Skip to content

Clareon™ CNWTT3 (UDI 00380652468558)Recalls & reported problems

Alcon Laboratories, Inc. · STERILE UV and Blue Light Filtering Hydrophobic Acrylic Foldable Diffractive Aspheric Toric Posterior Chamber Intraocular Lens · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00380652468558), received between Jul 11, 2022 and Mar 11, 2026 — classified as 6 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Scratched Material” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem6
  • Scratched Material2
  • Device Damaged by Another Device1
  • Device Dislodged or Dislocated1
  • Failure to Unfold or Unwrap1
  • Material Split, Cut or Torn1

Patient problems

  • Visual Impairment4
  • No Clinical Signs, Symptoms or Conditions3
  • Visual Disturbances3
  • Blurred Vision2
  • Insufficient Information2
  • Halo1
  • Headache1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2026Failure to Unfold or Unwrap

    A NON-HEALTH CARE PROFESSIONAL REPORTED WITH A DESCRIPTION, THE HAPTIC REMAINS GLUED TO THE OPTIC AND NO PATIENT IMPACT WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED AND STATED, THE HAPTIC REMAIN STUCK TO THE OPTIC INSIDE THE PATIENT EYE. THE LENS WAS EXPLANTED AND RE-IMPLANTED IN THE SAME PROCEDURE. ADDITIONAL INFORMATION WAS RECEIVED AND STATED, ANTERIOR HAPTIC WAS STUCK TO THE OPTIC. A MANUAL DEPLOYMENT WAS ATTEMPTED BEFORE EXPLANTATION AND NO ABNORMAL RESISTANCE WAS ENCOUNTERED DURING INSERTION. THE SURGERY WAS COMPLETED DURING THE SAME PROCEDURE WITH BACK UP LENS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24568468 ↗
  • InjuryJan 23, 2026Adverse Event Without Identified Device or Use Problem

    A NONHEALTH CARE PROFESSIONAL REPORTED THAT AFTER THE INTRAOCULAR LENS (IOL) IMPLANTATION SURGERY, THE PATIENT EXPERIENCED HALOS AND COULD NOT DRIVE AT NIGHT. THE HALOS WAS VERY BAD AROUND THE LIGHTS THAT HE COULD NOT DRIVE AT NIGHT. THERE ARE TWO FILES ASSOCIATED WITH THIS REPORT. THIS IS 1 OF 2.

    Patient problems: Headache; Visual Disturbances

    MAUDE report 24158659 ↗
  • MalfunctionNov 27, 2025Scratched Material; Scratched Material

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE OPTIC WAS DAMAGED. THE PROCEDURE WAS COMPLETED ON THE SAME DAY. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND RECEIVED STATING THAT THE LENS HAD BEEN REMOVED DURING INITIAL PROCEDURE (IN AND OUT) AND REPLACED WITH ANOTHER BACKUP LENS.

    Patient problems: Insufficient Information

    MAUDE report 23666074 ↗
  • MalfunctionOct 29, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A HEALTH CARE PROFESSIONAL REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, SURGEON NOTICED LENS OPTIC CRACKED WHEN THE LENS WAS PUSHED TO THE EDGE OF CARTRIDGE. THE SURGERY WAS COMPLETED ON THE SAME DAY AND THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23413984 ↗
  • MalfunctionOct 15, 2025Device Damaged by Another Device; Device Damaged by Another Device

    A NURSE REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE, THE LENS WAS SCRATCHED BY THE INJECTOR. ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE SCRATCH WAS IDENTIFIED BY THE SURGEON UNDER THE MICROSCOPE DURING LENS INSERTION. THE LENS WAS NOT IMPLANTED BUT REMOVED BEFORE FULL INSERTION. NO HARM WAS CAUSED TO THE PATIENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23299515 ↗
  • MalfunctionMay 27, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A PHYSICIAN REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE,RESIDUAL REFRACTIVE ERROR, TORIC WAS LOCATED AT 10 DEGREES (ORIGINALLY CALCULATED AXIS WAS 172). PATIENT WAS HAVING ASTIGMATISM AND IT RESOLVED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Visual Impairment

    MAUDE report 22088392 ↗
  • InjuryOct 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING THE INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT WAS UNHAPPY WITH THE REFRACTION OUTCOME. THE LENS WAS EXPLANTED FOLLOWING THE INITIAL PROCEDURE AND REPLACED WITH ANOTHER LENS.

    Patient problems: Visual Impairment

    MAUDE report 20520661 ↗
  • InjuryJun 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED BLURRY VISIO AND UNABLE TO READ STREET SIGNS. THE LENS WAS EXPLANTED AND REPLACED WITH MONOFOCAL LENS IN A SECONDARY PROCEDURE. THE CLINICAL REASON FOR THE EXPLANT WAS MECHANICAL COMPLICATION. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Blurred Vision

    MAUDE report 19521048 ↗
  • InjuryApr 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A HEALTH CARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED PERSISTENT GLARE AND HALOS. YAG LASER SURGERY WAS PERFORMED BUT HAS NOT HELPED. HEALTH CARE PROFESSIONAL REPORTED THE PATIENT BELIEVED THAT THE QUALITY OF LIFE HAS BEEN SEVERELY IMPACTED WITH NO HARM. THE SURGEON¿S PROGNOSIS INCLUDED PATIENT WAS STABLE. THERE ARE TWO MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS ASSOCIATED WITH THE LEFT EYE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Halo; Visual Disturbances

    MAUDE report 19213475 ↗
  • InjuryDec 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT AFTER AN INTRAOCULAR LENS (IOL) IMPLANTATION SURGERY, PATIENT HAD BLURRED VISUAL ACUITY. CLINICAL REASON WAS MENTIONED AS MYOPIC OUTCOME. THE IOL WAS EXCHANGED APPROXIMATELY 7 MONTHS FOLLOWING THE INITIAL IMPLANT PROCEDURE.

    Patient problems: Blurred Vision; Visual Impairment

    MAUDE report 18289877 ↗
  • InjuryAug 29, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FACILITY REPRESENTATIVE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT WAS UNABLE TO ADAPT TO THE IMPLANT AND COMPLAINED OF GLARE. THE PATIENT'S CORRECTED VISION WAS 0.9, BUT THEIR VISION DID NOT RECOVER AND REMAINED BELOW 0.5. THE IOL WAS EXCHANGED FOR COMPANY MONOFOCAL LENS. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Visual Disturbances; Visual Impairment

    MAUDE report 17641909 ↗
  • MalfunctionJul 11, 2022Scratched Material; Scratched Material

    A HEALTH CARE PROFESSIONAL REPORTED THAT DURING A CATARACT EXTRACTION WITH INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE LENS HAD A SCRATCH. THERE WAS A PATIENT CONTACT, AND THE SURGERY WAS COMPLETED THE SAME DAY. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND RECEIVED STATED THAT ISSUE WITH LENS WAS NOTED AFTER THE LENS WAS INSERTED. THE PATIENT RECEIVED THE REPLACEMENT LENS. THE SURGEON WAS NOT SURE HOW SCRATCH HAS OCCURRED. THERE WAS A PATIENT CONTACT. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 14977899 ↗

Other Lens, multifocal intraocular devices

All Lens, multifocal intraocular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page