Skip to content

Multiflex MTA4U0 (REF MTA4U0.190)Recalls & reported problems

Alcon Laboratories, Inc. · MULTIFLEX™*III SINGLE-PIECE PMMA STERILE ACL (IOL/AC), 13.0mm LENGTH, 5.5mm CONVEXOPLANO OPTIC, 0.5mm VAULTED HAPTICS. *REG. U.S. PAT. & TM OFF. · Class 3 · Implant

All models of Multiflex (28)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380652535878), received between Nov 10, 2022 and Sep 15, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Defective Device1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Material Split, Cut or Torn

    A non-healthcare professional reported during intraocular lens (IOL) implant procedure, the surgeon noticed lens broke half in the patient's eye after the IOL was implanted into the eye. The lens was explanted during initial procedure.Additional information has been requested and received stating that the procedure was completed on same day by using same model and diopter lens.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50770936 ↗
  • MalfunctionApr 24, 2025Break; Break

    A FACILITY REPRESENTATIVE REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, AFTER INSERTION WHILE TRYING TO REPOSITION THE LENS THE HAPTIC BROKE OFF. SUBSEQUENTLY THE LENS WAS REMOVED DURING INITIAL PROCEDURE AND COMPLETED WITH ANOTHER LENS. ADDITIONAL INFORMATION RECEIVED STATING THAT THE PROCEDURE WAS COMPLETED ON SAME DAY. THERE WAS NO PATIENT HARM. IN THE SURGEON¿S OPINION, REPOSITIONING LENS CONTRIBUTED TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21906141 ↗
  • MalfunctionNov 10, 2022Defective Device; Defective Device

    A MATERIALS MANAGER REPORTED THAT DURING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE LENS WAS ATTEMPTED TO BE IMPLANTED AND WOULD NOT FIT IN THE EYE, SO IT WAS EXPLANTED AND THE PHYSICIAN WENT WITH THE DIFFERENT SIZE LENS. THERE WAS A PATIENT CONTACT. THE LENS WAS TOO BIG FOR THE PATIENT EYE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 15770726 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page