Skip to content

Clareon CCWTT3 (UDI 00380652457460)Recalls & reported problems

Alcon Laboratories, Inc. · Clareon PanOptix Toric UVA · Class 3 · Implant

All models of Clareon (1142)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00380652457460), received between Jan 11, 2023 and Oct 22, 2025 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Device Contamination with Chemical or Other Material2

Patient problems

  • Blurred Vision3
  • Insufficient Information2
  • Halo1
  • Visual Disturbances1
  • Visual Impairment1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT REASON AS IOL CAME TOO STRONG AND CLINICAL REASON FOR EXPLANT BEING BLURRY VISION IOL CAME TOO STRONG. THE IOL WAS EXPLANTED AND EXCHANGED FOR AN UNKNOWN ADVANCED TECHNOLOGY INTRAOCULAR LENS FOLLOWING THE INITIAL IMPLANT PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Blurred Vision

    MAUDE report 23356034 ↗
  • InjuryJan 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON HEALTH CARE PROFESSIONAL REPORTED THAT AFTER SURGERY, THE PATIENT WAS UNABLE TO DRIVE AT NIGHT 2 HALOS. CLINICAL REASON BEING MENTIONED AS UNABLE TO TOLERATE GLARE, HALOS, QUALITY OF VISION. THE IOL(INTRAOCULAR LENS) WAS EXPLANTED AND EXCHANGED FOR AN UNKNOWN MONOFOCAL LENS IN THE SECONDARY IMPLANT PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Halo; Visual Disturbances; Visual Impairment

    MAUDE report 21055681 ↗
  • InjurySep 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT HAD MALADAPTATION. THE IOL WAS EXCHANGED FOR UNSPECIFIED LENS MODEL 2 MONTHS AND 10 DAYS AFTER THE INITIAL IMPLANT PROCEDURE. CLINICAL REASON FOR EXPLANT WAS REPORTED AS MALADAPTATION. THERE ARE TWO MEDICAL REPORTS ASSOCIATED WITH THIS PATIENT. THIS FILE BELONGS TO LEFT EYE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Blurred Vision

    MAUDE report 20134438 ↗
  • InjurySep 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON HEALTHCARE PROFESSIONAL REPORTED THAT, AFTER INTRAOCULAR LENS IMPLANTATION SURGERY, THE PATIENT EXPERIENCED BLURRY VISION. HENCE THE LENS WAS EXPLANTED AND REPLACED WITH ANOTHER LENS IN A SECONDARY PROCEDURE. THE CLINICAL REASON FOR EXPLANT WAS BLURRY VISION. ADDITIONAL INFORMATION RECEIVED AND REPORTED THAT THE IOL WAS EXCHANGED FORM (24.5) DIOPTER TO (21.5) DIOPTER LENS OF SAME MODEL INDICATING THE CORRECT LENS POWER WAS NOT IMPLANTED INITIALLY. THERE IS NO REPORTED DEFECT OR FAULT WITH THE IOL.

    Patient problems: Blurred Vision

    MAUDE report 17789947 ↗
  • MalfunctionMar 28, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    A PHYSICIAN REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE SURGEON NOTICED THAT ON THE LENS IT APPEARED TO BE A CRACK ON THE POSTERIOR SIDE OF A IOL. WHEN SURGEON EXAMINED FURTHER AND ALTHOUGH IT WAS A STRAIGHT LINE RESEMBLING A SCRATCH, HE COULD ACTUALLY PEEL BACK THE EDGE OF THE LINE. HE COULD NOT REMOVE WITH SUCTION FROM I/A (IRRIGATION AND ASPIRATION), AND IT TOOK HIM ABOUT 5 MINUTES TO PEEL IT. AFTER HE PEELED IT OFF THE BACK OF THE LENS, IT WAS EASILY SUCTIONED OUT BY I/A WITH NO DAMAGE VISIBLE IN ANY WAY ON THE OPTIC/IOL.

    Patient problems: Insufficient Information

    MAUDE report 16625936 ↗
  • MalfunctionJan 11, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    A PHYSICIAN REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE SURGEON NOTICED THAT ON THE LENS IT APPEARED TO BE A CRACK ON THE POSTERIOR SIDE OF A IOL. WHEN SURGEON EXAMINED FURTHER AND ALTHOUGH IT WAS A STRAIGHT LINE RESEMBLING A SCRATCH, HE COULD ACTUALLY PEEL BACK THE EDGE OF THE LINE. HE COULD NOT REMOVE WITH SUCTION FROM I/A (IRRIGATION AND ASPIRATION), AND IT TOOK HIM ABOUT 5 MINUTES TO PEEL IT. AFTER HE PEELED IT OFF THE BACK OF THE LENS, IT WAS EASILY SUCTIONED OUT BY I/A WITH NO DAMAGE VISIBLE IN ANY WAY ON THE OPTIC/IOL.

    Patient problems: Insufficient Information

    MAUDE report 16136049 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page