Clareon CNWET3 (UDI 00380652466301)Recalls & reported problems
Alcon Laboratories, Inc. · Clareon Vivity Toric · Class 3 · Implant
All models of Clareon (1142)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380652466301), received between Feb 6, 2023 and Apr 3, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Malposition of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Malposition of Device2
Patient problems
- Visual Disturbances1
- Visual Impairment1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 3, 2023Malposition of Device; Malposition of Device
A FACILITY REPRESENTATIVE REPORTED FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE LENS EXPLANTED DUE TO MILD ROTATION OF INTRAOCULAR LENS. THE LENS WAS EXPLANTED AND REPLACED WITH A MONOFOCAL LENS IN A SECONDARY PROCEDURE. THE PATIENT REASON FOR THE EXPLANT WAS GHOSTING/SHADOW VISION. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Visual Disturbances
MAUDE report 16664882 ↗MalfunctionFeb 6, 2023Malposition of Device; Malposition of Device
A PHYSICIAN REPORTED THAT AFTER INTRAOCULAR LENS (IOL) IMPLANTATION, THERE WAS LENS ROTATION CAUSING POOR VISION IN THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Visual Impairment
MAUDE report 16303628 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page