Skip to content

ClearCut®; Intrepid 8065772445Recalls & reported problems

Alcon Laboratories, Inc. · ClearCut® S Intrepid Micro-coaxial system Safety Slit Knife 2.4 SB · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00380657724451), received between Mar 25, 2021 and Oct 23, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Dull, Blunt” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 15, 2024Open, Classified

    Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.

    Z-2652-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Dull, Blunt4
  • Material Fragmentation1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2025Dull, Blunt; Dull, Blunt

    A HEALTHCARE PROFESSIONAL REPORTED THAT AN OPHTHALMIC CUTTING BLADE WAS FOUND DULL. PROCEDURE DETAILS AND PATIENT IMPACT WERE NOT REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND NONE RECEIVED TILL DATE.

    Patient problems: Insufficient Information

    MAUDE report 23367611 ↗
  • MalfunctionAug 8, 2025Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT OPHTHALMIC SURGICAL BLADES WERE NOTED TO BE DULL. PROCEDURE DETAILS AND PATIENT DETAILS WERE NOT REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 22748426 ↗
  • MalfunctionSep 17, 2024Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT DURING CATARACT SURGERY AN OPHTHALMIC SURGICAL KNIFE WAS DULL. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING WITH ANOTHER KNIFE. NO PATIENT IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20234890 ↗
  • MalfunctionJun 25, 2024Dull, Blunt; Dull, Blunt

    A HEALTHCARE PROFESSIONAL REPORTED THAT AN OPHTHALMIC KNIVES WERE FOUND DULL. PROCEDURE DETAILS AND PATIENT IMPACT WERE NOT REPORTED. ADDITIONAL INFORMATION HAS BEN REQUESTED AND NONE RECEIVED TILL DATE.

    Patient problems: Insufficient Information

    MAUDE report 19603784 ↗
  • MalfunctionMar 25, 2021Material Fragmentation; Material Fragmentation

    A PHYSICIAN REPORTED THAT OPHTHALMIC KNIVES HAVE REPEATEDLY LEFT RESIDUAL METAL PARTICLES IN EYE WOUND SITES DURING SEPARATE CATARACT SURGERIES. THERE WAS NO REPORT OF ANY PATIENT HARM. ADDITIONAL INFORMATION IS NOT AVAILABLE FOR THIS REPORT.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11568292 ↗

Other Knife, ophthalmic devices

All Knife, ophthalmic devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.