  • InjurySep 6, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NURSE REPORTED THAT POST IMPLANTATION OF AN INTRAOCULAR LENS (IOL) THE PATIENT EXPERIENCED DECREASED VISION. THE IOL WAS REPLACED WITH ANOTHER LENS 79 DAYS FOLLOWING THE INITIAL PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Visual Impairment

    MAUDE report 12428004 ↗
  • InjuryAug 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT, FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT WAS UNHAPPY WITH DISTANCE VISUAL ACUITY. THE IOL WAS REPLACED WITH ANOTHER IOL. THE CLINICAL REASON FOR EXPLANT WAS GIVEN AS RESIDUAL MIXTURE. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Visual Impairment

    MAUDE report 12352004 ↗
  • InjuryJun 18, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A PHYSICIAN REPORTED VIA EMAIL, THAT CONSUMER EXPERIENCED BLURRY VISION AND INTOLERANCE TO TECHNOLOGY, THAT LEAD TO THE EXPLANTATION OF THE INTRAOCULAR LENS. ADDITIONAL INFO HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION OBTAINED.

    Patient problems: Blurred Vision

    MAUDE report 12024661 ↗

FDA clearance and classification

The FDA regulates product code POE (extended depth of focus intraocular lens) under 21 CFR 886.3600 as a Class III device. 21 CFR 886.3600 ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Intended to be implanted to replace the natural lens of an eye.

Common questions about the AcrySof™ IQ Vivity™ DAT015

Has the AcrySof™ IQ Vivity™ DAT015 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380652424677), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the AcrySof™ IQ Vivity™ DAT015?

The FDA's MAUDE database lists 29 adverse event reports for this device, received between Mar 18, 2021 and Sep 12, 2024. In those reports, the most common specific device problems are “Crack” (2), “Defective Device” (2) and “Scratched Material” (2); the most common patient problems are “Blurred Vision” (11), “Visual Impairment” (11) and “Visual Disturbances” (10). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AcrySof™ IQ Vivity™ DAT015?

None of the reports are classified as deaths, 23 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AcrySof™ IQ Vivity™ DAT015 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the AcrySof™ IQ Vivity™ DAT015, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand AcrySof™ IQ Vivity™, model DAT015). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Extended depth of focus intraocular lens devices

All Extended depth of focus intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page