Skip to content

SatinSlit® 8065992561Recalls & reported problems

Alcon Laboratories, Inc. · SatinSlit® Knife 2.5mm Angled · Class 1

All models of SatinSlit (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380659925610), received between Mar 25, 2024 and Mar 25, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Material Puncture/Hole” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 15, 2024Class IIOpen, Classified

    Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.

    Z-2664-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Puncture/Hole1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 25, 2024Material Puncture/Hole; Material Puncture/Hole

    A HEALTHCARE PROFESSIONAL REPORTED THAT OPHTHALMIC SURGICAL BLADE HAD PUNCTURE THROUGH THE FOAM PACKAGING AND THE OPERATOR HAD A STAB INJURY UPON OPENING BEFORE CATARACT SURGERY WHICH WAS TREATED WITH BASIC FIRST AID. THE SURGERY WAS COMPLETED ON THE SAME DAY. THE PATIENT IMPACT WAS NOT REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 18976339 ↗

FDA clearance and classification

The FDA regulates product code HNN (knife, ophthalmic) under 21 CFR 886.4350 as a Class I device. 21 CFR 886.4350 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the SatinSlit® 8065992561

Has the SatinSlit® 8065992561 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2664-2024 (Class II), began on Jul 15, 2024: Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the SatinSlit® 8065992561?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Mar 25, 2024. That report lists device problem “Material Puncture/Hole” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SatinSlit® 8065992561?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SatinSlit® 8065992561 FDA approved?

Devices of this type (product code HNN) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Alcon Laboratories, Inc.

Who makes the SatinSlit® 8065992561, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand SatinSlit®, model 8065992561). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Knife, ophthalmic devices

All Knife, ophthalmic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page