Skip to content

Clareon PanOptix Pro PXCAT3 (REF PXCAT3.135)Recalls & reported problems

Alcon Laboratories, Inc. · CLRN PO PRO TORIC ATME UVA · Class 3 · Implant

All models of Clareon PanOptix Pro (222)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380658491192), received between May 20, 2026 and Jun 19, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Scratched Material” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Scratched Material1

Patient problems

  • Blurred Vision1
  • Visual Disturbances1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 19, 2026Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED BLURRY VISION AT ALL RANGES AND SHADOWING OF LETTERS WHEN TRYING TO READ. CLINICAL REASON BEING MENTIONED AS MECHANICAL COMPLICATION OF IOL. THE IOL WAS EXPLANTED AND EXCHANGED WITH AN UNSPECIFIED ATIOL (ADVANCED TECHNOLOGY INTRAOCULAR LENS) IN THE SECONDARY IMPLANT PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Blurred Vision

    MAUDE report 25558215 ↗
  • MalfunctionMay 20, 2026Scratched Material

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED HALOS AT NIGHT AND SMUDGE ON LENS. CLINICAL REASON FOR EXPLANT WAS PSEUDOPHAKIC. THE IOL WAS PLANNED TO EXCHANGE FOR UNSPECIFIED ADVANCED TECHNOLOGY INTRAOCULAR LENSES (ATIOL) FOLLOWING THE INITIAL IMPLANT PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Visual Disturbances

    MAUDE report 25230281 ↗

Other Lens, Multifocal Intraocular devices

All Lens, Multifocal Intraocular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page