ARCHITECT 1P31-25Recalls & reported problems
Abbott GmbH · ARCHITECT iGentamicin Reagent Kit · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740001742), received between Nov 27, 2019 and May 4, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
1 other recall cover devices cleared under the same FDA submission (K102699); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- High Test Results1
- Leak/Splash1
- Noise, Audible1
- Smoking1
Patient problems
- No Patient Involvement1
- Skin Irritation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 4, 2020High Test Results; High Test Results
ABBOTT IDENTIFIED THAT SAMPLES TESTED FOR ARCHITECT STAT HIGH SENSITIVE TROPONIN-I AND STAT HIGH SENSITIVITY TROPONIN-I ASSAYS (LN 3P25 AND LN 2R98) MAY SHOW INTERACTION WHEN PROCESSED DIRECTLY AFTER THE ARCHITECT IGENTAMICIN (LN 1P31-25) ASSAY AND PATIENT RESULTS MIGHT BE IMPACTED. THERE HAS BEEN NO REPORTED HARM OR IMPACT TO PATIENT MANAGEMENT AS A RESULT OF THIS ISSUE.
Patient problems: No Patient Involvement
MAUDE report 10020975 ↗MalfunctionNov 27, 2019Leak/Splash; Smoking; Noise, Audible; Leak/Splash; Smoking; Noise, Audible
THE CUSTOMER REPORTED THAT WHEN TAKING OUT THE ARCHITECT GENTAMICIN CONTAINER FROM THE ANALYZER THERE WAS A "SMALL EXPLOSION/POPPING NOISE". THE TECHNICIAN SAW VAPOR/SMOKE WITHOUT ANY SMELL COMING FROM THE CONTAINER. A FEW SECONDS LATER, THE TECHNICIAN REALIZED THAT SHE HAD A SPLASH ON THE OUTER PART OF HER LEFT WRIST, AND IT WAS A LITTLE IRRITATED, AND ON THE RING FINGER OF HER RIGHT HAND SHE FELT SOMETHING LIKE A KIND OF EMBEDDED SPLINTER. THE AFFECTED AREAS WERE WASHED BUT NO ADDITIONAL MEDICAL INTERVENTION WAS SOUGHT. THERE WAS NO ADVERSE IMPACT TO PATIENT MANAGEMENT, USER SAFETY, OR FACILITY DAMAGE REPORTED.
Patient problems: Skin Irritation
MAUDE report 9384445 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K102699. One submission often covers many models, so check the product and lot details in each record.
- Z-2117-2020 · Apr 24, 2020 · Open, ClassifiedARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin t
Other ENZYME IMMUNOASSAY, GENTAMICIN devices
- Syva® Emit® 2000 Gentamicin Plus Assay 10445333Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, GENTAMICIN2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.