Skip to content

ClearCut®; Intrepid 8065772265Recalls & reported problems

Alcon Laboratories, Inc. · ClearCut® S Intrepid Micro-coaxial system Safety Slit Knife 2.2 DB · Class 1

All models of ClearCut; Intrepid (9)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00380657722655), received between Sep 6, 2023 and Sep 26, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Dull, Blunt” (9 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 15, 2024Class IIOpen, Classified

    Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.

    Z-2651-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Dull, Blunt9

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 26, 2025Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT OPHTHALMIC KNIVES FROM FOUR DIFFERENT LOT NUMBERS WERE FOUND TO BE BLUNT DURING A CATARACT SURGERY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED ON THE SAME DAY BY USING A REPLACEMENT KNIFE, AND THERE WAS NO IMPACT TO THE PATIENT. NO FURTHER INFORMATION IS EXPECTED, AS THE CUSTOMER WAS UNWILLING TO PROVIDE ADDITIONAL DETAILS. THIS REPORT PERTAINS TO THE THIRD CASE, INVOLVING ONE UNKNOWN LOT NUMBERS. A TOTAL OF 15 BLUNT KNIVES WERE REPORTED ACROSS THREE CASES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23152992 ↗
  • MalfunctionJul 15, 2025Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT DURING THE CATARACT SURGERY, THREE OPHTHALMIC KNIVES WERE FOUND TO BE DULL. TWO FROM THE SAME LOT AND ONE FROM A DIFFERENT LOT. THE SURGERY WAS COMPLETED ON THE SAME DAY USING AN ALTERNATIVE BLADE. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22505675 ↗
  • MalfunctionJul 15, 2025Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT DURING THE CATARACT SURGERY, THREE OPHTHALMIC KNIVES WERE FOUND TO BE DULL. TWO FROM THE SAME LOT AND ONE FROM A DIFFERENT LOT. THE SURGERY WAS COMPLETED ON THE SAME DAY USING AN ALTERNATIVE BLADE. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22505674 ↗
  • MalfunctionFeb 26, 2025Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT AN FIVE OPHTHALMIC BLADES FROM THE SAME LOT WERE FOUND TO BE DULL DURING SURGERY. THE PROCEDURE WAS COMPLETED AFTER OPENING A NEW BLADE WITHOUT ANY PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21468491 ↗
  • MalfunctionFeb 20, 2025Dull, Blunt; Dull, Blunt

    A PHYSICIAN REPORTED THAT DURING A CATARACT SURGERY WITH INTRA OCULAR LENS (IOL) IMPLANT, OPHTHALMIC KNIVES WERE FOUND TO BE BLUNT. THE SURGERY WAS COMPLETED BY BACKUP KNIFE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21421956 ↗
  • MalfunctionJan 20, 2025Dull, Blunt; Dull, Blunt

    A PHYSICIAN REPORTED THAT DURING CATARACT SURGERY WHILE PREPARING THE MAIN CUT IN THE RIGHT EYE OF THE PATIENT, INSUFFICIENT SHARPNESS OF AN OPHTHALMIC SURGICAL BLADE WAS OBSERVED AND THE PUNCTURE COULD NOT BE PERFORMED. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING WITH ANOTHER BLADE. NO PATIENT IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21190127 ↗
  • MalfunctionAug 9, 2024Dull, Blunt; Dull, Blunt

    A PHARMACIST REPORTED THAT TIP OF AN OPHTHALMIC SURGICAL SCALPEL WAS CROOKED AND MAKING IT IMPOSSIBLE TO USE BEFORE CATARACT SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING WITH ANOTHER PRODUCT. NO PATIENT IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19949163 ↗
  • MalfunctionOct 6, 2023Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT AN OPHTHALMIC KNIFE WAS FOUND TO BE DULL DURING THE CATARACT SURGERY. THE SURGERY WAS COMPLETED BY THE ALTERNATIVE KNIFE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17886199 ↗
  • MalfunctionSep 6, 2023Dull, Blunt; Dull, Blunt

    A CUSTOMER REPORTED THAT AN OPTHALMIC OPERATING SURGICAL KNIFES WERE DULL. PROCEDURE AND PATIENT IMPACT DETAILS WERE NOT REPORTED. ADDITIONAL INFORMATION REQUESTED NONE RECEIVED TILL DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17696069 ↗

FDA clearance and classification

The FDA regulates product code HNN (knife, ophthalmic) under 21 CFR 886.4350 as a Class I device. 21 CFR 886.4350 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ClearCut®; Intrepid 8065772265

Has the ClearCut®; Intrepid 8065772265 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2651-2024 (Class II), began on Jul 15, 2024: Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ClearCut®; Intrepid 8065772265?

The FDA's MAUDE database lists 9 adverse event reports for this device, received between Sep 6, 2023 and Sep 26, 2025. In those reports, the most common specific device problem is “Dull, Blunt” (9). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ClearCut®; Intrepid 8065772265?

None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ClearCut®; Intrepid 8065772265 FDA approved?

Devices of this type (product code HNN) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Alcon Laboratories, Inc.

Who makes the ClearCut®; Intrepid 8065772265, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand ClearCut®; Intrepid, model 8065772265). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Knife, ophthalmic devices

All Knife, ophthalmic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page