Clareon CNWET3 (UDI 00380652466295)Recalls & reported problems
Alcon Laboratories, Inc. · Clareon Vivity Toric · Class 3 · Implant
All models of Clareon (1142)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380652466295), received between Apr 26, 2023 and May 18, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Deliver1
- Failure to Fold1
- Malposition of Device1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Halo1
- Visual Disturbances1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 18, 2026Failure to Fold; Failure to Deliver
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, IT WAS NOTICED THAT THE LENS FOLDED INCORRECTLY, GOT CAUGHT ON PLUNGER AND WAS NOT ABLE TO BE INSERTED PROPERLY. SUSPECTED LENS DAMAGE AS A RESULT. OPENED A NEW IOL AND FINISHED THE CASE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25200607 ↗MalfunctionDec 12, 2024Scratched Material; Scratched Material
A HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, IT WAS NOTED THAT THE LENS HAD MARK WHEN IMPLANTED. THE IOL WAS EXPLANTED DURING INITIAL PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20917245 ↗InjuryApr 26, 2023Malposition of Device; Malposition of Device
A FACILITY REPRESENTATIVE REPORTED FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED NIGHTTIME GLARE AND HALOS . THE LENS WAS EXPLANTED AND REPLACED WITH OTHER LENS IN A SECONDARY PROCEDURE. THE CLINICAL REASON FOR THE EXPLANT WAS LENS SLIGHTLY ROTATED . ADDITIONAL INFORMATION WAS REQUESTED AND RECEIVED STATING THE INTOLERANCE OF EDOF OPTICS DUE TO LARGER SCOTOPIC PUPIL SIZE WAS THE CAUSE OF THE EVENT.
Patient problems: Halo; Visual Disturbances
MAUDE report 16815012 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗