ARCHITECT 1L76-25Recalls & reported problems
Abbott Ireland Diagnostics · ARCHITECT Sirolimus Reagent Kit · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740001346), received between Aug 26, 2016 and Feb 20, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- High Test Results1
- Low Test Results1
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2018Low Test Results; Low Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT SIROLIMUS RESULTS FOR ONE PATIENT. THE CUSTOMER INDICATED THE UNDILUTED SAMPLE (B)(6) GENERATED AN INITIAL RESULT OF >30 NG/ML AND DILUTED 1:2 GENERATED 9.8 NG/ML. THE SAMPLE WAS RETESTED AND UNDILUTED GENERATED >30 NG/ML AND DILUTED 1:2 GENERATED 12.83 NG/ML. THE CUSTOMER FURTHER REPORTED A SECOND TECHNICIAN SETUP AND RAN THE SAMPLE OBTAINING AN UNDILUTED RESULT OF >30 NG/ML AND DILUTED 1:2 GENERATED 33.86 NG/ML. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7283278 ↗MalfunctionAug 26, 2016High Test Results; High Test Results
THE CUSTOMER STATED THAT PATIENTS BEING TREATED WITH EVEROLIMUS TEST POSITIVE FOR SIROLIMUS ON THE ARCHITECT ANALYZER. ON (B)(6) 2016 A PATIENT BEING TREATED WITH EVEROLIMUS GENERATED A POSITIVE SIROLIMUS RESULT BY ARCHITECT, BUT A RESULT OF <0.9NG/ML ON THE UPLC MASS SPEC. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5906437 ↗
Other CALIBRATOR, SECONDARY devices
- ARCHITECT 2K41-27Abbott Laboratories · CALIBRATOR, SECONDARY1 recall374 reports
- ARCHITECT 8K28-02Abbott Laboratories · CALIBRATOR, SECONDARY1 recall115 reports
- VITROS 8231474Ortho-Clinical Diagnostics, Inc. · Calibrator, secondary3 recalls0 reports
- BioPlex 2200 12000651Bio-Rad Laboratories, Inc. · Calibrator, secondary3 recalls0 reports
- ARCHITECT 08K2803Abbott GmbH · Calibrator, secondary1 recall93 reports
- ARCHITECT 7K63-25Abbott Ireland · CALIBRATOR, SECONDARY108 reports
- VITROS 8067324Ortho-Clinical Diagnostics, Inc. · Calibrator, secondary2 recalls0 reports
- VITROS 8568040Ortho-Clinical Diagnostics, Inc. · CALIBRATOR, SECONDARY1 recall18 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.