Skip to content

ClearCut® 8065993045Recalls & reported problems

Alcon Laboratories, Inc. · ClearCut® Slit Knife 3.0mm Angled · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380659930454), received between May 24, 2023 and Mar 25, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Dull, Blunt” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 15, 2024Open, Classified

    Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.

    Z-2669-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Dull, Blunt4

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2026Dull, Blunt

    A CUSTOMER REPORTED THAT, PRIOR TO THE CATARACT SURGERY, AN OPHTHALMIC KNIFE WAS FOUND TO BE BENT UPON OPENING. THE SURGERY WAS COMPLETED USING AN ALTERNATIVE KNIFE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688428 ↗
  • MalfunctionAug 8, 2025Dull, Blunt; Dull, Blunt

    A CUSTOMER REPORTED AN OPHTHALMIC KNIFE WAS DULL. THE PROCEDURE AND PATIENT IMPACT DETAILS WERE NOT REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 22747523 ↗
  • MalfunctionSep 11, 2024Dull, Blunt; Dull, Blunt

    A CUSTOMER REPORTED THAT DURING CATARACT SURGERY AN OPHTHALMIC CUTTING BLADE WAS FOUND DULL. THE SURGERY WAS COMPLETED WITH THE REPLACED BLADE WITH NO PATIENT HARM.

    Patient problems: Insufficient Information

    MAUDE report 20189642 ↗
  • MalfunctionMay 24, 2023Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT DURING A CATARACT SURGERY AN OPHTHALMIC KNIFE WAS DULL. SURGERY WAS COMPLETED USING ANOTHER PRODUCT. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16991209 ↗